INTERVENTIONAL ANGIOGRAPHY SYSTEM recallZ-2392-2023 · Canon Medical System, USA, INC. · initiated Jun 22, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Canon Medical System, USA, INC. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The C-arm, used to set the position and angle of the x-ray exposure field, may not function due to the tableside console PWB, which needs replacement. There may be a delay in the procedure, but procedures can continue using the operators console, the switches on the back of the column, switches on the side of the flat panel detector, and the emergency stop will continue to function normally.

FDA root cause category: Nonconforming Material/Component

What the company is doing

On 6/22/23 correction notices were emailed to customers who were advised of the following: If the table side control does not function then the procedure can continue using the operators console in the operator room, the switches on the back of the column and the switches on the side of the flat panel detector. The emergency stop should not be affected and will continue to function normally. Neither manual operation or auto positioning of the C-arm can be performed on the console during procedure, and the C-arm will not operate. A firm representative will contact you to schedule a time to replace the PWB on the console. Complete and return the customer reply form to [email protected] For questions, contact the InTouch team at [email protected] or your service representative at 800.521.1968

Recall details

Recall number
Z-2392-2023
Status
Open, Classified
Initiated by the company
Jun 22, 2023
Posted by the FDA
Aug 11, 2023
Recalling firm
Canon Medical System, USA, INC.
Quantity
12
Product code
OWB
FDA submissions
K210900
Distribution
US Nationwide Distribution: NY, CO, IL, CA, MD, MN, WA

Product description

INFX-8000C XGCP-930AA TABLE SIDE CONSOLE, INFX-8000C XGCP-930BA TABLE SIDE CONSOLE

Code information (models, lots, serial numbers)

UDI-DI: 04987670100192. INFX-8000C, Serial Numbers: BHA1992001, BHB2132003, SWC2242002, AFB1832003, SUC2262008, SUC2292010, BFB1972004, SUB2172002, SWC2262003