Skip to content

LifeSPARC recallZ-2465-2023 · Cardiac Assist, Inc · initiated Jun 30, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Cardiac Assist, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Epoxy used during manufacturing of the LifeSPARC Pump may have uncured epoxy. if undetectable curing may result in added risk of early pump failure,

FDA root cause category: Under Investigation by firm

What the company is doing

LivaNova - TandemLife made Initial contact by phone/email on June 30,2023 to ensure the affected pumps were quarantined and not used. Other details of the recall, including how the customer is to respond, are included in the customer letter issued July 14, 2023 provided via regular mail and/or email. Letter states reason for recall, health risk and action taken. For questions regarding this Medical Device Removal, please contact your TandemLife representative, or TandemLife Customer Quality at [email protected].

Recall details

Recall number
Z-2465-2023
Status
Open, Classified
Initiated by the company
Jun 30, 2023
Posted by the FDA
Aug 25, 2023
Recalling firm
Cardiac Assist, Inc
Quantity
2 units
Product code
KFM
FDA submissions
K183623
Distribution
US Nationwide distribution in the states of IL, FL.

Product description

LifeSPARC Pump (part number 5800-0000) contained within TandemLife LifeSPARC Priming Tray part number: 5800-1000. Note: The 5800-1000 model number is an individually packaged component contained within the Kit numbers: 5830-2916 LS TANDEMLIFE KIT DL29 5840-2417 LS TANDEMLIFE KIT - V24, A17

Code information (models, lots, serial numbers)

S/N and UDI: (1)Kit 5830-2916 Pump SN 00876971 UDI: ( 01)00814112020609(11)230411(17)250219(21)00876971 (2)Kit 5840-2417 Pump SN 00876972 UDI: (01)00814112020609(11)230411(17)250219(21)00876972

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

2 devices cleared under K183623 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page