Transducer 3D9-3v recallZ-2517-2023 · Philips Ultrasound, Inc. · initiated Jun 30, 2023

Completed

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Philips Ultrasound, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Transducer, an ultrasound system accessory, consists of two parts that are bonded together that may come apart due to chassis bonding issue, which may pose risk to users and patients of tissue damage, and electric shock if the built-in safety measures to prevent electric were to also fail.

FDA root cause category: Process control

What the company is doing

On 6/30/23, recall notices were sent to customers who were advised to do the following: 1) Stop usage and/or distribution of affected transducers and secure them in a location where they cannot inadvertently be used. 2) Circulate the notice to all users of this device so that they are aware of the issue. Retain this notice with your system(s) and ensure the letter is in a place likely to be seen/viewed. 3) Complete and return the response form via email to: [email protected] Field Service Engineers will visit your site to provide a replacement device, and collect the affected product. Contact Customer Care Solutions Center if you need further information or support concerning this issue at 1-800-722-9377. On 7/24/23 additional recall notices were distributed to customers informing them that additional devices were found to be affected.

Recall details

Recall number
Z-2517-2023
Status
Completed
Initiated by the company
Jun 30, 2023
Posted by the FDA
Sep 6, 2023
Recalling firm
Philips Ultrasound, Inc.
Quantity
449
Product code
ITX
FDA submissions
K042540, K043535, K081661, K132304, K222648, K120321
Distribution
Worldwide - US Nationwide distribution including in the states of FL, MN, AZ, NC, CA, MO, OH, WV, AL, NH, IN, IL, VT, PA, NY, NE, TX, WI, MI, KY, NV, MA, SD, LA, CO, TN, AK, GA, WA, UT, NM, ME, AR, SC and the countries of Italy, China, Netherlands, New Zealand, Hungary, Australia, France, Thailand, C¿te D'Ivoire, Germany, Canada, Mexico, Spain, Saudi Arabia, Uzbekistan, Greece, Yemen.

Product description

3D9-3v Transducer, accessory to Philips Ultrasound Systems: EPIQ Elite; Affiniti 30, 50, and 70; ClearVue 850; HD15; iU22; and Compact 5000

Code information (models, lots, serial numbers)

Show the full code information (4k characters)

UDI-DI:00884838061545, Serial Numbers: F07VBZ, F083JP, F086Z5, F088QC, F08JPF, F08K4H, F08VLP, F08VLY, F08W1C, F08W1D, F08W1J, F08W1K, F08W1L, F08W1M, F08W1Q, F08WXP, F08WXY, F08WZY, F08X01, F08X02, F08X03, F08X04, F08Y2T, F08ZR3, F08ZR4, F08ZR5, F092BQ, F092BR, F092BV, F097QQ, F097QV, F097QW, F09BTB, F09BTC, F09BTD, F09BTG, F09BTH, F09BTN, F09BTP, F09C1G, F09C1H, F09C1L, F09C1M, F09F3R, F09F3T, F09F3V, F09F3X, F09F3Y, F09F3Z, F09F40, F09F41, F09F42, F09F43, F09F45, F09F48, F09F49, F09F4C, F09F4D, F09F4F, F09F4G, F09F4H, F09JT0, F09JT2, F09JT3, F09MH4, F09MH5, F09MH6, F09MH7, F09MH8, F09MH9, F09MHB, F09MHC, F09MHG, F09MHH, F09MHK, F09MHL, F09MHM, F09MHN, F09MHP, F09N4J, F09N4K, F09N4M, F09N4N, F09N4P, F09NLM, F09NLN, F09NLP, F09NLT, F09V25, F09V26, F09V27, F09V28, F09V29, F09V2C, F09V2D, F09V2F, F09V2G, F09V2H, F09V2J, F09VH6, F09VH7, F09VH9, F09VHB, F09VWL, F09VWM, F09VWN, F09VWP, F09VWQ, F09XY4, F09XY5, F09XY6, F09XY7, F09XY8, F09XYB, F09XYD, F09XYF, F09XYG, F09XYH, F09YQ7, F09YQ8, F09YQ9, F09YQB, F0B16J, F0B16K, F0B16M, F0B16P, F0B1K5, F0B1K6, F0B1K7, F0B1K8, F0B1K9, F0B5KP, F0B5KQ, F0B5KR, F0B5KT, F0B5KV, F0B5KW, F0B6K9, F0B6KB, F0B6KC, F0B6KD, F0B6KF, F0B6KG, F0B7ND, F0B7NF, F0B7NG, F0B7NH, F0B7NJ, F0B7NK, F0B7NM, F0B7NN, F0B7NQ, F0B7NT, F0B7NX, F0B7NY, F0B7P0, F0B7P4, F0B7P5, F0B7X9, F0B7XC, F0B7XD, F0B7XG, F0B8G7, F0B8G8, F0B8G9, F0B8GB, F0B95D, F0B95G, F0BB01, F0BB02, F0BB03, F0BBGT, F0BBGX, F0BBGY, F0BBGZ, F0BCFQ, F0BCFR, F0BDMQ, F0BDMR, F0BDMT, F0BDMV, F0BFK8, F0BFKB, F0BFKC, F0BFXH, F0BFXJ, F0BFXK, F0BFXM, F0BFXN, F0BFXQ, F0BFXV, F0BHWP, F0BHWQ, F0BHWR, F0BHWT, F0BHWV, F0BHWW, F0BHZP, F0BHZR, F0BHZT, F0BHZV, F0BHZW, F0BHZZ, F0BJ00, F0BJ01, F0BJ9X, F0BJ9Y, F0BJB0, F0BJB1, F0BJB2, F0BJYQ, F0BJYR, F0BL03, F0BL06, F0BMC1, F0BMC4, F0BMC5, F0BMC6, F0BN4N, F0BN4T, F0BN4V, F0BR78, F0BR79, F0BR7B, F0BR7D, F0BRMH, F0BRMJ, F0BT87, F0BT88, F0BT8B, F0BT8C, F0BVPB, F0BVPC, F0BVPF, F0BW00, F0BW01, F0BW02, F0BW03, F0BW04, F0BW05, F0BXCT, F0BXCV, F0BXCW, F0BXCX, F0BXCY, F0BXT4, F0BXT5, F0BXT7, F0C3W1, F0C4DJ, B3TBRW, F01W12, F01ZG6, F01ZZ5, F024G2, F03Q35, F04DX8, F04DXC, F04DXD, F04F28, F04F29, F04F2G, F04F2H, F04F2L, F07GY0, F07XRT, F07XRV, F07XRX, F07Y4Q, F09NLQ, F09NLR, F0B95H, F0BBGW, B3PZBK, F03561, F03JJ7, F054CP, F08Y2Z, F05VJ4, F07QBQ, F021MP, F02WLR, F01GWF, B2ZHFB, B3PYF8, F00MC6, F01QCD, F03Z3Q, F05Q7Q, F09BTL, B2ZHX2, F04655, B3HX01, F09N4L, F026YN, F05Y8M, F045L3, B2ZVMB, F09BTM, F0BCFM, F0BB04, F0BFK9, F097QP, F061X4, F09F4B, F03K58, F08ZR1, F011M1, F01MPH, F02MX4, F01QCL, F06TCC, B35BYY, F09F3W, F0295R, F05DHN, F00XH9, F09XYC, B3Q08H, B1MKLW, B3HMKL, F009V6, F045L6, F06WH0, B3L4N2, F04F24, F01MV2, F0BT89, F0BVP9, B3FT4V, F04F26, F09F47, B3FH7X, F05J7K, F09JT4, B0WH4P, B3HY5P, B35JQ4, B3HLN6, F05PR5, F09MHD, F00Q08, F027TG, F049Y9, F09YQ6, F00MCQ, F00Z9K, F04Y32, F09F44, F0152T, F03LBN, B32MKY, B2JFNX, B3Q5WZ, B3L5BB, F0568H, B3FTPN, F00NQB, B2RVM9, B3LFRX, F009V5, B3LFT2, B3TCCV, F03MCP, B3L4FD, B1YQC1, F081WX, F05PR8, B3L4LZ, F033JD, F0B16N, B3LFJF, F02Y1L, F0BR7F, F0723Y, B2867Q, B3HX2C, F0152P, F07D6G, B3L4QJ, B2ZHB2, F009V0, B18NV8, F04444, B23M55, B3HWZT, B2ZR05, B32MK3, F02RND, F09C1J, F0B7NZ, F0B7P1, F0B7P3, F0BJ02, F0BL05, F0BL07, B25331, B3LDH7, F0603X, F097QR, F097QT, F09BTF, F09BTJ, F09BTK, F09C1K, F09F46, F09JT1, F09VH8, F0B7NL, F0B7NP, F0B7NR, F0B7NW, F0B7P2, F0B9ZZ, F0BB00, F0BBGV, F0BCFL, F0BCFN, F0BCFP, F0BDMW, F0BDMX, F0BFKD, F0BFXL, F0BFXP, F0BFXT, F0BFXW, F0BHZQ, F0BHZY, F0BJ9Z, F0BJYV, F0BMC3, B0T4K4, B2JDYH, B2PWBT, B2RL1T, B35JLT, B3TCYD, F01X0Z, F04TQ8, F05QVR, F06WH6, F07D6C, F09V2B, F0B7XB, F0B7XF, F0B8G6, F0B95B, F0B95C, F0BFK7

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page