Skip to content

For intraventricular indications mainly used for operations… recallZ-2518-2023 · Aesculap Implant Systems LLC · initiated Jul 20, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Aesculap Implant Systems LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Trocar manufactured with the shaft too long and does not meet manufacturing specifications. As a result, when used with the 30 neuroendoscope, a portion of the visual field is obscured.

FDA root cause category: Process change control

What the company is doing

Aesculap issued Urgent Medical Device Recall Notification Letter via USPS Certified Mail 20.JUL.2023. Letter states reason for recall, health risk and action to take: 1. Review the Recall Notification in its entirety and ensure that all users in your organization of the above-mentioned product, and all concerned persons are informed about this voluntary product recall. 2. Your Aesculap Sales Representative will be in contact with you to review your current inventory, aid in evaluating your devices and assist you in completing the attached Product Acknowledgement Form. 3. Do not destroy any affected product. 4. Utilizing the attached form, you and your Sales Representative will record the total number of individual impacted pieces. If you have no inventory remaining, a zero (0) will be entered on the form. 5. Return the completed "Product Acknowledgement Form" to Aesculap, Inc. Product Quality Excellence department by emailing the form to [email protected] or faxing the form to (610) 849-1197 after your inventory has been evaluated by your Sales Representative, even if the total inventory in your possession is zero (0). 6. Once Aesculap, Inc. receives your "Product Acknowledgement Form", a customer support representative will contact you with instructions on how to return any impacted pieces in your possession.

Recall details

Recall number
Z-2518-2023
Status
Open, Classified
Initiated by the company
Jul 20, 2023
Posted by the FDA
Sep 6, 2023
Recalling firm
Aesculap Implant Systems LLC
Quantity
25 units
Product code
GWG
FDA submissions
K983365
Distribution
US Nationwide distribution in the states of FL, ID, IL, MD, NC, NY, OR, RI, TX.

Product description

For intraventricular indications mainly used for operations within the brain ventricles. Product Code: FF399R

Code information (models, lots, serial numbers)

UDI-DI: 04038653065547 Lot Numbers: 52750146 52753381, 52756472, 52768317, 52805655, 52812596, 52816570, 52823878, 52830872. Serial Numbers: 4281,4282, 4284,4288, 4289,4291, 4308,4309, 4310, 4312, 4313,4336, 4340,4341, 4342,4402, 4411, 4412, 4413,4414, 4422,4439, 4447,4449, 4450

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page