Skip to content

TyTek Tension Pneumothorax Access Kit TPAK10: 10 gauge… recallZ-2611-2023 · Tytek Medical Inc · initiated Oct 5, 2020

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Tytek Medical Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Updated IFU to inlclude: (1) identify the risks associated with the hazard of needle decompression and (2) improve the artwork showing placement site location

FDA root cause category: Device Design

What the company is doing

TyTek New Product IFU was initially sent out to all who purchased the product on 08Dec2020 with notification on what was updated. An official field safety notice (FSN) was sent to customers on 02Aug2021.The changes to the IFUs were proactive in nature. There has been no change to the Products themselves. When used in a manner consistent with the IFUs, the Products are safe for their intended use, and we wish to reiterate that there have been no incidents whatsoever. For you reference I have attached the IFU that we originally sent you in December 2020. Please do not hesitate to contact us should you have any additional questions. The change in our IFUs for these products was implemented to: (1) identify the risks associated with the hazard of needle decompression and (2) improve the artwork showing placement site location.

Recall details

Recall number
Z-2611-2023
Status
Terminated
Initiated by the company
Oct 5, 2020
Posted by the FDA
Sep 19, 2023
Recalling firm
Tytek Medical Inc
Quantity
4400 units
Product code
GAA
Distribution
OH, SC, VA Foreign: Country Australia Denmark Finland Germany Hong Kong/China Israel SPAIN The Netherlands Turkey UAE UK

Product description

TyTek Tension Pneumothorax Access Kit TPAK10: 10 gauge, 3.25 inch- Intended for introduction into the body to facilitate removal of air from the pleural cavity in patients with a tension pneumothorax Model Number: TM-310

Code information (models, lots, serial numbers)

UDI: 00855204008020 All lots produced prior to Oct 5, 2020

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 92899).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page