Skip to content

Liquid Assayed Chemistry Control Premium Plus Level 3… recallZ-2609-2023 · Randox Laboratories Ltd. · initiated Aug 3, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Randox Laboratories Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

there has been a transcription error in the Instructions for Use (IFU) for the Liquid Assayed Chemistry Control Premium Plus Level 3. For Caeruloplasmin, the one and two standard deviation values printed on the sheet are incorrect, which may lead to misreporting patient results. Target and range values are correct, and risk is mitigated; most clinical labs will use the target and range values in clinical practice.

FDA root cause category: Under Investigation by firm

What the company is doing

A Medical Device Correction notification letter dated 8/3/23 was sent to customers. Action to be taken: " Review Quality Control data generated using the affected lot. " Discuss the contents of this notice with your Medical Director. " Complete and return the response form 12187-QA to [email protected] within five working days. Transmission of Field Safety Notice: Send a copy of the FSN to all affected customers and to those who need to be aware within your organisation. Please accept our apologies for any inconvenience caused. Thank you for your patience and understanding. If you have any questions or concerns please contact Randox Technical Services.

Recall details

Recall number
Z-2609-2023
Status
Open, Classified
Initiated by the company
Aug 3, 2023
Posted by the FDA
Sep 19, 2023
Recalling firm
Randox Laboratories Ltd.
Quantity
66 kits
Product code
JJY
FDA submissions
K140971
Distribution
US States: GA, OH, OR, PA, TX, VA

Product description

Liquid Assayed Chemistry Control Premium Plus Level 3, Catalog Number LAE4215

Code information (models, lots, serial numbers)

UDI-DI: 05055273208986; Lot Number: 1308UE

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates