Skip to content

Attest recallZ-2638-2023 · 3M Company - Health Care Business · initiated Aug 22, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from 3M Company - Health Care Business or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Specific lots of 3M Attest Super Rapid Readout Biological Indicators (1492V), 3M Attest Super Rapid 5 Steam-Plus Challenge (41482V, 41482VF), and 3M Attest Super Rapid Readout Steam Challenge (1496V) may contain biological indicators with a cap that melts, deforms and/or cracks following a steam sterilization cycle of 132C and 135C.

FDA root cause category: Process control

What the company is doing

3M issued an Urgent product recall notice to its consignees on 08/24/2023 by email and US mail. The notice explained the problem with the product and requested the following: Discontinue distribution of the product and dispose of the product. Distributors were directed to notify their customers. The firm is offering replacement with the return of the completed response form.

Recall details

Recall number
Z-2638-2023
Status
Open, Classified
Initiated by the company
Aug 22, 2023
Posted by the FDA
Sep 26, 2023
Recalling firm
3M Company - Health Care Business
Quantity
33504 devices
Product code
FRC
FDA submissions
K173519
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Belgium, Canada, China, Colombia, Hong Kong, Italy, Japan, Peru, Poland, South Korea, Spain, Taiwan.

Product description

3M Attest Super Rapid Readout Steam Challenge Pack, REF 1496V; designed to qualify and monitor dynamic-air removal steam sterilization processes

Code information (models, lots, serial numbers)

UDI/DI: 50707387762983( (shipper); Lot codes: 33P4NJ, exp. 01/01/2025; 33P56T, exp. 01/02/2025

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 92915).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page