Apellis Injection Kit 29g Injection Needle -single use… recallZ-2653-2023 · Apellis Pharmaceuticals, Inc. · initiated Aug 22, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Apellis Pharmaceuticals, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
19-gauge x 1" inch filter needles showed appearances of a small fraction of non-preferred characteristics. A causal relationship has not been established between the structural variations in this 19-gauge filter needle and the rare events of retinal vasculitis
FDA root cause category: Under Investigation by firm
What the company is doing
Apellis issued Medical Device Field Correction letters via email on 8/22/23 to Healthcare Practitioners and Distributors/Specialty Distributors and Pharmacy/ Logistics. The communication to the Healthcare Practitioners also includes update on the rare events of retinal vasculitis with real world use of SYFOVRE (pegcetacoplan injection). A causal relationship has not been established between the 19-gauge needle and the rare events of retinal vasculitis in the real world. Letter states reason for recall, health risk and action to take: Actions to be taken by the Customer/Distributor: Please note that as part of this field action, you are requested to: " Confirm your availability and participation by sending back an email to acknowledge receipt of this message. " Immediately cease distribution of the below injection kit lots and place any remaining inventory of the lots On Hold. " Provide a summary of current on-hand inventory of the injection kit lots below. " Provide a distribution report of the injection kit lots including customer name/address, lot shipped, quantity shipped, and date. Please provide the Distribution report in a maximum time of four business hours. Apellis will also be recommending that practitioners discontinue use of, and dispose, any remaining injection kits that contain the specific 19-gauge filter needle and recommend they instead use the injection kits containing the 18-gauge filter needle already in distribution. Apellis is now exclusively distributing injection kits with the 18-gauge filter needle. If you need to place a supplemental order to replace quarantined lots, please place them immediately through your normal ordering process and Apellis will expedite shipping replacement injection kits./ Apellis will also be notifying physicians that if they do not have an injection kit with the 18-gauge filter needle, they may request a new kit at no cost by emailing [email protected] or contacting their Apellis Territory Business Mana
Recall details
- Recall number
- Z-2653-2023
- Status
- Open, Classified
- Initiated by the company
- Aug 22, 2023
- Posted by the FDA
- Sep 29, 2023
- Recalling firm
- Apellis Pharmaceuticals, Inc.
- Quantity
- 61,943 units
- Product code
- FMI
- FDA submissions
- K222681
- Distribution
- US Nationwide distribution.
Product description
Apellis Injection Kit 29g Injection Needle -single use intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin and either an 18-gauge or a 19-gauge 5 micron filter needle. Ref: IVT-KIT-29G
Code information (models, lots, serial numbers)
UDI-DI: 00860008043672 Kit Lots: 223186 230036 230056 230236 230316 230326 230336 230406 231926 231946 231956 231986
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Apellis Injection Kit IVT-KIT-29GAltaviz, LLC · Needle, Hypodermic, Single Lumen1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.