Skip to content

BD Pyxis recallZ-2646-2023 · CareFusion 303, Inc. · initiated Aug 18, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from CareFusion 303, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

After dispensing a timed dose medication for a patient, the automated dispensing cabinet software may continue to show the already dispensed medication dose is still available for dispense, which could lead to duplicate medication administration.

FDA root cause category: Device Design

What the company is doing

On 8/18/23, correction notices were mailed to directors of Biomedical Engineering/Risk Management/Nursing/Pharmacy and distributors. They were asked to take the following actions: 1) Remind users that affected devices do not replace the electronic (or paper) patient record, specifically the medication order, which will remain accurate and serves as the source of truth for medication administration. Nursing best practice is to confirm the 5 rights of medication orders prior to administration. Some inpatient psychiatric units also utilize barcode medication administration (BCMA) which may detect duplicate administrations. If a duplicate medication is inadvertently administered, ensure the patient is appropriately monitored and follow organizational process for reporting of a medication error. 2. Please ensure this letter is forwarded to applicable facilities if devices have been transferred and that all appropriate personnel have been made aware of this issue. 3. Complete and return the response form. Adverse health consequences experienced with the use of this product should be reported to BD at 1-800-727-6102 or via bd.com/self-service The recall issue will be resolved in software v1.2, which will be provided to all customers approximately October 2023 via remote upgrade. On 8/26/24, correction-update notices were sent informing customers the field correction (software update to BD Pyxis MedBank MedPass software v1.2.1.17) was not effective. As a result, devices required an additional update to BD Pyxis MedBank MedPass software version v1.2.3.23.

Recall details

Recall number
Z-2646-2023
Status
Open, Classified
Initiated by the company
Aug 18, 2023
Posted by the FDA
Sep 28, 2023
Recalling firm
CareFusion 303, Inc.
Quantity
37
Product code
BRY
Distribution
US: FL, MI, KS

Product description

BD Pyxis MedBank MedPass Software, REF: 139088-01

Code information (models, lots, serial numbers)

UDI-DI: 10885403513817, Software versions prior to but not including v1.2.3.23

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates