DJO Surgical: Empowr Knee, Universal Tibia Baseplate, 6R… recallZ-2603-2023 · Encore Medical, LP · initiated Aug 4, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Encore Medical, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Orthopedic device components were not subjected to final sterilization before shipment, which if implanted may lead to varying degrees of infection, which may require treatment, prolonged monitoring, and may lead to revision surgery
FDA root cause category: Nonconforming Material/Component
What the company is doing
On 8/4/23, recall notices were mailed to surgeons and distributors. They were reminded that starting on 7/3/23 they were contacted by the firm, and asked to quarantine affected devices, and that the affected devices had since been retrieved by the firm's sales representatives. Surgeons were asked to take the following actions: 1) We ask that you reach out to your patient[s] who were implanted with affected components as identified above and consider whether it may be appropriate for your patient to be monitored more closely to watch for signs and symptoms of acute or chronic infection. 2) We have also provided a patient letter for you to provide to your patient[s] who was [were] implanted with an affected component[s] should you prefer a written notification to accompany your verbal contact. 3) Please make the hospital where you implanted the components aware of this action, as appropriate by providing a copy of this communication or verbally notifying the hospital. 4) Please review and complete the enclosed Voluntary Recall Response Form. If you have any questions phone the firm at 1(800)456-8696 between the hours of 7:30AM-6:00PM CST, or e-mail [email protected] In addition, distributors were asked to do the following: Provide the notification to any additional organizations or members within your facility or network where devices subject to this voluntary recall may have been transferred.
Recall details
- Recall number
- Z-2603-2023
- Status
- Open, Classified
- Initiated by the company
- Aug 4, 2023
- Posted by the FDA
- Sep 19, 2023
- Recalling firm
- Encore Medical, LP
- Quantity
- 20
- Product code
- JWH
- FDA submissions
- K173723
- Distribution
- US: AL, AZ, CA, CO, FL, IL, IN, KS, KY, LA, MA, MI, MN, MO, NC, NJ, NY, OH, PA, PR, RI, SC, TX, UT, VA, WA. OUS: Sweden, Australia, France, UK, Colombia, Germany
Product description
DJO Surgical: Empowr Knee, Universal Tibia Baseplate, 6R, REF: 352-02-106
Code information (models, lots, serial numbers)
UDI-DI/Lot/Expiration: 00190446258032/308V1112/May 25, 2029
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- DJO SURGICAL 352-02-106Encore Medical, L.P. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall3 reports
Other devices under the same FDA submission
7 devices cleared under K173723 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- DJO SURGICAL 352-02-105Encore Medical, L.P. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER2 reports
- DJO SURGICAL 352-01-108Encore Medical, L.P. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
- DJO SURGICAL 352-02-109Encore Medical, L.P. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
- DJO SURGICAL 352-04-102Encore Medical, L.P. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
- DJO SURGICAL 352-04-106Encore Medical, L.P. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
- DJO SURGICAL 354-10-040Encore Medical, L.P. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
- DJO SURGICAL 354-12-065Encore Medical, L.P. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 92969).
- Z-2604-2023 · Open, ClassifiedDJO Surgical: Delta Ceramic Femoral Head, 44 mm + 4.0 mm, REF: 400-03-443
- Z-2605-2023 · Open, ClassifiedDJO Surgical: CLP Hip, Stem, Polished Neck, Lateral Offset Implant, Size#3, REF: 426-11-030
- Z-2606-2023 · Open, ClassifiedDJO Surgical: Reverse Shoulder System (RSP) Semi Humeral Socket Insert, 32mm, REF: 508-01-032; Baseplate, Glenoid HA-Coat, RSP, 6.5mm X 30mm, REF: 508-32-104; Reverse Shoulder Syst…
- Z-2607-2023 · Open, ClassifiedDJO Surgical: Altivate Reverse, Humeral Stem, Small Shell, Size 6 x 108mm, REF: 533-06-108
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page