MAQUET CARDIOHELP recallZ-0290-2024 · Maquet Medical Systems USA · initiated Oct 12, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Maquet Medical Systems USA or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
It was found that the fixation (weld) employed by the supplier of the nut bolt to the slide rail plate was insufficient. The deficient weld allows the nut bolt to break off from the slide rail plate with slight manual force.
FDA root cause category: Device Design
What the company is doing
An URGENT MEDICAL DEVICE Correction notification letter dated 10/12/23 was sent to customers. Actions to be taken by the Customer: Our records indicate that you have received a CARDIOHELP-i System having a serial number that is affected by this voluntary Medical Device Correction. " Please examine your inventory immediately to determine if you have any affected CARDIOHELP System(s) in your inventory. " Should you have any affected CARDIOHELP-i System(s), your unit(s) will need to be returned to Getinge for correction. Getinge will contact your facility to arrange the return of your unit for correction. A CARDIOHELP loaner can be provided if needed. " Please do not to fixate any accessory at the slide rail until the slide rail holders have been replaced. " Affected CARDIOHELP-i can be used for intra-hospital transport only when placed on a Sprinter cart. " Inter-hospital transport is not permitted until the slide rail holders have been replaced. " Please forward this information to all current and potential CARDIOHELP-i users within your hospital / facility. " If you are a distributor who has shipped any affected products to customers, please forward this document to their attention for appropriate action. " Whether or not your facility has affected product(s) listed in this notice, please complete and sign the attached MEDICAL DEVICE- CORRECTION RESPONSE FORM (Page 4) to acknowledge that you have received this notification. Return the completed form to Getinge by e-mailing a scanned copy to [email protected] or by faxing the form to 1-866-313-4049. Type of Action by Getinge: A Getinge representative will contact you to arrange for the repair of your CARDIOHELP -i unit(s). Getinge will correct your CARDIOHELP -i unit(s) upon receipt and will return your unit(s) once correction has been completed. This voluntary correction only affects the products listed on Page 1; no other products are affected by this voluntary correction. We apologize for
Recall details
- Recall number
- Z-0290-2024
- Status
- Open, Classified
- Initiated by the company
- Oct 12, 2023
- Posted by the FDA
- Nov 9, 2023
- Recalling firm
- Maquet Medical Systems USA
- Quantity
- 1022 units
- Product code
- DTQ
- FDA submissions
- K133598
- Distribution
- US Nationwide.
Product description
Cardiohelp-i, blood oxygenation and carbon dioxide removal system, Part Numbers 701072780 (Cardiohelp-i), 701068550 (Slide Rail Holder)
Code information (models, lots, serial numbers)
UDI-DI: 04058863074863; Serial Numbers: 90414355 through 90414868
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- MAQUET CARDIOHELP CARDIOHELP-iMaquet Cardiopulmonary AG · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS5 recalls2 deaths64 reports
Product lines
Other devices under the same FDA submission
4 devices cleared under K133598 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- MAQUET CARDIOHELP CARDIOHELP SystemMaquet Cardiopulmonary AG · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS4 recalls18 deaths107 reports
- MAQUET CARDIOHELP BS 3/8x3/32 L1.7Maquet Cardiopulmonary AG · Console, Heart-Lung Machine, Cardiopulmonary Bypass2 recalls0 reports
- MAQUET CARDIOHELP CARDIOHELP Emergency Drive (w/o holder)Maquet Cardiopulmonary AG · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS1 recall0 reports
- MAQUET CARDIOHELP CARDIOHELP Base UnitMaquet Cardiopulmonary AG · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS2 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page