MAQUET CARDIOHELP CARDIOHELP-iRecalls & reported problems
Maquet Cardiopulmonary AG · The CARDIOHELP System is a blood oxygenation and carbon dioxide removal system used to pump blood through the extracorporeal bypass circuit for circulatory and/or pulmonary support for periods appropriate to cardiopulmonary bypass (up to six hours). It is also intended to provide circulatory and/or pulmonary support during procedures not requiring cardiopulmonary bypass (for periods up to six hours). The CARDIOHELP System in configuration with the HLS/HIT Set Advanced is intended to be used within the hospital environment and outside the hospital environment (for periods up to six hours), e.g. for intra- and inter-Hospital transport. The CARDIOHELP System in configuration with the QUADROX-iR is intended to be used in extracorporeal circulation during cardiopulmonary bypass in cardiac surgery (for periods for up to six hours). · Class 2
All models of MAQUET CARDIOHELP (5)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
5 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Maquet Cardiohelp-i is a heart-lung machine console used to oxygenate and pump blood. The 77 reports mostly describe malfunctions such as failure to read input signals, pressure and electrical problems, flow or pumping stops, and alarms. There are 77 reports, including 2 deaths and no injuries. It has 5 open recalls for leakage-current testing, venous probe cable breaks, incorrect IFU settings, a weak nut-bolt weld, and a loose transport-guard pin. Recent reports rose from 18 to 35.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 64
- Deaths reported
- 2
- Injuries reported
- 0
- Recalls naming it
- 5
Summary
The FDA's MAUDE database contains 64 adverse event reports that list this exact device (DI 04058863074863), received between Oct 12, 2023 and Sep 28, 2026 — classified as 2 deaths, 62 malfunctions.
The most frequently reported specific device problem is “Pressure Problem” (16 reports).
5 FDA recall records name this device's identifier, 5 still open.
4 other recalls cover devices cleared under the same FDA submission (K133598); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Nov 14, 2023Open, Classified
CARDIOHELP-i System was not properly tested to measure leakage current
Z-0595-2024 · Root cause: Process design · Recall details · FDA record ↗
- Nov 13, 2023Open, Classified
Firm received complaints of breaks in the insulation, shielding, or wires of the venous probe connection cable. Fracture of the venous probe cable may influence the management of extracorporeal support. Patient may be exposed to inappropriate blood flow, blood temperature, or hematocrit/hemoglobin levels as a result of the CARDIOHELP-i not receiving information from the venous probe.
Z-0605-2024 · Root cause: Component design/selection · Recall details · FDA record ↗
- Nov 6, 2023Open, Classified
Certain factory settings for the device were incorrectly described in the IFU-related to the warning limits for PVen and PAux, and the default settings for the venous bubble sensor (VBS) and the automatic lock in the Minimized Extracorporeal Circulation (MECC) ThApp
Z-0555-2024 · Root cause: Labeling design · Recall details · FDA record ↗
- Oct 12, 2023Open, Classified
It was found that the fixation (weld) employed by the supplier of the nut bolt to the slide rail plate was insufficient. The deficient weld allows the nut bolt to break off from the slide rail plate with slight manual force.
Z-0290-2024 · Root cause: Device Design · Recall details · FDA record ↗
- Feb 18, 2022Open, Classified
Potential for a loose cylindrical pin on the Transport Guard that may result in the jamming of the locking mechanism causing the Transport Guard to not open or, if the locking pin is fully disengaged from the assembly, a full disassembly of the Transport Guard may occur, resulting in a delay of support if a product exchange is necessary.
Z-0897-2022 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Pressure Problem16
- Failure to Read Input Signal13
- Electrical /Electronic Property Problem9
- Infusion or Flow Problem8
- Pumping Stopped6
- Corroded4
- Crack3
- Mechanical Problem3
Patient problems
- No Clinical Signs, Symptoms or Conditions63
- Sepsis2
Recent adverse event reports
The 25 most recent of 64 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 28, 2026Electrical /Electronic Property Problem
The event occurred in USA with no patient involvement. It was reported that the venous bubble sensor is defective. No harm to any person has been reported. As a bubble sensor defective alarm could lead to a zero flow mode, a report is required. Complaint ID (b)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50910894 ↗MalfunctionSep 23, 2026Incorrect, Inadequate or Imprecise Result or Readings; Infusion or Flow Problem
The event occurred in USA during training, with no patient involvement. It was reported that the RPM changed speeds randomly when demonstrating the disposable disconnect. No harm to any person has been reported. As a Flow Bubble Sensor issue or flow issue, can lead to a pump stop, if the intervention is set by the user, a report is required. Complaint (b)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50868009 ↗MalfunctionSep 22, 2026Mechanical Problem
The event occurred in the USA during visual inspection with no patient involvement. It was reported that there was a visible crack on the body of the flow bubble sensor. No harm to any person has been reported. As a Flow Bubble Sensor issue, can lead to a pump stop, if the intervention is set by the user, a report is required. Complaint ID: (b)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50852616 ↗MalfunctionSep 16, 2026Mechanical Problem
The event occurred in the USA. The failure was found prior to use. It was reported that two pins on a cable are bent and one is broken in half. No harm to any person has been reported. Additional information was received that the Cardiohelp came back from system restore with this damage. The affected cable is the cable of the flow bubble sensor, which was damaged during transport from system restore. As a damaged Flow bubble sensor can lead to a pump stop, if the intervention is set by the user, a report is required. Complaint ID (b)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50784959 ↗MalfunctionSep 9, 2026Mechanical Problem
It was reported that Emergency Drive was broken during use however customer does not want to provide demographics. Later it was found that the latch of the emergency drive was sticky and therefore was not moving. No harm to any person has been reported. As there is a potential Risk, that in emergency situations the patient cannot be supported via the HLS Set and Emergency drive till a new Cardiohelp is connected, a report is required. Complaint # (b)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50684120 ↗MalfunctionSep 8, 2026Electrical /Electronic Property Problem
The event occurred in USA during treatment. It was reported that the venous bubble sensor failed. The venous bubble sensor was switched out. No harm to any person has been reported. As a bubble sensor defective alarm could lead to a zero flow mode, a report is required. Complaint ID (b)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50665634 ↗MalfunctionSep 2, 2026Electrical /Electronic Property Problem; Pumping Stopped
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50562811 ↗MalfunctionSep 2, 2026Device Alarm System; Infusion or Flow Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50559083 ↗MalfunctionAug 5, 2026False Alarm
Complaint # (b)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50138345 ↗MalfunctionMar 24, 2026Failure to Read Input Signal
THE FAILURE WAS DETECTED WITH NO PATIENT INVOLVEMENT. IT WAS REPORTED THAT THE FLOW PROBE (FLOW BUBBLE SENSOR) IS NOT WORKING. NO HARM TO ANY PERSON HAS BEEN REPORTED. AS ANY FLOW ISSUE AND BUBBLE ALARM IS A HIGH PRIORITY ALARM, WHICH LEADS TO PUMP STOP, IF INTERVENTION IS SET BY THE USER, A REPORT IS NEEDED. COMPLAINT ID: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24679436 ↗MalfunctionMar 18, 2026Corroded; Pressure Problem
THE EVENT OCCURRED IN USA DURING A TRAINING. IT WAS REPORTED THAT THE PRESSURES COULD NOT BE ZEROED. THE CUSTOMER STATED THAT THE HLS CABLE IS GIVING ERRATIC PRESSURE VALUES ACROSS MULTIPLE CIRCUITS. THE CUSTOMER BELIEVES THAT THE HLS CABLE IS TWISTED AND BENT AND SUSPECTS IT IS DAMAGED. NO HARM TO ANY PERSON HAS BEEN REPORTED. A PRESSURE READING ISSUE, CAN LEAD TO A PUMP STOP, IF THE INTERVENTION IS SET BY THE USER, THEREFORE A REPORT IS REQUIRED. COMPLAINT ID: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24626028 ↗MalfunctionMar 5, 2026Device Sensing Problem
THE EVENT OCCURRED IN USA DURING PATIENT TREATMENT. IT WAS REPORTED THAT THE ¿VENOUS BUBBLE SENSOR DEFECTIVE¿ ERROR MESSAGE WAS DISPLAYED. THE CARDIOHELP DEVICE WAS NOT REPLACED, THE VENOUS BUBBLE SENSOR WAS CHANGED WITH ANOTHER BUBBLE SENSOR THAT WORKED. FURTHER IT WAS CONFIRMED THAT THERE WAS NO VISIBLE DAMAGE ON THE VENOUS BUBBLE SENSOR OR ITS CABLE. NO HARM TO ANY PERSON HAS BEEN REPORTED. AS A BUBBLE SENSOR DEFECTIVE ALARM COULD LEAD TO A ZERO-FLOW MODE, A REPORT IS REQUIRED AS THERE IS BARELY A FLOW. COMPLAINT ID (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24516685 ↗MalfunctionMar 3, 2026Failure to Read Input Signal
THE EVENT OCCURRED IN THE USA. IT WAS REPORTED, THAT THE CARDIOHELP ALARMED "VENOUS BUBBLE SENSOR DEFECTIVE" DURING TREATMENT. THE CUSTOMER USED ANOTHER SENSOR AND THIS SOLVED THE FAILURE. THE DEFECTIVE VENOUS BUBBLE SENSOR SHOWS A SLIGHT FRAYING ON THE CORD. THE CARDIOHELP WAS NOT REPLACED. NO HARM TO ANY PERSON HAS BEEN REPORTED. COMPLAINT ID: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24495160 ↗MalfunctionFeb 17, 2026Device Sensing Problem; Infusion or Flow Problem
COMPLAINT ID (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24362235 ↗MalfunctionFeb 4, 2026Obstruction of Flow; Infusion or Flow Problem; Pressure Problem; Noise, Audible
COMPLAINT ID (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24257059 ↗MalfunctionJan 20, 2026Crack
COMPLAINT ID: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24114291 ↗MalfunctionDec 24, 2025Failure to Charge; Charging Problem; Power Problem; Failure to Charge; Charging Problem; Power Problem
COMPLAINT ID (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23886534 ↗MalfunctionDec 4, 2025Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED THAT THERE WERE ISSUES GETTING A FLOW READING FROM THE CARDIOHELP DEVICE. NO HARM TO ANY PERSON HAS BEEN REPORTED. AS THERE WAS A FLOW ISSUE DURING TREATMENT, WHICH CAN CAUSE A PUMP STOP IF THE INTERVENTION IS SET BY THE USER, A REPORT IS REQUIRED. COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23715507 ↗MalfunctionDec 4, 2025Failure to Read Input Signal; Failure to Read Input Signal
COMPLAINT ID (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23714247 ↗MalfunctionNov 25, 2025Crack; Crack
COMPLAINT ID: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23642385 ↗MalfunctionNov 24, 2025Pumping Stopped; Pressure Problem; Pumping Stopped; Pressure Problem
COMPLAINT ID: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23630406 ↗MalfunctionNov 6, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
COMPLAINT ID: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23484166 ↗MalfunctionOct 17, 2025No Visual Prompts/Feedback; No Visual Prompts/Feedback
THE EVENT OCCURRED IN USA DURING TREATMENT. IT WAS REPORTED THAT THE CARDIOHELP SCREEN WAS NOT RESPONDING. WHEN INCREASING OR DECREASING RPM'S VIA THE ROTARY KNOB, THE SCREEN WAS NOT SHOWING THE INCREASE/DECREASE, HOWEVER, THE LED RPM'S AT THE TOP OF THE PUMP WAS SHOWING THE INCREASE/DECREASE OF RPM'S. CUSTOMER TRIED TO USE THE TOUCHSCREEN TO INCREASE/DECREASE THE RPM'S WITH NO RESPONSE. THERE WAS NO VISIBLE CRACK OR DENT ON THE SCREEN. THE CUSTOMER WAS ABLE TO SWITCH THE HLS SET TO A NEW CARDIOHELP. NO HARM TO ANY PERSON WAS REPORTED. AS THE DEVICE WAS REPLACED DURING TREATMENT, A REPORT IS REQUIRED. COMPLAINT ID (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23322915 ↗MalfunctionOct 17, 2025Pressure Problem; Pressure Problem
THE EVENT OCCURRED IN THE USA. IT WAS REPORTED THAT THERE WAS A PRESSURE READING ISSUE. NO HARM TO ANY PERSON HAS BEEN REPORTED. AS A PRESSURE READING ISSUE CAN LEAD TO A PUMP STOP IF THE INTERVENTION IS SET BY THE USER, A REPORT IS REQUIRED. COMPLAINT ID (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23322911 ↗MalfunctionSep 8, 2025Failure to Read Input Signal; Failure to Read Input Signal
IT WAS REPORTED THAT THE HLS CABLE IS READING DISCONNECTED, WHILE CONNECTED. THE FAILURE OCCURRED WITHOUT PATIENT INVOLVEMENT. NO HARM TO ANY PERSON HAS BEEN REPORTED. AS A PRESSURE ISSUE WAS REPORTED, WHICH IS A POTENTIAL RISK FOR THE PATIENT, A REPORT IS NEEDED. COMPLAINT ID: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22990902 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K133598. One submission often covers many models, so check the product and lot details in each record.
- Z-1300-2026 · Jan 9, 2026 · Open, ClassifiedBubble Sensor for 3/8" x 3/32" tubing; Model Number: 701055720;
- Z-1831-2024 · Apr 15, 2024 · Open, ClassifiedCARDIOHELP-i System. Blood oxygenation and carbon dioxide removal system.
- Z-1053-2024 · Dec 28, 2023 · Open, ClassifiedBubble Sensor (REF: 70105.5720)
- Z-0420-2024 · Oct 10, 2023 · Open, ClassifiedCARDIOHELP Emergency Drive, Part Numbers 701048002 and 701076205. The CARDIOHELP System is a blood oxygenation and carbon dioxide removal system.
Other CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS devices
- Terumo® Advanced Perfusion System 1 801188Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass3 recalls447 reports
- MAQUET CARDIOHELP CARDIOHELP SystemMaquet Cardiopulmonary AG · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS4 recalls18 deaths107 reports
- Terumo® Advanced Perfusion System 1 816300Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass300 reports
- Terumo® Advanced Perfusion System 1 801763Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass1 death230 reports
- Terumo® Advanced Perfusion System 1 816571Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass230 reports
- ESSENZ HLMLivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS1 recall1 death110 reports
- Terumo® Advanced Perfusion System 1 5773Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass1 recall78 reports
- S5 Roller Pump 150 RP 150LivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS1 death120 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page