Why it was recalled

Technopath Manufacturing Ltd. has identified an issue with the glass vial commodity used to fill the Multichem S Plus Level 1, Multichem S Plus Level 2, Multichem S Plus Level 3, Multichem IA Plus and Multichem P Controls. Some customers have experienced broken vials when receiving or thawing Technopath Multichem IA Plus and Multichem S Plus Controls. Breakages have been observed upon delivery, while thawing or while opening the vial. The breakages can result in the bottom of the vial breaking off, breakage in the neck thread area or cracking on the sides of the vial. This may expose the user to biohazardous material if cut and may cause a delay in patient results.

FDA root cause category: Unknown/Undetermined by firm

What the company is doing

LGC Group notified their sole US consignee (distribution partner) of the recall on 08/25/2023 via email. The notice explained the issue and the risk with the request to notify their end users. The end user was advised to take additional care when handling the glass vials due to the potential of breakages upon receipt, while thawing, or while opening the vials.

Recall details

Recall number
Z-0104-2024
Status
Open, Classified
Initiated by the company
Aug 25, 2023
Posted by the FDA
Oct 12, 2023
Recalling firm
Techno-Path Manufacturing Ltd.
Quantity
117155 units
Product code
JJY
Distribution
US Nationwide distribution in the state of Illinois.

Product description

Multichem S Plus, Part Codes: a) 08P88-10, b) 08P88-11, c) 08P88-12

Code information (models, lots, serial numbers)

a) 08P88-10, UDI/DI 05391523440776, Lot Numbers: 108032101, 110072101, 119112101, 126052201, 130112201, 131112201; b) 08P88-11, UDI/DI 05391523440783, Lot Numbers: 108032102, 110072102, 119112102, 126052202, 130112202, 131112202; c) 08P88-12, UDI/DI 05391523440790, Lot Numbers: 108032103, 110072103, 119112103, 126052203, 130112203, 131112203

Devices named in this recall

3 devices in the FDA device registry whose identifier appears in the recall's code information.