Skip to content

Multichem P, Part Code 08P90-10 recallZ-0106-2024 · Techno-Path Manufacturing Ltd. · initiated Aug 25, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Techno-Path Manufacturing Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Technopath Manufacturing Ltd. has identified an issue with the glass vial commodity used to fill the Multichem S Plus Level 1, Multichem S Plus Level 2, Multichem S Plus Level 3, Multichem IA Plus and Multichem P Controls. Some customers have experienced broken vials when receiving or thawing Technopath Multichem IA Plus and Multichem S Plus Controls. Breakages have been observed upon delivery, while thawing or while opening the vial. The breakages can result in the bottom of the vial breaking off, breakage in the neck thread area or cracking on the sides of the vial. This may expose the user to biohazardous material if cut and may cause a delay in patient results.

FDA root cause category: Unknown/Undetermined by firm

What the company is doing

LGC Group notified their sole US consignee (distribution partner) of the recall on 08/25/2023 via email. The notice explained the issue and the risk with the request to notify their end users. The end user was advised to take additional care when handling the glass vials due to the potential of breakages upon receipt, while thawing, or while opening the vials.

Recall details

Recall number
Z-0106-2024
Status
Open, Classified
Initiated by the company
Aug 25, 2023
Posted by the FDA
Oct 12, 2023
Recalling firm
Techno-Path Manufacturing Ltd.
Quantity
2964 units
Product code
JJY
Distribution
US Nationwide distribution in the state of Illinois.

Product description

Multichem P, Part Code 08P90-10

Code information (models, lots, serial numbers)

UDI/DI 05391523440769, Lot Numbers: 04530621P, 04561021P, 04610523P

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 93017).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page