Universal Viewer Workflow Manager, Image processing… recallZ-0143-2024 · GE Healthcare · initiated Sep 8, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE Healthcare or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
When using Universal Viewer or Centricity Universal Viewer with Workflow Manager (WFM), if a third-party reporting application is used to launch exams, it could display information on a different patient than WFM and the Viewer.
FDA root cause category: Software design
What the company is doing
GE Healthcare notified consignees on about 09/08/2023 via letter. Consignees were instructed that until the software update is available the device can still be utilized through an updated browser setting or through discontinuing the use of bi-directional reporting. A GE Healthcare representative will contact each consignee to arrange for the correction. Consignees were instructed to complete and return the provided acknowledgement response.
Recall details
- Recall number
- Z-0143-2024
- Status
- Open, Classified
- Initiated by the company
- Sep 8, 2023
- Posted by the FDA
- Oct 23, 2023
- Recalling firm
- GE Healthcare
- Quantity
- 57 units
- Product code
- LLZ
- FDA submissions
- K211312
- Distribution
- Nationwide distribution to AR, AZ, CA, CO, DE, FL, GA, IL, IN, ME, MI, MN, MO, NE, NJ, NM, NV, NY, OH, OK, PA, SC, SD, TN, TX. International distribution to Austria, Canada, France, Germany, Hong Kong, Israel, Italy, Philippines, Singapore, Thailand, Turkey, Vietnam.
Product description
Universal Viewer Workflow Manager, Image processing radiological system
Code information (models, lots, serial numbers)
UDI/DI 00195278379610, Software versions 7.0 SP1.1, 7.0 SP1, 7.0 SP0.0.5, 7.0 SP0.1.0. All product IDs with an affected software version are impacted.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Universal Viewer 8Ge Healthcare · System, image processing, radiological2 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 93053).
- Z-0142-2024 · Open, ClassifiedCentricity Universal Viewer Workflow Manager, Image processing radiological system
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page