Skip to content

Centricity recallZ-0142-2024 · GE Healthcare · initiated Sep 8, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from GE Healthcare or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

When using Universal Viewer or Centricity Universal Viewer with Workflow Manager (WFM), if a third-party reporting application is used to launch exams, it could display information on a different patient than WFM and the Viewer.

FDA root cause category: Software design

What the company is doing

GE Healthcare notified consignees on about 09/08/2023 via letter. Consignees were instructed that until the software update is available the device can still be utilized through an updated browser setting or through discontinuing the use of bi-directional reporting. A GE Healthcare representative will contact each consignee to arrange for the correction. Consignees were instructed to complete and return the provided acknowledgement response.

Recall details

Recall number
Z-0142-2024
Status
Open, Classified
Initiated by the company
Sep 8, 2023
Posted by the FDA
Oct 23, 2023
Recalling firm
GE Healthcare
Quantity
85 units
Product code
LLZ
FDA submissions
K182419
Distribution
Nationwide distribution to AR, AZ, CA, CO, DE, FL, GA, IL, IN, ME, MI, MN, MO, NE, NJ, NM, NV, NY, OH, OK, PA, SC, SD, TN, TX. International distribution to Austria, Canada, France, Germany, Hong Kong, Israel, Italy, Philippines, Singapore, Thailand, Turkey, Vietnam.

Product description

Centricity Universal Viewer Workflow Manager, Image processing radiological system

Code information (models, lots, serial numbers)

UDI/DI 00840682145794, Software versions 7.0 SP1.1, 7.0 SP1, 7.0 SP0.0.5, 7.0 SP0.1.0. All product IDs with an affected software version are impacted.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 93053).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page