THUNDERBEAT 5 mm,45 cm,Front-actuated Grip Type S recallZ-0109-2024 · Olympus Corporation of the Americas · initiated Sep 12, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Olympus Corporation of the Americas or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Thunderbeats blue seal button may remain in the engaged position after the button is released and not immediately return to a neutral position may result in prolonged surgery
FDA root cause category: Under Investigation by firm
What the company is doing
UPDATE: Olympus issued a revised letter for additional lots titled: Correction to letter dated September 13, 2023- Additional Lot numbers on January 11, 2024. Olympus issued Urgent Medical Device Recall Letter on 9/13/23 to Healthcare Professional. Letter states reason for recall, health risk and action to take: 1. Inspect your inventory and identify any THUNDERBEAT models and lots subject to this action. Please check all areas of the hospital to determine if any of these devices remain in inventory. The lot number can be found on the box or pouch. 2. Call your Olympus customer service representative at 1-800-848-9024, option 2, with the quantity, model, and lot number of affected devices. Olympus will issue a Return Material Authorization to return any affected product at no charge. Olympus will issue a credit to your facility upon return of affected product. 3. Access the Olympus recall portal to indicate that you have received this notification: a. Go to https://olympusamerica.com/recall. b. Enter the recall number: 0431 4. If you have distributed these devices outside your facility, please notify your customers of this matter immediately. Your notification to your customers may be enhanced by including a copy of this letter. Olympus requests that you report complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1.
Recall details
- Recall number
- Z-0109-2024
- Status
- Open, Classified
- Initiated by the company
- Sep 12, 2023
- Posted by the FDA
- Oct 12, 2023
- Recalling firm
- Olympus Corporation of the Americas
- Quantity
- 271 units
- Product code
- LFL
- FDA submissions
- K211838
- Distribution
- US Nationwide distribution.
Product description
THUNDERBEAT, 5MM, 45CM, Front-actuated Grip Type S Model: TB-0545FCS Indicated for open, laparoscopic (including single-site surgery) general surgery and gynecological surgery (including urologic, thoracic, plastic and reconstructive, bowel resections, cholecystectomies, Nissen fundoplication, adhesiolysis, oophorectomy, hysterectomies (both vaginal assisted and abdominal) etc) and endoscopic surgery or in any procedure in which cutting, vessel ligation (sealing and cutting), coagulation, grasping, and dissection is performed
Code information (models, lots, serial numbers)
UDI-DI: 4953170409684 Lot Numbers: KR310806 KR310809 KR313734 KR319658 KR326619 KR326628 KR333661 Lots Added January 11, 2024: KR310806 KR310809 KR333661 KR342666
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- THUNDERBEAT 5 mm,45 cm, Front-actuated Grip Type S TB-0545FCSOlympus Medical Systems Corp. · Electrosurgical, cutting & coagulation & accessories3 recalls16 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 93071).
- Z-0108-2024 · Open, ClassifiedTHUNDERBEAT, 5MM, 35CM, Front-actuated Grip Type S. Model: TB-0535FCS Indicated for open, laparoscopic (including single-site surgery) general surgery and gynecological surgery (in…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page