THUNDERBEAT 5 mm,45 cm, Front-actuated Grip Type S TB-0545FCSRecalls & reported problems

Olympus Medical Systems Corp. · THUNDERBEAT 5 mm,45 cm,Front-actuated Grip Type S · Class 2

All models of THUNDERBEAT 5 mm,45 cm,Front-actuated Grip Type S (3)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
16
Deaths reported
0
Injuries reported
4
Recalls naming it
3

Summary

The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 04953170409684), received between Dec 7, 2018 and Jan 27, 2026 — classified as 4 injuries, 12 malfunctions.

The most frequently reported specific device problem is “Break” (12 reports).

3 FDA recall records name this device's identifier, 3 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 2 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 22, 2025Class IIOpen, Classified

    Firm is initiating a removal due to continued reports of adverse events.

    Z-1371-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Nov 25, 2024Class IIOpen, Classified

    The probe tips of the hand instruments are being damaged or breaking including pad damage and detachment.

    Z-0994-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  3. Sep 12, 2023Class IIOpen, Classified

    Thunderbeats blue seal button may remain in the engaged position after the button is released and not immediately return to a neutral position may result in prolonged surgery

    Z-0109-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break12
  • Detachment of Device or Device Component4
  • Activation Problem2
  • Degraded1
  • Peeled/Delaminated1

Patient problems

  • No Clinical Signs, Symptoms or Conditions12
  • Foreign Body In Patient2
  • No Consequences Or Impact To Patient2
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 27, 2026Degraded

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE ULTRASONIC SURGICAL DEVICE TISSUE PAD IN THE GRASPING SECTION WAS WORN AWAY. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24190417 ↗
  • InjuryJan 15, 2026Break; Detachment of Device or Device Component

    NO ADDITIONAL INFORMATION WAS RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24087582 ↗
  • MalfunctionDec 23, 2025Break; Break

    THE CUSTOMER REPORTED THAT, EARLY ON IN A THERAPEUTIC COLORECTAL RESECTION PROCEDURE, THE PAD OF THEIR THUNDERBEAT DEVICE BECAME LOOSE AFTER SEAL ACTIVATION. THE CUSTOMER FINISHED THE PROCEDURE WITH A REPLACEMENT OLYMPUS DEVICE, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23885048 ↗
  • MalfunctionDec 8, 2025Peeled/Delaminated; Peeled/Delaminated

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE THUNDERBEAT TISSUE PAD HAD SPLIT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23744238 ↗
  • MalfunctionOct 30, 2025Break; Break

    IT WAS REPORTED THE ULTRASONIC PROBE BROKE IN HALF DURING A THERAPEUTIC GASTRIC BYPASS PROCEDURE. THE PROCEDURE WAS COMPLETED WITH AN OLYMPUS BACK-UP DEVICE. THERE WERE NO REPORTS OF PATIENT HARM OR INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23425847 ↗
  • MalfunctionJul 3, 2025Break; Break

    IT WAS REPORTED THE THUNDERBEAT FRONT GRIP TYPE S PROBE BROKE DURING USE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22396216 ↗
  • InjuryOct 7, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    IT WAS REPORTED THAT TWO ULTRASONIC SURGICAL DEVICE USED DURING A THERAPEUTIC CELIOSCOPY (DISSECTION AROUND A POLYPROPYLENE ELEMENT), HAD THE SCISSOR BROKE (LOST JAWS) IN THE PATIENT'S BELLY DURING ADHESIOLYSIS. AFTER THE SECOND BREAKAGE, A MONOPOLAR HOOK AND ENDOBAG WERE USED TO REMOVE THE BROKEN PIECES. THE FIRST DEVICE WAS USED FOR ABOUT 30 MINUTES AND THE SECOND ONE FOR ABOUT 10 MINUTES. THERE WERE NO REPORTS OF PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20385176 ↗
  • InjuryOct 7, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    IT WAS REPORTED THAT TWO ULTRASONIC SURGICAL DEVICES USED DURING A THERAPEUTIC CELIOSCOPY (DISSECTION AROUND A POLYPROPYLENE ELEMENT), HAD THE SCISSOR BROKE (LOST JAWS) IN THE PATIENT'S BELLY DURING ADHESIOLYSIS. AFTER THE SECOND BREAKAGE, A MONOPOLAR HOOK AND ENDOBAG WERE USED TO REMOVE THE BROKEN PIECES. THE FIRST DEVICE WAS USED FOR ABOUT 30 MINUTES AND THE SECOND ONE FOR ABOUT 10 MINUTES. THERE WERE NO REPORTS OF PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20385140 ↗
  • InjurySep 3, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    IT WAS REPORTED THE THUNDERBEAT JAW BROKE OFF DURING A THERAPEUTIC SLEEVE GASTRECTOMY PROCEDURE AN FELL INTO THE PATIENT CAVITY. THE DEVICE WAS REPLACED AND THE PROCEDURE WAS COMPLETED. THERE WAS NO REPORTED PATIENT HARM.

    Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20130209 ↗
  • MalfunctionJul 22, 2024Break; Break

    DURING THE DEVICE EVALUATION, THE THUNDERBEAT FRONT-ACTUATED GRIP EXHIBITED THE PROBE BROKEN AT ITS DISTAL END. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19801796 ↗
  • MalfunctionJul 1, 2022Break; Break

    AS REPORTED FOR THIS EVENT BY THE CUSTOMER, SHORTLY AFTER STARTING A THERAPEUTIC LAPAROSCOPY-ASSISTED DISTAL GASTRECTOMY (LADG) PROCEDURE, THE DEVICE PROBE BROKE OFF INTO THE PATIENT. THE BROKEN PIECE WAS RETRIEVED FROM THE PATIENT. THE PROCEDURE WAS COMPLETED WITH A NEW SIMILAR DEVICE. THERE IS NO HARM OR ADVERSE IMPACT TO THE PATIENT. THE OUTPUT OF THE DEVICE WAS DONE ONLY A FEW TIMES WITH THE TISSUE PAD HARDLY WORN WHEN THE EVENT OCCURRED. THE SETTINGS AT THE TIME OF THE EVENT ARE NOT KNOWN. THE USER IS EXPERIENCED IN THE USE OF THE DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14889432 ↗
  • MalfunctionMay 26, 2021Break; Break

    AS REPORTED FOR THIS EVENT, DURING AN UNKNOWN PROCEDURE THE DEVICE TEFLON COATING BROKE OFF INTO THE PATIENT AND WAS RETRIEVED. THERE IS NO HARM OR ADVERSE IMPACT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11886487 ↗
  • MalfunctionDec 8, 2020Break; Break

    WE RECEIVED THE FOLLOWING REPORT. DURING A PROCEDURE OF COELIOSCOPY, THE SUBJECT DEVICE WAS USED. THE DISTAL END OF THE DEVICE BROKE OFF AND FELL INSIDE THE PATIENT. THE FRAGMENT WAS RETRIEVED. THE INTENDED PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10964039 ↗
  • MalfunctionOct 30, 2019Activation Problem; Activation Problem

    THE MANUFACTURER WAS INFORMED THAT THE DEVICE FAILED DURING A PROCEDURE. THE GENERATOR INDICATED A PROBE DAMAGE ERROR CODE. THE DEVICE WAS REPLACED AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9255369 ↗
  • MalfunctionAug 5, 2019Activation Problem; Activation Problem

    SERVICE CENTER WAS INFORMED THAT DURING A PROCEDURE THE THUNDERBEAT DEVICE FAILED. THE GENERATOR INDICATED VIA AN ERROR CODE THAT THERE WAS PROBE DAMAGE. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO PATIENT INJURY REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8861175 ↗
  • MalfunctionDec 7, 2018Break; Break

    ON NOVEMBER 12, 2018 OLYMPUS RECEIVED A MEDWATCH, REPORT# (B)(4), WHICH STATES THAT DURING A LAPAROSCOPIC HYSTERECTOMY PROCEDURE, THE METAL TIP OF THE THUNDERBEAT HANDPIECE BROKE OFF THE IN THE PATIENT. THE TIP WAS RETRIEVED FROM THE PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 8144691 ↗

Other Electrosurgical, cutting & coagulation & accessories devices