Centricity recallZ-0140-2024 · GE Healthcare · initiated Sep 14, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE Healthcare or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
GE HealthCare has become aware of an issue where images from two different patients may be contained in a single study, when stored in Centricity PACS-IW V3.7.x, Centricity PACS-IW with Universal Viewer V5.0 and Centricity Universal Viewer V6.0 with PACS-IW foundation.
FDA root cause category: Software design
What the company is doing
GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 09/14/2023 by letter. The notice explained the problem, risk to the patient, and provided two workaround options for the continued use of the instruments until the software update is ready for installation. Questions: please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative
Recall details
- Recall number
- Z-0140-2024
- Status
- Open, Classified
- Initiated by the company
- Sep 14, 2023
- Posted by the FDA
- Oct 20, 2023
- Recalling firm
- GE Healthcare
- Quantity
- 329 units
- Product code
- LLZ
- FDA submissions
- K150420
- Distribution
- US and Austria, Azerbaijan, Bosnia and Herzegovina, Brazil, Canada, Chile, Colombia, France, Germany, India, Ireland, Italy, Kuwait, Malaysia, Netherlands, Netherlands Antilles, Panama, Poland, Russia, Saudi Arabia, Slovenia, South Africa, Spain, Switzerland, United Kingdom, Uruguay, Venezuela, Vietnam
Product description
Centricity Universal Viewer, software versions 6.0 SP0 through 6.0 SP7.1; Radiological Image Processing System
Code information (models, lots, serial numbers)
Show the full code information (15k characters)
UDI/DI 840682103800, Installed Product ID #: 102123-1-Centricity Universal Viewer-00558570, 102123-1-Centricity Universal Viewer-02157634, 1011-1-Centricity Universal Viewer-00558109, 1011-1-Centricity Universal Viewer-02097166, 1011-1-Centricity Universal Viewer-02097165, 100666-1-Centricity Universal Viewer-14194768, 100666-1-Centricity Universal Viewer-20084199, 9789-1-Centricity Universal Viewer-00558645, 4413-2-Centricity Universal Viewer-02288447, 4267-5-Centricity Universal Viewer-19618406, 100763-1-Centricity Universal Viewer-00558366, 101171-1-Centricity Universal Viewer-02398531, 101171-1-Centricity Universal Viewer-02398531, 104535-3-Centricity Universal Viewer-02396026, 104535-3-Centricity Universal Viewer-02396027, 104535-3-Centricity Universal Viewer-02396028, 104535-3-Centricity Universal Viewer-02396029, 11438-1-Centricity Universal Viewer-00558464, 104535-4-Centricity Universal Viewer-02396030, 104535-4-Centricity Universal Viewer-01806096, 4415-1-Centricity Universal Viewer-00558359, 4415-1-Centricity Universal Viewer-01973510, 4308-1-Centricity Universal Viewer-00558676, 4308-1-Centricity Universal Viewer-21200827, 4308-1-Centricity Universal Viewer-21200830, 4318-1-Centricity Universal Viewer-00558703, 4318-1-Centricity Universal Viewer-20080694, 4048-1-Centricity Universal Viewer-00558695, 4048-1-Centricity Universal Viewer-14767809, 4048-1-Centricity Universal Viewer-14767808, 4048-1-Centricity Universal Viewer-14767807, 4048-1-Centricity Universal Viewer-14767806, 4048-1-Centricity Universal Viewer-02256595, 4460-1-Centricity Universal Viewer-02255583, 4460-1-Centricity Universal Viewer-01819618, 4460-7-Centricity Universal Viewer-02255586, 4460-7-Centricity Universal Viewer-02255587, 101280-1-Centricity Universal Viewer-19630914, 101280-1-Centricity Universal Viewer-19630915, 101280-1-Centricity Universal Viewer-19630916, 101280-1-Centricity Universal Viewer-20553950, 9768-1-Centricity Universal Viewer-00558661, 9768-1-Centricity Universal Viewer-01972254, 4552-1-Centricity Universal Viewer-01294074, 6579-11-Centricity Universal Viewer-02339896, 4353-2-Centricity Universal Viewer-01589220, 4698-1-Centricity Universal Viewer-00558313, 100029-1-Centricity Universal Viewer-02256808, 100717-1-Centricity Universal Viewer-01293807, 100717-1-Centricity Universal Viewer-01293808, 13162-8-Centricity Universal Viewer-01452133, 13162-8-Centricity Universal Viewer-14776447, 13162-8-Centricity Universal Viewer-14776449, 100019-1-Centricity Universal Viewer-00558666, 100019-1-Centricity Universal Viewer-01477448, 100019-1-Centricity Universal Viewer-01477447, 100019-1-Centricity Universal Viewer-15993728, 101725-2-Centricity Universal Viewer-00558642, 4687-1-Centricity Universal Viewer-00558334, 100489-1-Centricity Universal Viewer-01819402, 100489-1-Centricity Universal Viewer-20791988, 104249-1-Centricity Universal Viewer-00558667, 104287-2-Centricity Universal Viewer-00555505, 4451-1-Centricity Universal Viewer-00558107, 100366-1-Centricity Universal Viewer-00558522, 100366-1-Centricity Universal Viewer-14777071, 100718-2-Centricity Universal Viewer-02350739, 100718-1-Centricity Universal Viewer-02373799, 4742-2-Centricity Universal Viewer-01766912, 4742-2-Centricity Universal Viewer-20584004, 4742-2-Centricity Universal Viewer-20685485, 4583-1-Centricity Universal Viewer-01367198, 12968-1-Centricity Universal Viewer-02277843, 12968-1-Centricity Universal Viewer-02278145, 2337-5-Centricity Universal Viewer-01810490, 2337-5-Centricity Universal Viewer-19670358, 12902-1-Centricity Universal Viewer-00602638, 3245-1-Centricity Universal Viewer-01451973, 4585-3-Centricity Universal Viewer-01448365, 4585-3-Centricity Universal Viewer-20502096, 101039-1-Centricity Universal Viewer-01373579, 101039-1-Centricity Universal Viewer-14580319, 13345-1-Centricity Universal Viewer-00602707, 9852-2-Centricity Universal Viewer-00558644, 9852-2-Centricity Universal Viewer-02267839, 9852-2-Centricity Universal Viewer-01271417, 1530-1-Centricity Universal Viewer-00558491, 1530-12-Centricity Universal Viewer-01301357, 9764-1-Centricity Universal Viewer-01722652, 4391-1-Centricity Universal Viewer-02205222, 4391-3-Centricity Universal Viewer-01459013, 4698-11-Centricity Universal Viewer-19667347, 4698-11-Centricity Universal Viewer-15961473, 100159-1-Centricity Universal Viewer-00558324, 13295-1-Centricity Universal Viewer-01292684, 13295-1-Centricity Universal Viewer-19609993, 13295-1-Centricity Universal Viewer-19611189, 101598-1-Centricity Universal Viewer-00667138, 100971-1-Centricity Universal Viewer-00558469, 4382-1-Centricity Universal Viewer-02268547, 102768-1-Centricity Universal Viewer-02179353, 102768-1-Centricity Universal Viewer-02255613, 102768-1-Centricity Universal Viewer-16174584, 100636-1-Centricity Universal Viewer-01815255, 4637-1-Centricity Universal Viewer-00558372, 4637-1-Centricity Universal Viewer-02176556, 9094-1-Centricity Universal Viewer-01766830, 9094-1-Centricity Universal Viewer-01977629, 4433-1-Centricity Universal Viewer-00558704, 4433-1-Centricity Universal Viewer-17737529, 4682-1-Centricity Universal Viewer-02274840, 100864-1-Centricity Universal Viewer-00558400, 4669-1-Centricity Universal Viewer-00558691, 4669-1-Centricity Universal Viewer-14768587, 100104-1-Centricity Universal Viewer-16204735, 100104-1-Centricity Universal Viewer-01299102, 4440-1-Centricity Universal Viewer-00558353, 4440-1-Centricity Universal Viewer-02332746, 104491-1-Centricity Universal Viewer-01288014, 12961-1-Centricity Universal Viewer-00558236, 4449-1-Centricity Universal Viewer-00558276, 4449-1-Centricity Universal Viewer-20689478, 4305-7-Centricity Universal Viewer-16204605, 4305-7-Centricity Universal Viewer-16204606, 4305-7-Centricity Universal Viewer-16204607, 4305-7-Centricity Universal Viewer-16204608, 4305-7-Centricity Universal Viewer-16204609, 4305-7-Centricity Universal Viewer-16204612, 4305-7-Centricity Universal Viewer-16204613, 4306-3-Centricity Universal Viewer-00558573, 4306-1-Centricity Universal Viewer-02263351, 4306-3-Centricity Universal Viewer-02345023, 4306-3-Centricity Universal Viewer-02288781, 4362-1-Centricity Universal Viewer-01299095, 4362-1-Centricity Universal Viewer-19732280, 100581-1-Centricity Universal Viewer-01814001, 100092-1-Centricity Universal Viewer-15531177, 100092-1-Centricity Universal Viewer-14767331, 100092-1-Centricity Universal Viewer-00558648, 100092-1-Centricity Universal Viewer-00558697, 8746-1-Centricity Universal Viewer-00558308, 8746-1-Centricity Universal Viewer-01448329, 1057-1-Centricity Universal Viewer-02180910, 1057-1-Centricity Universal Viewer-00558402, 1057-1-Centricity Universal Viewer-14961014, 104336-1-Centricity Universal Viewer-00346858, 101690-1-Centricity Universal Viewer-00558680, 101690-1-Centricity Universal Viewer-20091140, 13125-1-Centricity Universal Viewer-00558110, 13125-1-Centricity Universal Viewer-20497227, 4351-1-Centricity Universal Viewer-01627896, 4351-1-Centricity Universal Viewer-16023960, 101170-1-Centricity Universal Viewer-00558305, 4483-1-Centricity Universal Viewer-00558457, 4483-1-Centricity Universal Viewer-20583582, 4742-3-Centricity Universal Viewer-02182462, 9796-1-Centricity Universal Viewer-00558571, 100770-1-Centricity Universal Viewer-01804127, 9752-1-Centricity Universal Viewer-02393053, 9752-1-Centricity Universal Viewer-00558454, 13225-1-Centricity Universal Viewer-02032334, 100769-1-Centricity Universal Viewer-01298556, 100769-1-Centricity Universal Viewer-02395799, 2403-7-Centricity Universal Viewer-01973509, 100468-1-Centricity Universal Viewer-02287985, 100468-1-Centricity Universal Viewer-02288300, 12966-1-Centricity Universal Viewer-01283020, 12966-1-Centricity Universal Viewer-20025013, 4683-1-Centricity Universal Viewer-00558456, 4601-1-Centricity Universal Viewer-00558699, 101310-1-Centricity Universal Viewer-02039946, 4347-1-Centricity Universal Viewer-00558520, 4487-1-Centricity Universal Viewer-01291589, 4487-1-Centricity Universal Viewer-02097167, AH4325-Centricity Universal Viewer-00160202, AH5291-Centricity Universal Viewer-00394902, AP6264-Centricity Universal Viewer-00000180, AP6469-Centricity Universal Viewer-01816421, AH0116-Centricity Universal Viewer-00160203, AH0106-Centricity Universal Viewer-00160204, 613493-Centricity Universal Viewer-02175791, 313741-Centricity Universal Viewer-01422997, 313741-Centricity Universal Viewer-02097256, 313741-Centricity Universal Viewer-15591262, 000282202-Centricity Universal Viewer-02291736, 571064-Centricity Universal Viewer-01293639, A54852-Centricity Universal Viewer-00043560, M41980-Centricity Universal Viewer-00044515, HC0399-Centricity Universal Viewer-01282972, PC1695-Centricity Universal Viewer-01447701, 786787-Centricity Universal Viewer-01283171, HBON01-Centricity Universal Viewer-00260801, 691749-Centricity Universal Viewer-01539092, 509940-Centricity Universal Viewer-01813036, HC0529-Centricity Universal Viewer-01364776, 683198-Centricity Universal Viewer-02263377, HC0245-Centricity Universal Viewer-00059319, 600791-Centricity Universal Viewer-02099049, 427990-Centricity Universal Viewer-01813160, 485135-Centricity Universal Viewer-01447706, 2336375-Centricity Universal Viewer-01444341, HC1414-Centricity Universal Viewer-01814335, PC1961-Centricity Universal Viewer-00001720, PC0023-Centricity Universal Viewer-01282279, 531344-Centricity Universal Viewer-00000707, CC0154-Centricity Universal Viewer-01813738, 509585-Centricity Universal Viewer-01826272, 601860-Centricity Universal Viewer-01373156, HC0565-Centricity Universal Viewer-00545595, HC4357-Centricity Universal Viewer-02090320, TP0575-Centricity Universal Viewer-01825777, 81111-Centricity Universal Viewer-01286673, A59927-Centricity Universal Viewer-00040382, A61971-Centricity Universal Viewer-01891850, AMS39-Centricity Universal Viewer-00545606, 268034-Centricity Universal Viewer-18862276, PL3673-Centricity Universal Viewer-00038810, PL1494-Centricity Universal Viewer-00038799, PL2970-Centricity Universal Viewer-00038797, PL1124-Centricity Universal Viewer-00038795, PL1499-Centricity Universal Viewer-00038800, PL1501-Centricity Universal Viewer-00038802, PL1190-Centricity Universal Viewer-21245580, PL1961-Centricity Universal Viewer-01981010, PL2624-Centricity Universal Viewer-00038804, PL1005-Centricity Universal Viewer-02336290, 600062-Centricity Universal Viewer-02095166, PL3362-Centricity Universal Viewer-20631670, PL0326-Centricity Universal Viewer-02040892, PL2975-Centricity Universal Viewer-00162605, PL2971-Centricity Universal Viewer-00038808, RU9255-Centricity Universal Viewer-19727239, RU8414-Centricity Universal Viewer-01445730, RU7519-Centricity Universal Viewer-01816694, RU4255-Centricity Universal Viewer-02331160, RU0003-Centricity Universal Viewer-02335931, RU9059-Centricity Universal Viewer-01445746, RU9327-Centricity Universal Viewer-02337466, RU1011-Centricity Universal Viewer-02330918, RU8366-Centricity Universal Viewer-18045776, RU1112-Centricity Universal Viewer-01453917, RU1112-Centricity Universal Viewer-01977813, RU1112-Centricity Universal Viewer-01980259, RU1516-Centricity Universal Viewer-02331146, RU1717-Centricity Universal Viewer-01476628, RU0472-Centricity Universal Viewer-02092677, RU3052-Centricity Universal Viewer-02330856, RU1107-Centricity Universal Viewer-01813331, RU9216-Centricity Universal Viewer-01982768, 600101-Centricity Universal Viewer-01273190, RU1544-Centricity Universal Viewer-02330871, RU2526-Centricity Universal Viewer-02331246, RU5388-Centricity Universal Viewer-01719556, RU0994-Centricity Universal Viewer-02331254, RU3794-Centricity Universal Viewer-19673911, RU9226-Centricity Universal Viewer-19673912, RU6968-Centricity Universal Viewer-02335840, RU8356-Centricity Universal Viewer-01445768, RU9595-Centricity Universal Viewer-01812133, 600105-Centricity Universal Viewer-02331152, RU9697-Centricity Universal Viewer-19673913, 2422939-Centricity Universal Viewer-01509505, RU0120-Centricity Universal Viewer-02335932, RU9546-Centricity Universal Viewer-01812747, RU3998-Centricity Universal Viewer-01553790, RU6164-Centricity Universal Viewer-19720917, SA1339-Centricity Universal Viewer-00116444, SI4014-Centricity Universal Viewer-01889278, 16218680-Centricity Universal Viewer-02266272, ZA2267-Centricity Universal Viewer-02001877, ZA2541-Centricity Universal Viewer-01985829, ZA2472-Centricity Universal Viewer-00116565, 482896-Centricity Universal Viewer-00545574, 941804-Centricity Universal Viewer-02287043, 85021-Centricity Universal Viewer-01426096, 318055-Centricity Universal Viewer-01457316; UDI/DI none, 4413-2-Centricity Universal Viewer-01300927, 4413-2-Centricity Universal Viewer-01300929, 4267-5-Centricity Universal Viewer-20884363, 4267-5-Centricity Universal Viewer-20884364, 4267-5-Centricity Universal Viewer-20884365, 4446-2-Centricity Universal Viewer-00558643, 4491-1-Centricity Universal Viewer-00558690, 100256-3-Centricity Universal Viewer-00558700, 100256-3-Centricity Universal Viewer-01810788, 102143-1-Centricity Universal Viewer-00558685, 101044-1-Centricity Universal Viewer-00558694, 100837-1-Centricity Universal Viewer-00558435, 100837-1-Centricity Universal Viewer-17594527, 13356-2-Centricity Universal Viewer-00558681, 4379-1-Centricity Universal Viewer-02184891, 102335-1-Centricity Universal Viewer-00558675, 13240-2-Centricity Universal Viewer-00558103, 13228-1-Centricity Universal Viewer-00558678, 101274-1-Centricity Universal Viewer-00558679, 101274-1-Centricity Universal Viewer-20279342, 1027-1-Centricity Universal Viewer-00558706, 1027-1-Centricity Universal Viewer-20759776, 13125-1-Centricity Universal Viewer-02268546, 102359-1-Centricity Universal Viewer-00558696, 4441-1-Centricity Universal Viewer-00558677, 4578-1-Centricity Universal Viewer-00558340, 2403-7-Centricity Universal Viewer-00558347, 11042-1-Centricity Universal Viewer-02285888, 11042-1-Centricity Universal Viewer-19603374, 280801-Centricity Universal Viewer-19737221, A91901-Centricity Universal Viewer-00044549, CC4105-Centricity Universal Viewer-01282278, HC4105-Centricity Universal Viewer-00001393, 233637-Centricity Universal Viewer-15773152, A54120-Centricity Universal Viewer-00040449, KW0000-Centricity Universal Viewer-01416913, BON01-Centricity Universal Viewer-00044131, PL3349-Centricity Universal Viewer-00038811, PL3277-Centricity Universal Viewer-00038831, RU3866-Centricity Universal Viewer-01445731, RU8945-Centricity Universal Viewer-01445740, RU2580-Centricity Universal Viewer-01445743, RU7175-Centricity Universal Viewer-01445748, RU1646-Centricity Universal Viewer-01445763, RU1107-Centricity Universal Viewer-01273207, RU3453-Centricity Universal Viewer-01273249, TBD-Centricity Universal Viewer-01273292, MPX352-Centricity Universal Viewer-19629107, DP7603-Centricity Universal Viewer-20093544, FP5028-Centricity Universal Viewer-00110911, 651948-Centricity Universal Viewer-01813739, 124774-Centricity Universal Viewer-00000208, 237870-Centricity Universal Viewer-01444278
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- CENTRICITY UNIVERSAL VIEWER 6.0Ge Healthcare · System, image processing, radiological8 recalls0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page