Cios Alpha recallZ-0195-2024 · Siemens Medical Solutions USA, Inc · initiated Sep 12, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Software issue for Cios Alpha, Cios Flow, and Cios Spin VA30 systems, fulfillment of new requirements from DIN and IEC standards for Cios VA30 systems required may result in patient misdiagnosis/repeat exposure
FDA root cause category: Software design
What the company is doing
Siemens Medical Solutions USA issued Urgent Medical Device Correction Letter (AX035/23/S).Letter states reason for recall, health risk and action to take. Siemens will correct the software error via Update Instruction AX036/23/S. With software version VA30K, the user interface of Cios VA30 systems will be changed to consider the current standards DIN 6862-3 and IEC 60601-2-54. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Customer Safety Advisory Notice. Please respond within 14 days of receiving this notification. If you have received this notification by email via Adobe GigaSign, kindly sign with your digital signature to acknowledge that you have read and understand the content provided. If you have received this customer notification by delivery by the US Postal Service or by FedEx, please send an email to [email protected] acknowledging that you have read and understand the content provided. Please ensure that all users of the affected products within your organization and others who may need to be informed will receive the relevant safety information provided within this notice and will comply with the recommendations therein.
Recall details
- Recall number
- Z-0195-2024
- Status
- Open, Classified
- Initiated by the company
- Sep 12, 2023
- Posted by the FDA
- Oct 27, 2023
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Quantity
- 237 units
- Product code
- OWB
- FDA submissions
- K210055
- Distribution
- Nationwide
Product description
Siemens Cios Alpha (VA30) mobile fluoroscopic C-arm X-ray System, designed for the surgical environment. 11105200
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
***Updated 12/4/23*** UDI-DI: 04056869153490 S/N: 40013 40014 40015 40018 40020 40023 40030 40032 40033 40034 40035 40036 40039 40040 40046 40047 40048 40049 40050 40051 40052 40054 40055 40057 40058 40062 40070 40073 40075 40076 40077 40078 40080 40083 40084 40093 40094 40095 40102 40103 40104 40107 40108 40109 40110 40111 40116 40117 40119 40121 40122 40123 40124 40125 40126 40130 40134 40135 40137 40139 40157 40158 40160 40161 40162 40163 40164 40167 40168 40169 40170 40171 40174 40175 40176 40188 40189 40190 40191 40192 40193 40194 40196 40206 40208 40215 40220 40226 40228 40230 40231 40234 40236 40238 40239 40240 40242 40244 40245 40246 40248 40250 40251 40252 40255 40256 40257 40259 40261 40263 40264 40265 40266 40267 40268 40270 40271 40273 40276 40314 40315 40317 40318 40322 40323 40324 40336 40339 42001 42002 42003 42004 42005 42006 42010 42018 42024 42025 42028 42032 42033 42037 42040 42041 42042 42045 42046 42052 42057 42061 42068 42069 42082 42087 42094 42101 42111 42140 42141 42146 42147 42148 42156 42160 42161 42185 42186 42187 42188 42189 42190 42191 42196 42199 42202 43000 43002 43003 43006 43007 43013 43022 43023 43025 43026 43027 43028 43029 43031 43033 43034 43037 43042 43044 43045 43046 43048 43049 43055 43058 43060 43061 43062 43063 43072 43073 43076 43079 43086 43095 43105 43106 43108 43112 43142 43143 43150 43161 43163 43177 43193 43208 43222 43226 43231 43257 43261 43279 43283 43286 43289 43291 43296 43297 43298 43319 43339
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Cios Alpha 11105200Siemens Healthcare GmbH · Interventional fluoroscopic x-ray system1 recall1 death6 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 93127).
- Z-0196-2024 · Open, Classified***Updated 12/4/23*** (1) Cios Flow (VA30) mobile fluoroscopic C-arm X-ray System. Designed for the surgical environment and provides comprehensive image acquisition modes. Mater…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page