Skip to content

Cios Alpha 11105200Recalls & reported problems

Siemens Healthcare GmbH · mobile C-arm; software version VA31A · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
6
Deaths reported
1
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04056869153490), received between Jan 19, 2021 and Jan 2, 2026 — classified as 1 death, 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 12, 2023Open, Classified

    Software issue for Cios Alpha, Cios Flow, and Cios Spin VA30 systems, fulfillment of new requirements from DIN and IEC standards for Cios VA30 systems required may result in patient misdiagnosis/repeat exposure

    Z-0195-2024 · Root cause: Software design · FDA record ↗

Most reported problems

Device problems

  • Insufficient Device Problem Information2
  • Radiation Output Failure2
  • Electrical /Electronic Property Problem1
  • Output Problem1
  • Unintended Electrical Shock1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Aneurysm1
  • Electric Shock1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 2, 2026Insufficient Device Problem Information

    SIEMENS BECAME AWARE OF A MALFUNCTION THAT OCCURRED WHILE OPERATING THE CIOS ALPHA VA3X SYSTEM. DURING AN EMERGENCY PATIENT PROCEDURE, THE IMAGE INTENSIFIER MALFUNCTIONED, WHICH LED TO A LONG TERM DELAY. WE HAVE NO INDICATIONS OF ANY ADVERSE EFFECTS ON THE HEALTH STATUS OF THE INVOLVED PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23943480 ↗
  • MalfunctionNov 8, 2023Radiation Output Failure; Radiation Output Failure

    SIEMENS BECAME AWARE OF A MALFUNCTION THAT OCCURRED WHILE OPERATING THE CIOS ALPHA VA30 SYSTEM. DURING AN INTERVENTIONAL PROCEDURE, NO X-RAY EXPOSURE WAS AVAILABLE. THE PATIENT HAD TO BE MOVED, AND THE PROCEDURE WAS COMPLETED ON AN ALTERNATE UNIT. WE HAVE NO INDICATIONS OF ANY ADVERSE EFFECTS ON THE HEALTH STATUS OF THE INVOLVED PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18091898 ↗
  • DeathOct 6, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    SIEMENS BECAME AWARE OF AN INCIDENT THAT OCCURRED WHILE OPERATING THE CIOS ALPHA SYSTEM. THE USER REPORTED THAT DURING AN EMERGENCY INTERVENTIONAL EXAMINATION TYPE ABDOMINAL AORTIC ANEURYSM. THE QUALITY IMAGE OF THE SYSTEM WAS NOT SUITABLE TO PERFORM THE EXAMINATION PROTOCOL ENDOPROSTHESIS. AS A RESULT, THE PHYSICIAN DECIDED TO PERFORM A SURGERY. DURING THE SURGICAL PROCEDURE THAT LASTED 4 HOURS, THE PATIENT PASSED AWAY. SIEMENS HAS REQUESTED ADDITIONAL INFORMATION IN ORDER TO CONDUCT AN INVESTIGATION OF THE REPORTED EVENT.

    Patient problems: Aneurysm

    MAUDE report 17886165 ↗
  • InjurySep 13, 2022Electrical /Electronic Property Problem; Unintended Electrical Shock; Electrical /Electronic Property Problem; Unintended Electrical Shock

    IT WAS REPORTED TO SIEMENS THAT AN INCIDENT OCCURRED WITH THE CIOS ALPHA SYSTEM. THE USER RECEIVED AN ELECTRIC SHOCK WHILE REMOVING THE X.10 CONNECTOR. THE OPERATOR COMPLAINED OF PAIN IN HIS FOREARM AND WAS EXAMINED IN THE ER. THE OPERATOR DID NOT SUFFER ANY INJURIES AND WAS FOUND TO BE IN GOOD HEALTH. SIEMENS HAS REQUESTED ADDITIONAL INFORMATION IN ORDER TO CONDUCT AN INVESTIGATION OF THE REPORTED EVENT. THE REPORTED EVENT OCCURRED IN SWITZERLAND.

    Patient problems: Electric Shock

    MAUDE report 15412481 ↗
  • MalfunctionSep 9, 2022Output Problem; Output Problem

    IT WAS REPORTED TO SIEMENS THAT A MALFUNCTION OCCURRED WHILE OPERATING THE CIOS ALPHA SYSTEM. DURING AN INTERVENTIONAL PROCEDURE WITH A PATIENT IN CRITICAL CONDITION, THE USER REPORTED THAT THE BRAKE ON THE SYSTEM COULD NOT BE RELEASED. THE CLINICAL PROCEDURE HAD TO BE TERMINATED. SIEMENS HAS REQUESTED ADDITIONAL INFORMATION IN ORDER TO CONDUCT AN INVESTIGATION OF THE REPORTED EVENT. HOWEVER, ADDITIONAL INFORMATION WAS RECEIVED THAT THE INVOLVED PATIENT HAD NOT BEEN INJURED. THE REPORTED EVENT IN JAPAN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15393267 ↗
  • MalfunctionJan 19, 2021Radiation Output Failure; Radiation Output Failure

    IT WAS REPORTED TO SIEMENS THAT A MALFUNCTION OCCURRED WHILE OPERATING THE CIOS ALPHA SYSTEM. DURING AN INTERVENTIONAL PROCEDURE, THE USER REPORTED THAT THE SYSTEM WAS BUSY AND NO FLUORO WAS POSSIBLE. THE PROCEDURE WAS CONTINUED AND COMPLETED ON AN ALTERNATE SYSTEM. WE ARE UNAWARE OF ANY IMPACT TO THE STATE OF HEALTH OF THE PATIENT INVOLVED. SIEMENS HAS REQUESTED ADDITIONAL INFORMATION IN ORDER TO CONDUCT AN INVESTIGATION OF THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11192688 ↗

Other Interventional fluoroscopic x-ray system devices

All Interventional fluoroscopic x-ray system devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.