GRAFTONand GRAFTON PLUSDemineralized Bone Matrix (DBM) recallZ-0166-2024 · Medtronic Sofamor Danek USA Inc · initiated Sep 26, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Sofamor Danek USA Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Specific serial numbers of GraftonTM DBM product due to the potential that a step during the inspection of the outer Tyvek pouch may not have been performed according to requirements. Based on this, batches may have been released with potential non-conformances in the outer Tyvek pouch.
FDA root cause category: Employee error
What the company is doing
Medtronic issued an URGENT: MEDICAL PRODUCT RECALL notice to its consignees on 09/26/2023 via UPS 2-day delivery. The notice explained the problem, potential risk, and requested the return of the affected devices.
Recall details
- Recall number
- Z-0166-2024
- Status
- Open, Classified
- Initiated by the company
- Sep 26, 2023
- Posted by the FDA
- Oct 25, 2023
- Recalling firm
- Medtronic Sofamor Danek USA Inc
- Quantity
- 583 units
- Product code
- MBP
- FDA submissions
- K043048, K051195
- Distribution
- US, Colombia, S. Korea, New Zealand, India, Taiwan
Product description
Grafton Plus DBM Paste, Model Numbers: a) T45001, b) T45005, c) T45010; Demineralized Bone Matrix
Code information (models, lots, serial numbers)
Show the full code information (7k characters)
a) T45001: GTIN 00643169122949, Serial Numbers: A48245-111, A58751-114, A57260-060, A57260-063, A58751-112, A58751-113, A57260-083, A57260-084, A58428-149, A58428-150, A57260-082, A58428-148, A58751-111, A57260-078, A58428-141, A48245-075, A48245-076, A57260-075, A58751-096, A57260-079, A58428-147, A58428-128, A57260-072, A57260-059, A57260-067, A57260-068, A57260-069, A48245-105, A48245-106, A57256-160, A58751-100, A57256-161, A48245-113, A57256-164, A58799-167, A58799-168, A58799-169, A58799-170, A58799-171, A57256-170, A48245-107, A48245-108, A57256-136, A58751-108, A48245-079, A48245-082, A48245-083, A48245-084, A48245-086, A58751-070, A58751-071, A58751-072, A58751-073, A58751-074, A58751-075, A58751-076, A58751-115, A58799-087, A48245-101, A58799-165, A58799-166, A48245-032, A48245-033, A48245-034, A48245-035, A48245-036, A57260-058, A58799-143, A58799-144, A48245-098, A48245-099, A58428-099, A58428-100, A58428-101, A58428-102, A58428-103, A57260-061, A48245-026, A48245-100, A48245-027, A48245-028, A48245-029, A48245-030, A48245-031, A58751-077, A58751-078, A58751-079, A58751-080, A57260-064, A57260-065, A57260-066, A58428-086, A58428-087, A58428-088, A57256-145, A57256-146, A57256-147, A57256-148, A57256-149, A57256-150, A48245-018, A58799-099, A58799-100, A58799-113, A58799-114, A48245-053, A58799-123, A48245-080, A48245-081, A58428-107, A58428-108, A58428-109, A58428-110, A57256-137, A57256-138, A57256-139, A57256-140, A58799-107, A58799-108, A58799-109, A58751-083, A58751-084, A58751-085, A48245-022, A48245-023, A58428-111, A58428-112, A58428-113, A58428-114, A58751-086, A58751-087, A58751-088, A58751-089, A57256-172, A58428-115, A58428-116, A58799-119, A58799-120, A58799-124, A58799-125, A48245-037, A48245-038, A57256-133, A57256-134, A57256-135, A57256-142, A57256-143, A57256-144, A58751-081, A58799-172, A58799-173, A58799-174, A58799-175, A58799-176, A58799-177, A58799-092, A58799-093, A58428-117, A58428-118, A48245-019, A48245-020, A48245-021, A58799-128, A58799-129, A58799-130, A58799-131, A58799-132, A48245-087, A48245-088, A57260-080, A57260-081, A58428-119, A58799-134, A58751-090, A48245-042, A48245-043, A48245-095, A48245-096, A48245-097, A58428-120, A58428-121, A58428-122, A58751-091, A58799-135, A58799-136, A58799-137, A58799-138, A58799-139, A48245-074, A48245-094, A48245-044, A48245-045, A48245-046, A58751-093, A58751-094, A58751-095, A58428-123, A58428-124, A58428-125, A58428-144, A58428-145, A58428-146, A48245-085, A48245-116, A48245-039, A48245-040, A48245-041, A58428-104, A58428-105, A58428-106, A58799-102, A58799-103, A58799-104, A58799-105, A48245-089, A48245-090, A48245-091, A48245-092, A48245-093, A48245-077, A57260-076, A57260-077, A58428-126, A58428-127, A58799-121, A58799-122, A58799-126, A58799-127, A58799-094, A58799-095, A48245-112, A58751-097, A58751-092, A58799-097, A58799-098, A58428-130, A58428-131, A58428-132, A58428-133, A58428-134, A57260-073, A57260-074, A58799-088, A58799-089, A58799-090, A58799-091, A58799-140, A58799-141, A58799-142, A58799-145, A58799-146, A58799-147, A58799-148, A58799-149, A58799-150, A58799-151, A58799-152, A58799-153, A58799-154, A58799-155, A58799-156, A58799-157, A58799-158, A58799-159, A58799-160, A58799-161, A58799-162, A57256-171, A48245-054, A48245-055, A58751-098, A58751-099, A48245-117, A48245-118, A48245-119, A48245-120, A57260-070, A57260-071, A48245-102, A58799-101, A58799-163, A58799-164, A58799-106, A58799-110, A58799-111, A58799-112, A58799-115, A58799-116, A58799-117, A58799-118, A57256-141, A58428-139, A48245-109, A48245-110, A57256-157, A57256-158, A57256-159, A48245-114, A48245-115, A48245-056, A48245-057, A48245-068, A48245-069, A48245-070, A48245-071, A48245-072, A48245-073, A48245-078, A58751-101, A58751-102, A58751-103, A58751-104, A58751-105, A48245-048, A48245-049, A48245-050, A48245-051, A48245-052, A58428-090, A58428-091, A58428-092, A58428-093, A58428-135, A58428-136, A58751-106, A58799-133, A58751-109, A58751-110, A58751-082, A58799-096, A57256-162, A57256-163, A58428-142, A48245-058, A58428-129, A57256-165, A48245-024, A48245-025, A58428-089, A58428-143, A58751-067, A58751-068, A58751-069, A57256-166, A58428-140, A48245-047, A58428-137, A58428-138, A57256-151, A57256-152, A57256-153, A57256-154, A57256-155, A57256-156, A58428-094, A58428-095, A58428-096, A58428-097, A58428-098, A48245-059, A48245-060, A48245-061, A48245-062, A48245-063, A48245-064, A48245-065, A48245-066, A48245-067, A57260-062, A58751-107, A57256-167, A57256-168, A57256-169, A48245-103, A48245-104; b) T45005: GTIN 00643169123021, Serial Numbers: A58746-016, A57923-013, A59083-037, A57923-022, A57878-072, A58720-014, A58720-015, A57261-045, A57261-046, A57261-047, A59083-048, A58973-023, A58973-024, A58973-025, A58973-026, A58973-027, A59132-021, A59132-022, A59132-023, A59132-024, A59132-025, A59132-026, A59132-027, A59132-028, A59132-029, A59132-030, A57878-063, A57878-064, A57878-065, A57878-066, A57878-067, A58973-029, A58973-033, A58746-013, A58746-015, A58746-014, A57923-021, A59083-049, A57923-011, A57923-012, A58746-018, A58746-019, A58746-020, A58746-021, A58746-024, A57878-068, A57878-069, A58973-028, A58746-023, A58746-025, A58746-029, A58746-030, A58746-031, A58746-032, A57878-070, A57878-071, A58746-026, A58746-027, A57923-014, A57923-015, A57261-039, A57261-040, A57261-041, A57261-042, A57261-043, A57261-044, A58746-028, A59083-036, A59083-038, A58746-011, A58746-012, A58746-017, A58973-011, A58973-012, A58973-013, A58973-014, A58973-015, A58973-016, A58973-017, A58973-018, A58973-019, A58973-020, A58973-021, A58973-022, A58720-023, A57878-073, A57878-074, A57261-038, A59083-039, A59083-040, A59083-041, A57261-056, A57878-075, A57878-076, A57261-048, A57261-049, A57261-050, A58720-011, A58720-012, A57261-051, A57261-052, A59083-042, A57261-055, A57261-057, A59083-043, A57261-053, A57261-054, A58720-013, A58720-016, A58720-017, A59083-044, A57923-016, A57923-017, A57923-018, A57923-023, A58973-030, A58973-031, A59083-045, A59129-032, A59129-041, A57923-019, A58720-018, A58720-019, A58720-020, A58720-021, A58720-022, A58973-032, A57923-020, A59083-046, A59083-047, A58746-022, A59132-011, A59132-012, A59132-013, A59132-014, A59132-015, A59132-016, A59132-017, A59132-018, A59132-019, A59132-020; c) T45010: GTIN 00643169123038, Serial Numbers: A59231-018, A59231-019, A56269-064, A59231-014, A56269-066, A59010-019, A59010-021, A59010-022, A59010-023, A56269-071, A56269-072, A59356-017, A59356-018, A59131-019, A59131-020, A59191-026, A59231-020, A59231-015, A59231-016, A59231-017, A59010-024, A59010-025, A56269-065, A59131-011, A59131-012, A59131-013, A59356-011, A59356-012, A59356-013, A59231-013, A56269-061, A56269-062, A56269-063, A59010-026, A59010-027, A59131-014, A59131-015, A59131-016, A59131-017, A59131-018, A59191-027, A59191-028, A59191-029, A59191-030, A59191-031, A59191-032, A59191-033, A59010-020, A56269-067, A56269-068, A59191-034, A59191-035, A59191-036, A59356-020, A59191-037, A59231-011, A59356-019, A56269-069, A56269-070, A59010-016, A59010-017, A59231-012, A59356-014, A59356-015, A56269-060, A59010-018, A59356-016
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T45010Medtronic Sofamor Danek, Inc. · Filler, bone void, osteoinduction (w/o human growth factor)2 recalls2 reports
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T45001Medtronic Sofamor Danek, Inc. · Filler, bone void, osteoinduction (w/o human growth factor)2 recalls0 reports
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T45005Medtronic Sofamor Danek, Inc. · Filler, bone void, osteoinduction (w/o human growth factor)2 recalls0 reports
Product lines
Other devices under the same FDA submission
30 devices cleared under K043048, K051195 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T44135Medtronic Sofamor Danek, Inc. · Filler, bone void, osteoinduction (w/o human growth factor)2 recalls1 report
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T44150Medtronic Sofamor Danek, Inc. · FILLER, BONE VOID, CALCIUM COMPOUND2 recalls1 report
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T42275Medtronic Sofamor Danek, Inc. · FILLER, BONE VOID, CALCIUM COMPOUND2 recalls1 report
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T44115Medtronic Sofamor Danek, Inc. · Filler, bone void, osteoinduction (w/o human growth factor)2 recalls1 report
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T44145Medtronic Sofamor Danek, Inc. · FILLER, BONE VOID, CALCIUM COMPOUND2 recalls0 reports
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T42100Medtronic Sofamor Danek, Inc. · Filler, bone void, osteoinduction (w/o human growth factor)2 recalls0 reports
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T42200Medtronic Sofamor Danek, Inc. · FILLER, BONE VOID, CALCIUM COMPOUND2 recalls0 reports
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T42210Medtronic Sofamor Danek, Inc. · Filler, bone void, osteoinduction (w/o human growth factor)2 recalls0 reports
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T42280Medtronic Sofamor Danek, Inc. · FILLER, BONE VOID, CALCIUM COMPOUND2 recalls0 reports
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T43105Medtronic Sofamor Danek, Inc. · Filler, bone void, osteoinduction (w/o human growth factor)1 recall4 reports
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T43102Medtronic Sofamor Danek, Inc. · FILLER, BONE VOID, CALCIUM COMPOUND1 recall2 reports
- GRAFTON® DBM T44150INTMedtronic Sofamor Danek, Inc. · Filler, bone void, osteoinduction (w/o human growth factor)1 recall1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 93163).
- Z-0161-2024 · Open, ClassifiedGrafton DBM Crunch, Model Numbers: T44115, Demineralized Bone Matrix
- Z-0162-2024 · Open, ClassifiedGrafton DBM (Flex), Model Numbers: T42100, Demineralized Bone Matrix
- Z-0163-2024 · Open, ClassifiedGrafton DMB Matrix, Model Numbers: a) T42200, b) T42210; Demineralized Bone Matrix
- Z-0164-2024 · Open, ClassifiedGrafton DBM Matrix Strips, Model Numbers: a) T42275, b) T42280; Demineralized Bone Matrix
- Z-0165-2024 · Open, ClassifiedGrafton DBM Orthoblend, Model Numbers: a) T44125INT, b) T44135, c) T44145, d) T44150; Demineralized Bone Matrix
- Z-0167-2024 · Open, ClassifiedGrafton DBM Putty, Model Numbers: a) T43102, b) T43102INT, c) T43103INT, d) T43105, e) T43105INT, f) T43110, g) T43110INT; Demineralized Bone Matrix
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page