MAQUET CARDIOHELP recallZ-0420-2024 · Maquet Medical Systems USA · initiated Oct 10, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Maquet Medical Systems USA or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Product removal due to possible blocking or impairment of the CARDIOHELP Emergency Drive. This issue can lead to the patient being exposed to inappropriate low blood flow. Potential associated harms are ischemia and hypoxia.
FDA root cause category: Under Investigation by firm
What the company is doing
On October 10, 2023, the firm issued an Urgent Medical Device Removal letter to all affected consignees. Customers should examine their inventory to determine if they have any affected CARDIOHELP Emergency Drives in their inventory. A representative from Getinge will contact customers to ensure they have received the recall information and have taken the necessary actions. Getinge will arrange the supply of loaner replacements where needed and arrange for the return of affected devices for repair. " You may also request loaner Emergency Drive(s) from Getinge Technical Support at (888) 943-8872 (select option 4,2,1), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone) " If you have affected product that is currently being used in patient treatment, inform Getinge immediately at (888) 943-8872, option 1, option 1. Please ensure that an alternate means of emergency support is available at all times, which may include a nonaffected Emergency Drive (if available) or a fully charged backup CARDIOHELP console. If you have any questions, please contact Getinge Technical Support at (888) 943-8872 (select option 4,2,1), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone). ***Updated 1/8/24*** Additional serial numbers were added to the scope.
Recall details
- Recall number
- Z-0420-2024
- Status
- Open, Classified
- Initiated by the company
- Oct 10, 2023
- Posted by the FDA
- Dec 7, 2023
- Recalling firm
- Maquet Medical Systems USA
- Quantity
- Updated 1/8/24 - 1858 Total distributed in U.S.; 69 Total affected units in U.S.
- Product code
- DTQ
- FDA submissions
- K133598
- Distribution
- US Nationwide - Worldwide Distribution: Domestic distribution to: AL, CA, DC, FL, HI, IA, IL, IN, KY, MD, MI, MN, MO, NE, OH, SC, VA Foreign distribution to: Austria, Australia, Brazil, Canada, Chile, China, Czech Republic, Finland, France, Germany, Hong Kong, India, Ireland, Italy, Japan, Jordan, Mexico, New Zealand, Pakistan, Portugal, Russia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, United Kingdom
Product description
CARDIOHELP Emergency Drive, Part Numbers 701048002 and 701076205. The CARDIOHELP System is a blood oxygenation and carbon dioxide removal system.
Code information (models, lots, serial numbers)
Part No. 701048002: UDI - 04037691643526; Serial Numbers 90425259 to 90425748 (Excluding 90425438, 90425443, 90425568, 90425570, 90425588, 90425716) [***Updated 1/8/24 to include Serial numbers 90421507, 90421980, 90424277. Serial numbers 90425366 and 90423121 are additionally included in the scope due to internal complaints, but these units were corrected prior to the field action initiation and therefore are not included in customer notification***] Part No. 701076205: No UDI; Serial Numbers 90425438, 90425443, 90425568, 90425570, 90425588, 90425716, 90425747, 90425748
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- MAQUET CARDIOHELP CARDIOHELP Emergency Drive (w/o holder)Maquet Cardiopulmonary AG · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS1 recall0 reports
Product lines
Other devices under the same FDA submission
4 devices cleared under K133598 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- MAQUET CARDIOHELP CARDIOHELP-iMaquet Cardiopulmonary AG · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS5 recalls2 deaths64 reports
- MAQUET CARDIOHELP CARDIOHELP SystemMaquet Cardiopulmonary AG · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS4 recalls18 deaths107 reports
- MAQUET CARDIOHELP BS 3/8x3/32 L1.7Maquet Cardiopulmonary AG · Console, Heart-Lung Machine, Cardiopulmonary Bypass2 recalls0 reports
- MAQUET CARDIOHELP CARDIOHELP Base UnitMaquet Cardiopulmonary AG · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS2 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page