DigitalDiagnost C90 recallZ-0276-2024 · Philips North America Llc · initiated Sep 21, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Philips North America Llc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for units suspended on the ceiling to fail and fall.

FDA root cause category: Under Investigation by firm

What the company is doing

The firm distributed recall notifications dated 8/31/2023 via FedEx to affected consignees. Consignees are instructed to immediately stop use of affected devices and contact their local Philips service representative if the Ceiling Suspension becomes unstable, if moving the Ceiling Suspended X-Ray Tube requires a high amount of force, or if moving the Ceiling Suspended X-Ray Tube is noticeably hampered. The recall notice is to be circulated among all device users, so they are aware of this issue and placed as an addendum in the IFU. Consignees are to complete the provided response form and return it to Philips no later than 30 days from notification receipt via email to [email protected]. Philips will contact consignees to schedule a time for a Field Service Engineer to visit customer sites to inspect affected systems and correct the issue. Consignees with any questions are to contact the Customer Care Solutions Center at 1-800-722-9377.

Recall details

Recall number
Z-0276-2024
Status
Open, Classified
Initiated by the company
Sep 21, 2023
Posted by the FDA
Nov 8, 2023
Recalling firm
Philips North America Llc
Quantity
143 units
Product code
JAA
FDA submissions
K202564
Distribution
Nationwide distribution.

Product description

DigitalDiagnost C90 High Performance, used to acquire, process, store, display and export digital radiographic images.

Code information (models, lots, serial numbers)

Show the full code information (6k characters)

Model No. 712034; UDI:(01)00884838090699(21)SN19000013, (01)00884838090699(21)10000203, (01)00884838090699(21)10000008, (01)00884838090699(21)10000030, (01)00884838090699(21)10000043, (01)00884838090699(21)10000037, (01)00884838090699(21)19861128, (01)00884838090699(21)19861129, (01)00884838090699(21)10000064, (01)00884838090699(21)19861143, (01)00884838090699(21)10000024, (01)00884838090699(21)SN10001001, (01)00884838090699(21)10001043, (01)00884838090699(21)10000077, (01)00884838090699(21)10000076, (01)00884838090699(21)10000104, (01)00884838090699(21)10000102, (01)00884838090699(21)10000126 , (01)00884838090699(21) 10000185 , (01)00884838090699(21) 10001084 , (01)00884838090699(21) 22860136 , (01)00884838090699(21) 10001144 , (01)00884838090699(21) 19861160 , (01)00884838090699(21) 10000198 , (01)00884838090699(21) 22860421 , (01)00884838090699(21) SN10001002 , (01)00884838090699(21) 10000212 , (01)00884838090699(21) 19861136 , (01)00884838090699(21) 10001057 , (01)00884838090699(21) 19861165 , (01)00884838090699(21) 10000196 , (01)00884838090699(21) 19861199 , (01)00884838090699(21) 19861197 , (01)00884838090699(21) 19861215 , (01)00884838090699(21) 19861233 , (01)00884838090699(21) 10001106 , (01)00884838090699(21) 10001154 , (01)00884838090699(21) 19861265 , (01)00884838090699(21) 10001104 , (01)00884838090699(21) 10001040 , (01)00884838090699(21) 10001042 , (01)00884838090699(21) 10001019 , (01)00884838090699(21) 10001018 , (01)00884838090699(21) 10001052 , (01)00884838090699(21) 21861498 , (01)00884838090699(21) 10001124 , (01)00884838090699(21) 21861529 , (01)00884838090699(21) 10001114 , (01)00884838090699(21) 10001126 , (01)00884838090699(21) 21861493 , (01)00884838090699(21) 10001109 , (01)00884838090699(21) 10001130 , (01)00884838090699(21) 10001135 , (01)00884838090699(21) 10001203 , (01)00884838090699(21) 10001155 , (01)00884838090699(21) 21861460 , (01)00884838090699(21) 10001145 , (01)00884838090699(21) 10001270 , (01)00884838090699(21) 10001191 , (01)00884838090699(21) 10001178 , (01)00884838090699(21) 10001275 , (01)00884838090699(21) 10001182 , (01)00884838090699(21) 10001169 , (01)00884838090699(21) 10001183 , (01)00884838090699(21) 10001199 , (01)00884838090699(21) 10001192 , (01)00884838090699(21) 21861398 , (01)00884838090699(21) 21861536 , (01)00884838090699(21) 21861503 , (01)00884838090699(21) 21861399 , (01)00884838090699(21) 10001179 , (01)00884838090699(21) 22860037 , (01)00884838090699(21) 10001208 , (01)00884838090699(21) 10001232 , (01)00884838090699(21) 10001215 , (01)00884838090699(21) 22860041 , (01)00884838090699(21) 10001216 , (01)00884838090699(21) 10001264 , (01)00884838090699(21) 22860038 , (01)00884838090699(21) 10001231 , (01)00884838090699(21) 10001236 , (01)00884838090699(21) 10001240 , (01)00884838090699(21) 10001234 , (01)00884838090699(21) 10001251 , (01)00884838090699(21) 10001241 , (01)00884838090699(21) 21861456 , (01)00884838090699(21) 22860442 , (01)00884838090699(21) 21861471 , (01)00884838090699(21) 21861532 , (01)00884838090699(21) 21861418 , (01)00884838090699(21) 21861531 , (01)00884838090699(21) 21861402 , (01)00884838090699(21) 22860064 , (01)00884838090699(21) 22860126 , (01)00884838090699(21) 21861540 , (01)00884838090699(21) 22860063 , (01)00884838090699(21) 22860117 , (01)00884838090699(21) 21861497 , (01)00884838090699(21) 21861496 , (01)00884838090699(21) 21861542 , (01)00884838090699(21) 21861525 , (01)00884838090699(21) 22860052 , (01)00884838090699(21) 22860050 , (01)00884838090699(21) 22860051 , (01)00884838090699(21) 22860075 , (01)00884838090699(21) 22860076 , (01)00884838090699(21) 22860072 , (01)00884838090699(21) 22860154 , (01)00884838090699(21) 22860104 , (01)00884838090699(21) 22860342 , (01)00884838090699(21) 22860060 , (01)00884838090699(21) 22860501 , (01)00884838090699(21) 22860077 , (01)00884838090699(21) 22860097 , (01)00884838090699(21) 22860326 , (01)00884838090699(21) 22860305 , (01)00884838090699(21) 22860155 , (01)00884838090699(21) 22860161 , (01)00884838090699(21) 22860370 , (01)00884838090699(21) 22860374 , (01)00884838090699(21) 22860469 , (01)00884838090699(21) 22860432 , (01)00884838090699(21) 22860436 , (01)00884838090699(21) 22860407 , (01)00884838090699(21) 22860519 , (01)00884838090699(21) 22860531 , (01)00884838090699(21) 22860477 , (01)00884838090699(21) 23860432 , (01)00884838090699(21) 22860634 , (01)00884838090699(21) 22860625 , (01)00884838090699(21) 22860615 , (01)00884838090699(21) 10001125 , (01)00884838090699(21) 22860580 , (01)00884838090699(21) 10001261 , (01)00884838090699(21) 10001128 , (01)00884838090699(21) 23860317 , (01)00884838090699(21) 10001180 , (01)00884838090699(21) 23860376 , (01)00884838090699(21) 22860360 , (01)00884838090699(21) 23860389 , (01)00884838090699(21) 23860518 , (01)00884838090699(21) 10000177 , (01)00884838090699(21) 22860451; Serial No. SN19000013, 10000203, 10000008, 10000030, 10000043, 10000037, 19861128, 19861129, 10000064, 19861143, 10000024, SN10001001, 10001043, 10000077, 10000076, 10000104, 10000102, 10000126, 10000185, 10001084, 22860136, 10001144, 19861160, 10000198, 22860421, SN10001002, 10000212, 19861136, 10001057, 19861165, 10000196, 19861199, 19861197, 19861215, 19861233, 10001106, 10001154, 19861265, 10001104, 10001040, 10001042, 10001019, 10001018, 10001052, 21861498, 10001124, 21861529, 10001114, 10001126, 21861493, 10001109, 10001130, 10001135, 10001203, 10001155, 21861460, 10001145, 10001270, 10001191, 10001178, 10001275, 10001182, 10001169, 10001183, 10001199, 10001192, 21861398, 21861536, 21861503, 21861399, 10001179, 22860037, 10001208, 10001232, 10001215, 22860041, 10001216, 10001264, 22860038, 10001231, 10001236, 10001240, 10001234, 10001251, 10001241, 21861456, 22860442, 21861471, 21861532, 21861418, 21861531, 21861402, 22860064, 22860126, 21861540, 22860063, 22860117, 21861497, 21861496, 21861542, 21861525, 22860052, 22860050, 22860051, 22860075, 22860076, 22860072, 22860154, 22860104, 22860342, 22860060, 22860501, 22860077, 22860097, 22860326, 22860305, 22860155, 22860161, 22860370, 22860374, 22860469, 22860432, 22860436, 22860407, 22860519, 22860531, 22860477, 23860432, 22860634, 22860625, 22860615, 10001125, 22860580, 10001261, 10001128, 23860317, 10001180, 23860376, 22860360, 23860389, 23860518, 10000177, 22860451.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 93200).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page