Skip to content

Philips Affiniti 70, Diagnostic Ultrasound System Model #… recallZ-0284-2024 · Philips Ultrasound, Inc. · initiated Oct 9, 2023

Completed

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Philips Ultrasound, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Their is the potential that diagnostic ultrasound systems operating with software version 10.0 may present a delayed image instead of a real time image.

FDA root cause category: Software design

What the company is doing

On October 9, 2023, Philips issued a Urgent Medical Device Correction notification to affected consignees. Philips asked consignees to take the following actions: 1. Identify if your firm has an affected product. 2. In the 2D tab in the system touchscreen, disable the AutoSCAN feature every time a transducer or a Transducer preset is selected to prevent the system from potentially experiencing image lag issue. The AutoSCAN feature is disabled when the button is not highlighted in amber. 3. Please complete and return the attached response form to Philips promptly and no later than 15 days from receipt. Completing this form confirms receipt of the Urgent Medical Device Correction Letter, understanding of the issue, and required actions to be taken. 4. Circulate this Urgent Medical Device Correction Letter to all users of this device so that they are aware of the issue. Please retain this letter with your system(s) and ensure the letter is in a place likely to be seen/viewed. 5. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, or by regular mail, or by fax.

Recall details

Recall number
Z-0284-2024
Status
Completed
Initiated by the company
Oct 9, 2023
Posted by the FDA
Nov 9, 2023
Recalling firm
Philips Ultrasound, Inc.
Quantity
7 systems
Product code
IYN
FDA submissions
K231190
Distribution
US: AZ, CA, CT, FL, GA, ID, IL, KY, MD, MT, NE, NV, OH, TN, TX, UT, VA OUS: Canada

Product description

Philips Affiniti 70, Diagnostic Ultrasound System Model # 795210

Code information (models, lots, serial numbers)

Following serial numbers operating with software version 10.0. Model 795210/UDI: (01)00884838106611/Serial Numbers: US523F1930 US918F1158 US918F1159 US723F1377 US523F1496 US523F1497 US217F0292

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 93218).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page