Philips Affiniti 70, Diagnostic Ultrasound System Model #… recallZ-0284-2024 · Philips Ultrasound, Inc. · initiated Oct 9, 2023
Completed
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips Ultrasound, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Their is the potential that diagnostic ultrasound systems operating with software version 10.0 may present a delayed image instead of a real time image.
FDA root cause category: Software design
What the company is doing
On October 9, 2023, Philips issued a Urgent Medical Device Correction notification to affected consignees. Philips asked consignees to take the following actions: 1. Identify if your firm has an affected product. 2. In the 2D tab in the system touchscreen, disable the AutoSCAN feature every time a transducer or a Transducer preset is selected to prevent the system from potentially experiencing image lag issue. The AutoSCAN feature is disabled when the button is not highlighted in amber. 3. Please complete and return the attached response form to Philips promptly and no later than 15 days from receipt. Completing this form confirms receipt of the Urgent Medical Device Correction Letter, understanding of the issue, and required actions to be taken. 4. Circulate this Urgent Medical Device Correction Letter to all users of this device so that they are aware of the issue. Please retain this letter with your system(s) and ensure the letter is in a place likely to be seen/viewed. 5. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, or by regular mail, or by fax.
Recall details
- Recall number
- Z-0284-2024
- Status
- Completed
- Initiated by the company
- Oct 9, 2023
- Posted by the FDA
- Nov 9, 2023
- Recalling firm
- Philips Ultrasound, Inc.
- Quantity
- 7 systems
- Product code
- IYN
- FDA submissions
- K231190
- Distribution
- US: AZ, CA, CT, FL, GA, ID, IL, KY, MD, MT, NE, NV, OH, TN, TX, UT, VA OUS: Canada
Product description
Philips Affiniti 70, Diagnostic Ultrasound System Model # 795210
Code information (models, lots, serial numbers)
Following serial numbers operating with software version 10.0. Model 795210/UDI: (01)00884838106611/Serial Numbers: US523F1930 US918F1158 US918F1159 US723F1377 US523F1496 US523F1497 US217F0292
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Affiniti 70 Ultrasound System 795210Philips Ultrasound LLC · System, Image Processing, Radiological1 recall3 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 93218).
- Z-0283-2024 · CompletedPhilips EPIQ Diagnostic Ultrasound System Models: EPIQ 7G(model # 795200), EPIQ 7C (Model # 795201), EPIQ 7W (Model # 795202), EPIQ 5G (model # 795204), EPIQ 5W (model # 795206), …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page