Affiniti 70 Ultrasound System 795210Recalls & reported problems
Philips Ultrasound LLC · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00884838106611), received between May 12, 2025 and Jul 2, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Battery Problem” (1 report).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Oct 9, 2023Class IICompleted
Their is the potential that diagnostic ultrasound systems operating with software version 10.0 may present a delayed image instead of a real time image.
Z-0284-2024 · Root cause: Software design · Recall details · FDA record ↗
Most reported problems
Device problems
- Battery Problem1
- Crack1
- Incorrect Measurement1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 2, 2025Crack; Crack
IT WAS REPORTED THAT THE INTRACAVITY C10-3V TRANSDUCER SEAL WAS CRACKED BETWEEN THE TWO HALVES OF THE TRANSDUCER, RESULTING IN CLEANING FLUID ENTERING AND CAUSING A GREEN SUBSTANCE AT THE BOTTOM OF THE TRANSDUCER CONNECTOR. THE TRANSDUCER WAS ASSOCIATED WITH THE AFFINITI 70 ULTRASOUND SYSTEM AT THE CUSTOMER¿S SITE. THE ISSUE WAS DISCOVERED OUTSIDE OF CLINICAL USE, DURING CLEANING AND INSPECTION, AND THE TRANSDUCER WAS REMOVED FROM SERVICE. NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. THE PHILIPS SERVICE ENGINEER REPLACED THE TRANSDUCER TO ADDRESS THE CUSTOMER¿S IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22391635 ↗MalfunctionJun 18, 2025Battery Problem; Battery Problem
IT WAS REPORTED THAT AN AMETEK POWERVAR 3KVA UNINTERRUPTIBLE POWER SUPPLY (UPS) HAD OVERHEATED AND EMITTED AN ODOR. THE UPS HAD SWOLLEN INTERNAL BATTERIES WHICH SHOWED SIGNS OF LEAKAGE. THE UPS WAS USED WITH THE AFFINITI 70 ULTRASOUND SYSTEM AT THE CUSTOMER¿S SITE. THE FAILURE OCCURRED OUTSIDE OF CLINICAL USE AND NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22285526 ↗MalfunctionMay 12, 2025Incorrect Measurement; Incorrect Measurement
A CUSTOMER REPORTED THAT WHEN PERFORMING EXAMS WITH THE C6-2 TRANSDUCER ON THEIR AFFINITI 70 ULTRASOUND SYSTEM AND GATHERING THE DOPPLER MEASUREMENT OF THE FETAL CALCULUS BETWEEN ONE WAVEFORM AND ANOTHER, THERE WAS AN EMPTY SPACE THAT THE DOCTOR CONSIDERED TO BE ZERO. THE CUSTOMER ALLEGES THIS HAS THE POTENTIAL TO LEAD TO A MISDIAGNOSIS. THE PATIENT DID NOT REQUIRE ANY LIFESAVING TREATMENT OR THERAPY OR HAVE A DECLINE IN THEIR STATE OF HEALTH. THE EXAM WAS COMPLETED USING ANOTHER SYSTEM. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22001989 ↗
Other System, Image Processing, Radiological devices
- Interventional Workspot Interventional Workspot - 1Philips Medical Systems Nederland B.V. · System, Image Processing, Radiological13 recalls24 reports
- Synapse Radiology PACS Synapse Radiology PACS Software LicenseFujifilm Healthcare Americas Corporation · System, image processing, radiological9 recalls38 reports
- CENTRICITY UNIVERSAL VIEWER ZFP CLIENT 6.0Ge Healthcare · System, image processing, radiological8 recalls0 reports
- CENTRICITY UNIVERSAL VIEWER 6.0Ge Healthcare · System, image processing, radiological8 recalls0 reports
- CENTRICITY UNIVERSAL VIEWER 7.0Ge Healthcare · System, image processing, radiological6 recalls0 reports
- IntelliSpace Cardiovascular 8Philips Medical Systems Nederland B.V. · System, Image Processing, Radiological2 recalls132 reports
- CA1000 2.0 SP 14Ge Healthcare · System, image processing, radiological4 recalls0 reports
- VMA 2Ortho Kinematics, Inc. · System, Image Processing, Radiological4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page