Skip to content

RSP Humeral Socket Insert Size 36mm Semi-Constrained +4 mm… recallZ-0282-2024 · Encore Medical, LP · initiated Sep 25, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Encore Medical, LP or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Two different humeral socket Insert devices were swapped during packaging, which could result in surgeons not having the correct device for surgery, which could cause a delay in surgery.

FDA root cause category: Labeling Change Control

What the company is doing

On 09/25/23, recall notices were emailed to Agents who were asked to do the following: 1) Pass the notice to all those within your organization or to any organization where the potentially affected product has been transferred. 2) Return all affected devices using the RMA number. 3) Complete and return the response form. Questions can be directed to [email protected]

Recall details

Recall number
Z-0282-2024
Status
Open, Classified
Initiated by the company
Sep 25, 2023
Posted by the FDA
Nov 8, 2023
Recalling firm
Encore Medical, LP
Quantity
20
Product code
PHX
FDA submissions
K051075
Distribution
US: MN, TX, IN, OH, SC, RI, ME, NJ, NY, KS, LA, FL, CA

Product description

RSP Humeral Socket Insert Size 36mm Semi-Constrained +4 mm, REF: 509-01-436, orthopedic shoulder implant component.

Code information (models, lots, serial numbers)

UDI-DI: 00888912144667, Lot: 390P1157

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

27 devices cleared under K051075 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 93221).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page