AC Power Adapter for use with the LIFEPAK 15… recallZ-0567-2024 · Physio-Control, Inc. · initiated Nov 6, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Physio-Control, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
AC Power Adapter (ACPA) may not charge the monitor/defibrillator batteries and may not allow the monitor/defibrillator to power on while plugged into the ACPA.
FDA root cause category: Component change control
What the company is doing
On 11/06/23, Recall and Field Safety Alert notices were sent to customers. The recall notices asked customers to do the following: 1) Complete and return business reply form by email to [email protected] 2) Firm recommends continued use. Follow the OI recommendations for daily inspection and testing according to the checklist provided in the Operating Instructions. a. The checklist includes disconnecting the defibrillator from the power adapter, turning the device on, and confirming that two batteries are installed and charged. Then, the defibrillator should be reconnected to the power adapter and users should confirm that the battery charging LED on the defibrillator is illuminated or flashing. b. If an ACPA is experiencing this failure mode, this will present to the user as the Aux Power Indicator on the LP15 flashing instead of solidly illuminating and the green LED strip on the ACPA flashing instead of solidly illuminating: 3) Please notify impacted customers that were shipped or may have been shipped this affected product by sharing this recall notification letter; Email Stryker at [email protected] with a list of customers who received/may have received this product and we can work with you on notifications to impacted customers. If you have any questions or feel that your ACPA may be experiencing issues, please contact Customer Service at 1 800 787 9537, option 2, from 8:00 AM to 7:00 PM (Eastern Time), Monday - Friday or by email at [email protected]
Recall details
- Recall number
- Z-0567-2024
- Status
- Open, Classified
- Initiated by the company
- Nov 6, 2023
- Posted by the FDA
- Dec 15, 2023
- Recalling firm
- Physio-Control, Inc.
- Quantity
- 18,732
- Product code
- MKJ
- FDA submissions
- P160026
- Distribution
- US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY.
Product description
AC Power Adapter for use with the LIFEPAK 15 Monitor/Defibrillator, REF: 11140-000098
Code information (models, lots, serial numbers)
UDI-DI: 00883873940204, Adapters manufactured between 11/3/2021 - 4/24/2023
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- NA 11140Physio-Control, Inc. · Auxiliary power supply (AC or DC) for low-energy dc-defibrillator1 recall1 report
Other devices under the same FDA submission
27 devices cleared under P160026 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors 99507Physio-Control, Inc. · Automated external defibrillators (non-wearable)1 death228 reports
- LIFEPAK 1000 defibrillator 99425Physio-Control, Inc. · Automated external defibrillators (non-wearable)1 death188 reports
- LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors 99402Physio-Control, Inc. · Automated external defibrillators (non-wearable)4 deaths173 reports
- LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors 99507Physio-Control, Inc. · Automated external defibrillators (non-wearable)4 deaths129 reports
- LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors 99507Physio-Control, Inc. · Automated external defibrillators (non-wearable)124 reports
- LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors 99507Physio-Control, Inc. · Automated external defibrillators (non-wearable)5 deaths101 reports
- LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors 99507Physio-Control, Inc. · Automated external defibrillators (non-wearable)91 reports
- LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors 99507Physio-Control, Inc. · Automated external defibrillators (non-wearable)3 deaths78 reports
- LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors 99402Physio-Control, Inc. · Automated external defibrillators (non-wearable)77 reports
- LIFEPAK 1000 defibrillator 99425Physio-Control, Inc. · Automated external defibrillators (non-wearable)76 reports
- LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors 99507Physio-Control, Inc. · Automated external defibrillators (non-wearable)1 recall3 reports
- LIFEPAK 15 monitor/defibrillator 97577Physio-Control, Inc. · Automated external defibrillators (non-wearable)1 recall2 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page