LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors 99507 (REF 99507-000002)Recalls & reported problems

Physio-Control, Inc. · LP20E - CONFIGURED, AHA2005 PACING, LI-ION, EN · Class 3

All models of LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors (23)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Physio-Control LIFEPAK 20 defibrillator has 228 adverse event reports, including 227 malfunctions, no injuries and 1 death. The reports mostly describe the unit not powering on or failing to finish starting up, which could stop it from delivering a shock if needed. Many reports note no patient involvement. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
228
Deaths reported
1
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 228 adverse event reports that list this exact device (DI 00883873813768), received between Dec 20, 2018 and Jun 9, 2026 — classified as 1 death, 227 malfunctions.

The most frequently reported specific device problem is “Failure to Power Up” (88 reports).

17 other recalls cover devices cleared under the same FDA submission (P160026); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 6 reportsDec 2023: 1 reportsJan 2024: 2 reportsFeb 2024: 6 reportsMar 2024: 4 reportsApr 2024: 2 reportsMay 2024: 4 reportsJun 2024: 4 reportsJul 2024: 4 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Power Up88
  • Electrical /Electronic Property Problem64
  • Failure to Deliver Shock/Stimulation13
  • Unexpected Shutdown13
  • Energy Output Problem9
  • Inappropriate/Inadequate Shock/Stimulation9
  • Display or Visual Feedback Problem8
  • Device Sensing Problem6

Patient problems

  • No Clinical Signs, Symptoms or Conditions179
  • No Patient Involvement50
  • No Known Impact Or Consequence To Patient11
  • Cardiac Arrest1

Recent adverse event reports

The 25 most recent of 228 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 9, 2026Failure to Power Up

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WAS NOT POWERING ON. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25440725 ↗
  • MalfunctionApr 30, 2026Failure to Power Up

    A CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT POWER ON. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25041197 ↗
  • MalfunctionApr 28, 2026Failure to Power Up

    A CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT POWER ON. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25017191 ↗
  • MalfunctionSep 17, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WAS UNABLE TO COMPLETE THE BOOT UP CYCLE. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF IT WAS NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23070287 ↗
  • MalfunctionJun 23, 2025Failure to Power Up; Failure to Power Up

    A CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE DOES NOT POWER ON. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY, IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22306250 ↗
  • MalfunctionMar 5, 2025Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WILL NOT POWER ON. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21530204 ↗
  • MalfunctionSep 17, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE DISPLAYED A WHITE SCREEN UPON DEVICE BOOT UP AND WOULD NOT COMPLETE A BOOT CYCLE. IN THIS STATE, THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20234684 ↗
  • MalfunctionAug 9, 2024Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT POWER ON WITH AC POWER. IN THIS STATE, THE DEVICE MAY NOT BE ABLE TO PROVIDE DEFIBRILLATION THERAPY, IF IT WERE NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19951877 ↗
  • MalfunctionJul 25, 2024Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE FAILED TO POWER UP. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19824770 ↗
  • MalfunctionJul 9, 2024Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WILL NOT POWER ON USING AC POWER. AS A RESULT, DEFIBRILLATION THERAPY MAY BE DELAYED OR UNAVAILABLE, IF NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19706063 ↗
  • MalfunctionJul 9, 2024Power Problem; Power Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE AC POWER WAS INOPERABLE. IN THIS STATE, DEFIBRILLATION THERAPY MAY BE DELAYED OR PREVENTED FROM BEING DELIVERED TO THE PATIENT IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19705542 ↗
  • MalfunctionJul 9, 2024Power Problem; Power Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE AC POWER WAS INOPERABLE. IN THIS STATE, DEFIBRILLATION THERAPY MAY BE DELAYED OR PREVENTED FROM BEING DELIVERED TO THE PATIENT IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19705391 ↗
  • MalfunctionJun 27, 2024Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT POWER ON. IN THIS STATE THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19629798 ↗
  • MalfunctionJun 13, 2024Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE IS NOT POWERING ON. IN THIS STATE THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19527989 ↗
  • MalfunctionJun 4, 2024Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE UNEXPECTEDLY LOST POWER. IN THIS STATE, THE DEVICE MAY NOT BE ABLE TO PROVIDE DEFIBRILLATION THERAPY, IF IT WERE NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19461168 ↗
  • MalfunctionJun 3, 2024Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WAS NOT POWERING ON. IN THIS STATE THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19447423 ↗
  • MalfunctionMay 30, 2024Device Sensing Problem; Device Sensing Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE DID NOT DELIVER SHOCK ENERGY WHEN NEEDED. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19424068 ↗
  • MalfunctionMay 28, 2024Device Sensing Problem; Device Sensing Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE DID NOT DETECT THE PATIENT WHEN CONNECTED WITH THE ELECTRODES. IN THIS STATE THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19411805 ↗
  • MalfunctionMay 8, 2024Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED STRYKER TO REPORT THEIR DEVICE WOULD NOT POWER ON. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19270847 ↗
  • MalfunctionMay 7, 2024Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT POWER ON. IN THIS STATE, THE DEVICE MAY NOT BE ABLE TO PROVIDE DEFIBRILLATION THERAPY, IF IT WERE NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19260501 ↗
  • MalfunctionApr 11, 2024Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WILL NOT POWER ON. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19087238 ↗
  • MalfunctionApr 9, 2024Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT POWER ON. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED TO THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19070096 ↗
  • MalfunctionMar 28, 2024Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT POWER ON. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED TO THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18996415 ↗
  • MalfunctionMar 11, 2024Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT POWER ON. IN THIS STATE, THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18873075 ↗
  • MalfunctionMar 6, 2024Energy Output Problem; Energy Output Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT THEIR DEVICE'S DEFIBRILLATION TOLERANCE WAS OUT OF RANGE DURING TESTING. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18847290 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P160026. One submission often covers many models, so check the product and lot details in each record.

  • Z-2065-2026 · Apr 13, 2026 · Open, Classified
    LIFEPAK 15, REF: 99577-000025, 99577-000047, 99577-001256, 99577-001217, 99577-001241, 99577-001255, 99577-001219, 99577-001957, 99577-001935, 99577-001964, 99577-001955, 99577-001
  • Z-2497-2025 · Jul 22, 2025 · Open, Classified
    LIFEPAK 15 Part / Catalog Numbers / UDI-DI code: V15-7-000022 / 99577-001957 / 883873871690 V15-2-000051 / 99577-000046 / 883873988022 V15-2-000099 / 99577-000093 / 883873979020 V
  • Z-2498-2025 · Jul 22, 2025 · Open, Classified
    LIFEPAK 1000 Part / Catalog Numbers / UDI-DI code: 320371500023 / 99425-000025 / 885074698953 320371500229 / 99425-000025 / 885074698953 331706400001 / 99425-000025 / 885074698953
  • Z-1232-2025 · Feb 3, 2025 · Open, Classified
    LIFEPAK 20e REF 99507-000001 The AED mode is to be used only on patients in cardiopulmonary arrest.
  • Z-1184-2025 · Jan 21, 2025 · Open, Classified
    LIFEPAK 15 V4/Masimo Rainbow Sensors: Masimo RD rainbow Adt 8 SpCO, SpO2, and SpMet, Adult Adhesive Sensors REF 11996-000515 Masimo LNCS-II rainbow DCI 8 SpCO, Adult Reusable Sen
  • Z-0230-2025 · Aug 30, 2024 · Open, Classified
    LIFEPAK 15 (LP15) Model/Catalog Numbers: o 97577-000101 o 99577-000046 o 99577-000055 o 99577-000061 o 99577-000076 o 99577-001250 o 99577-001255 o 99577-001256 o 99577-001368 o 99
  • Z-0567-2024 · Nov 6, 2023 · Open, Classified
    AC Power Adapter for use with the LIFEPAK 15 Monitor/Defibrillator, REF: 11140-000098
  • Z-0305-2024 · Sep 29, 2023 · Open, Classified
    LIFEPAK 15, REF: 99577-001960, Part Number: V15-7-000026
  • Z-0300-2024 · Sep 29, 2023 · Open, Classified
    LIFEPAK 15, REF: 99577-001373, Part Number V15-7-000002
  • Z-0303-2024 · Sep 29, 2023 · Open, Classified
    LIFEPAK 15, REF: 99577-001957, Part Number V15-5-000586 and V15-7-000022
  • Z-0296-2024 · Sep 29, 2023 · Open, Classified
    LIFEPAK 15, REF: 99577-000046, Part Number V15-2-000051
  • Z-0297-2024 · Sep 29, 2023 · Open, Classified
    LIFEPAK 15, REF: 99577-000055, Part Number V15-2-000061
  • Z-0301-2024 · Sep 29, 2023 · Open, Classified
    LIFEPAK 15, REF: 99577-001955, Part Number V15-5-000584 and V15-7-000020
  • Z-0302-2024 · Sep 29, 2023 · Open, Classified
    LIFEPAK 15, REF: 99577-001956, Part Number V15-7-000021
  • Z-0298-2024 · Sep 29, 2023 · Open, Classified
    LIFEPAK 15, REF: 99577-000061, Part Number V15-2-000067

…and 2 more.

Other Automated external defibrillators (non-wearable) devices

All Automated external defibrillators (non-wearable) devices →