Cooper Surgical H/S ELLIPTOSPHERE CATH 5 FR Set-For… recallZ-0503-2024 · CooperSurgical, Inc. · initiated Mar 31, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from CooperSurgical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Incorrect bonding material used for assembly of the catheter, may fail, resulting in detachment of component parts, broken components, or leakage. This breakage may result in delay of procedure
FDA root cause category: Component design/selection
What the company is doing
CooperSurgical issued Urgent Medical Device Safety Notice dated March 22, 2023 via UPS Certified Delivery Mail. Letter states reason for recall, health risk and action to take:Our records indicate that you may have purchased the affected Product from CooperSurgical. Please inspect stock and complete the attached Acknowledgement and Receipt Form. Once completed please return the form to CooperSurgical to acknowledge receipt of the notice. If you do not have affected Product in inventory, please use the same enclosed Form to indicate that and return it to CooperSurgical so that we may document receipt of this Recall Notice. A corrective action has been initiated to ensure this failure does not reoccur. We sincerely apologize for any inconvenience caused by this notice. CooperSurgical is committed to high quality, safe and effective products. Please feel free to reach us at 203-601-5200 ext. 3300 or via email at [email protected].
Recall details
- Recall number
- Z-0503-2024
- Status
- Open, Classified
- Initiated by the company
- Mar 31, 2022
- Posted by the FDA
- Dec 6, 2023
- Recalling firm
- CooperSurgical, Inc.
- Quantity
- 169 units
- Product code
- LKF
- FDA submissions
- K013972
- Distribution
- Nationwide Foreign: BELGIUM CANADA MALAYSIA NETHERLANDS SPAIN SWEDEN SWITZERLAND
Product description
Cooper Surgical H/S ELLIPTOSPHERE CATH 5 FR Set-For administering contrast media during Hysterosalpingography or Hysterosonography procedures to detect uterine pathology such as polyps, fibroids, adhesions or endometrial thickening, and/or patency of fallopian tubes Model: 61-4005
Code information (models, lots, serial numbers)
UDI-DI:60888937016446 S/N: 313121, 313568
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- H/S Elliptosphere™ Catheter • 5 French 61-4005Coopersurgical, Inc. · Cannula, Manipulator/Injector, Uterine1 recall0 reports
Other devices under the same FDA submission
1 device cleared under K013972 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- H/S Elliptosphere™ Catheter with Integrated Stylet 61-4005SCoopersurgical, Inc. · Cannula, Manipulator/Injector, Uterine1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 93239).
- Z-0502-2024 · Open, ClassifiedCooper Surgical H/S ELLIPTOSPHERE CATH 5 Fr -For administering contrast media during Hysterosalpingography or Hysterosonography procedures to detect uterine pathology such as polyp…
- Z-0504-2024 · Open, ClassifiedCooperSurgical H/S CATHETER SET FR 5 with Integated Stylet Model: 61-4005S For administering contrast media during Hysterosalpingography or Hysterosonography procedures to detect u…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page