Skip to content

CooperSurgical H/S CATHETER SET FR 5 with Integated Stylet recallZ-0504-2024 · CooperSurgical, Inc. · initiated Mar 31, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from CooperSurgical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Incorrect bonding material used for assembly of the catheter, may fail, resulting in detachment of component parts, broken components, or leakage. This breakage may result in delay of procedure

FDA root cause category: Component design/selection

What the company is doing

CooperSurgical issued Urgent Medical Device Safety Notice dated March 22, 2023 via UPS Certified Delivery Mail. Letter states reason for recall, health risk and action to take:Our records indicate that you may have purchased the affected Product from CooperSurgical. Please inspect stock and complete the attached Acknowledgement and Receipt Form. Once completed please return the form to CooperSurgical to acknowledge receipt of the notice. If you do not have affected Product in inventory, please use the same enclosed Form to indicate that and return it to CooperSurgical so that we may document receipt of this Recall Notice. A corrective action has been initiated to ensure this failure does not reoccur. We sincerely apologize for any inconvenience caused by this notice. CooperSurgical is committed to high quality, safe and effective products. Please feel free to reach us at 203-601-5200 ext. 3300 or via email at [email protected].

Recall details

Recall number
Z-0504-2024
Status
Open, Classified
Initiated by the company
Mar 31, 2022
Posted by the FDA
Dec 6, 2023
Recalling firm
CooperSurgical, Inc.
Quantity
90 units
Product code
LKF
FDA submissions
K013972
Distribution
Nationwide Foreign: BELGIUM CANADA MALAYSIA NETHERLANDS SPAIN SWEDEN SWITZERLAND

Product description

CooperSurgical H/S CATHETER SET FR 5 with Integated Stylet Model: 61-4005S For administering contrast media during Hysterosalpingography or Hysterosonography procedures to detect uterine pathology such as polyps, fibroids, adhesions or endometrial thickening, and/or patency of fallopian tubes

Code information (models, lots, serial numbers)

UDI-DI: 60888937016453 S/N: 309967

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

1 device cleared under K013972 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.