Medtronic Mo.Ma Ultra Proximal Cerebral Protection Device… recallZ-0321-2024 · Medtronic Inc · initiated Oct 12, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medtronic Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The Mo.Ma Ultra device manifold is labelled with 2 stickers, an inferior and superior sticker, one on each side of the manifold. These stickers label the common carotid artery (CCA) proximal balloon and external carotid artery (ECA) distal balloon inflation and deflation ports. Both sides of the complaint units manifold were labelled with the superior label leading to an incorrect identification of the proximal and distal inflation /deflation ports when viewing the devices from the inferior side.

FDA root cause category: Process control

What the company is doing

Medtronic issued an URGENT: MEDICAL DEVICE RECALL NOTIFICATION on 10/04/2023 via UPS. The notice explained the issue, potential risk, and requested the return of the device. For questions, consignees were directed to contact their Medtronic Field Sales Representative.

Recall details

Recall number
Z-0321-2024
Status
Open, Classified
Initiated by the company
Oct 12, 2023
Posted by the FDA
Nov 17, 2023
Recalling firm
Medtronic Inc
Quantity
173 units
Product code
NTE
FDA submissions
K092177
Distribution
Worldwide distribution - US Nationwide and the countries of Algeria, Armenia, France, Germany, Iraq, Italy, Japan, Netherlands, Poland, Romania, Spain, Sweden, Switzerland, United Kingdom.

Product description

Medtronic Mo.Ma Ultra Proximal Cerebral Protection Device, Model Number MOM0130069X6

Code information (models, lots, serial numbers)

UDI/DI 08033477090702:

Other devices under the same FDA submission

3 devices cleared under K092177 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 93258).