Elekta Medical Linear Accelerator recallZ-0443-2024 · Elekta, Inc. · initiated Oct 25, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Elekta, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of uncontrolled extension.

FDA root cause category: Under Investigation by firm

What the company is doing

Elekta released Field Safety Notice to its consignees on the 10/25/2023 via email. The notice explained the issue, potential risk, and provides information for the safe use of the instrument until Elekta Field Service Representatives can perform the correction.

Recall details

Recall number
Z-0443-2024
Status
Open, Classified
Initiated by the company
Oct 25, 2023
Posted by the FDA
Dec 1, 2023
Recalling firm
Elekta, Inc.
Quantity
1019 units
Product code
IYE
FDA submissions
K210500
Distribution
worldwide

Product description

ELEKTA LIMITED Elekta Synergy, REF MRT 8501, MRT 8511; Medical Charged-Particle Radiation Therapy System Accelerator, Linear

Code information (models, lots, serial numbers)

Show the full code information (8k characters)

UDI/DI 05060191071505, Serial Numbers: 154549, 151917, 151963, 152072, 152081, 152131, 152382, 151838, 152318, 152807, 151951, 152416, 153147, 153148, 153963, 151821, 151845, 151940, 152160, 152162, 152164, 152389, 152393, 152394, 152634, 152635, 152636, 152664, 152665, 152725, 152992, 151850, 151920, 152082, 152089, 152111, 152546, 152548, 152299, 152403, 152503, 152676, 152950, 151857, 151690, 151744, 151745, 151766, 151831, 151859, 151900, 151950, 152046, 152137, 152194, 152202, 152216, 152258, 152276, 152376, 152390, 152391, 152399, 152414, 152430, 152431, 152513, 152521, 152557, 152576, 152645, 152647, 152712, 152719, 152740, 152743, 153093, 153115, 153126, 153258, 153299, 153475, 153554, 153804, 153911, 154415, 151852, 151894, 151971, 152185, 152241, 152297, 152422, 152429, 152584, 152616, 152628, 152631, 152681, 152698, 152746, 152747, 152770, 152804, 152811, 152812, 152813, 152985, 153256, 153286, 153546, 152014, 152208, 153952, 152883, 152714, 154342, 152724, 151366, 152047, 152074, 152075, 152133, 152197, 152290, 152291, 152395, 152462, 152465, 152646, 152682, 152945, 153476, 153620, 153640, 154071, 151815, 152157, 152602, 152726, 152729, 152750, 152763, 152806, 152987, 151771, 151905, 154255, 152482, 152580, 152582, 152583, 152586, 152639, 152735, 152738, 152785, 152797, 152850, 153124, 153290, 152119, 152603, 152942, 152744, 153122, 152452, 151972, 152016, 152038, 152052, 152251, 152267, 152287, 152523, 152572, 152578, 152605, 152606, 152611, 152826, 152880, 152924, 152925, 152931, 153162, 153259, 153292, 153293, 153520, 153669, 153365, 154483, 151973, 152545, 152060, 151895, 152102, 152190, 151775, 151830, 151843, 151871, 151893, 151985, 151987, 152655, 152656, 152657, 152658, 152720, 153757, 153758, 153763, 153774, 153777, 151492, 152201, 152653, 151033, 151176, 151398, 151416, 151517, 151640, 151683, 151700, 151710, 151736, 151759, 151791, 151810, 151860, 151885, 151892, 151902, 151954, 151956, 151989, 152043, 152098, 152116, 152123, 152214, 152232, 152331, 152475, 152530, 152559, 152703, 152767, 152815, 153045, 153086, 153867, 154515; UDI/DI: 05060191071550, Serial Numbers: 152008, 153854, 153863, 153954, 153995, 109002, 152007, 152369, 152379, 152619, 152695, 152696, 152896, 152897, 152905, 152926, 153007, 153012, 153014, 153024, 153152, 153276, 153033, 153220, 151836, 153449, 154138, 151868, 152188, 152533, 152093, 152363, 152472, 154042, 152005, 152056, 152259, 152272, 152354, 152500, 152689, 152796, 153022, 153943, 154263, 154384, 153728, 151928, 152124, 152166, 152242, 152244, 152651, 151673, 153038, 151403, 151624, 151663, 151948, 151970, 151979, 152010, 152020, 152024, 152090, 152095, 152145, 152178, 152203, 152253, 152268, 152296, 152334, 152385, 152408, 152411, 152415, 152449, 152453, 152474, 152487, 152516, 152529, 152532, 152560, 152568, 152588, 152594, 152610, 152666, 152671, 152685, 152693, 152704, 152773, 152774, 152775, 152776, 152780, 152810, 152814, 152820, 152823, 152837, 152856, 152857, 152872, 152876, 152877, 152884, 152886, 152910, 152911, 152914, 152932, 152941, 152966, 152986, 153026, 153027, 153063, 153073, 153078, 153079, 153083, 153105, 153131, 153132, 153139, 153146, 153158, 153191, 153196, 153207, 153261, 153263, 153264, 153271, 153272, 153273, 153275, 153280, 153305, 153309, 153317, 153390, 153391, 153392, 153393, 153394, 153395, 153405, 153419, 153420, 153421, 153422, 153431, 153458, 153469, 153477, 153480, 153484, 153502, 153515, 153519, 153541, 153555, 153565, 153588, 153590, 153591, 153592, 153613, 153630, 153643, 153648, 153653, 153659, 153663, 153665, 153700, 153707, 153712, 153717, 153729, 153734, 153735, 153736, 153754, 153764, 153767, 153787, 153800, 153810, 153815, 153817, 153823, 153873, 153912, 153928, 153942, 153947, 153958, 153974, 153988, 153999, 154007, 154036, 154054, 154068, 154073, 154080, 154086, 154097, 154103, 154106, 154177, 154183, 154248, 154281, 154288, 154294, 154304, 154315, 154377, 154391, 154393, 154506, 152239, 152737, 154433, 153899, 153733, 153740, 153744, 153793, 153914, 153984, 151787, 151853, 151854, 152461, 152569, 152574, 153099, 151796, 152435, 152441, 151841, 152199, 152852, 153938, 153970, 154273, 154368, 151846, 151890, 151914, 151998, 152034, 152035, 152069, 152092, 152105, 152125, 152138, 152147, 152173, 152230, 152235, 152236, 152245, 152284, 152309, 152340, 152360, 152387, 152409, 152417, 152418, 152515, 152561, 152562, 152563, 152565, 152575, 152630, 152716, 152855, 152874, 152875, 152879, 152882, 152973, 152976, 152978, 152999, 153034, 153062, 153094, 153182, 153283, 153387, 153404, 153489, 153559, 153580, 153639, 153781, 153786, 153831, 153848, 153856, 153957, 154104, 154129, 154316, 154507, 153172, 153616, 152779, 153575, 153742, 151943, 151996, 152042, 152066, 152114, 152141, 152165, 152189, 152224, 152254, 152256, 152264, 152266, 152313, 152443, 152492, 152524, 152567, 152591, 152607, 152648, 152844, 152900, 152908, 152923, 153065, 153169, 153313, 153533, 153567, 153635, 153755, 153853, 154259, 151932, 151930, 152002, 152054, 152100, 152109, 152215, 152247, 152322, 152343, 152507, 152613, 152757, 152906, 153002, 153011, 153028, 153029, 153160, 153195, 153215, 153251, 153331, 153424, 153556, 153566, 153606, 153641, 153812, 153821, 153944, 153966, 154044, 154107, 151596, 151819, 151862, 151870, 152032, 152087, 152104, 152106, 152180, 152204, 152243, 152260, 152275, 152294, 152304, 152324, 152345, 152351, 152357, 152494, 152495, 152528, 152544, 152550, 152551, 152553, 152577, 152624, 152637, 152667, 152727, 152758, 152759, 152762, 152800, 152805, 152830, 152892, 152893, 152938, 152953, 152980, 152991, 152998, 153018, 153039, 153051, 153056, 153057, 153058, 153074, 153075, 153088, 153100, 153101, 153108, 153109, 153176, 153197, 153201, 153203, 153210, 153223, 153224, 153233, 153235, 153238, 153265, 153268, 153269, 153301, 153302, 153341, 153344, 153345, 153347, 153463, 153473, 153478, 153495, 153497, 153500, 153501, 153504, 153570, 153572, 153574, 153634, 153645, 153666, 153672, 153684, 153694, 153710, 153785, 153814, 153874, 153875, 153882, 153883, 153917, 153923, 153953, 153964, 153973, 154009, 154157, 154160, 154194, 154196, 154215, 154220, 154223, 154241, 154261, 154326, 154359, 154437, 154441, 151879, 151974, 152206, 152851, 152073, 152595, 152710, 152711, 152733, 152861, 152975, 153087, 153577, 153651, 151656, 152015, 152947, 153509, 152295, 152311, 152428, 152458, 152622, 152623, 152632, 152633, 152697, 152741, 152868, 152887, 152949, 152983, 153081, 153082, 153725, 153726, 153780, 153849, 153880, 154030, 154141, 154344, 151966, 151662, 152535, 152915, 153348, 154203, 152871, 152988, 152989, 153116, 153117, 153209, 153221, 153247, 153252, 152048, 152050, 152713, 153355, 154371, 151800, 151835, 152323, 152903, 153543, 153604, 153691, 154000, 154210, 152336, 152625, 152626, 152700, 152701, 152234, 152705, 153208, 153219, 154540, 152346, 152981, 153859, 154095, 151833, 151899, 151899, 152152, 152172, 152281, 152426, 152793, 153016, 153153, 153202, 153378, 153416, 153530, 153584, 153696, 153697, 153361, 152847, 151990, 152085, 152099, 152108, 152122, 152212, 152227, 152240, 152273, 152282, 152300, 152442, 152445, 152447, 152459, 152464, 152525, 152537, 152540, 152541, 152547, 152660, 152795, 152827, 152846, 152848, 152933, 153333, 153503, 153698, 153825, 154466, 152071, 154125, 152263, 152412, 153090, 153436, 154448, 151600, 153724, 151697, 151697, 151907, 151908, 152135, 152148, 152156, 152168, 152321, 152396, 152400, 152401, 152402, 152564, 152592, 152654, 152730, 152731, 152752, 152822, 152825, 152836, 152889, 152890, 152891, 153000, 153042, 153043, 153173, 153174, 153175, 153212, 153241, 153371, 153382, 153386, 153415, 153445, 153807, 154026, 154353, 154461, 151705, 151763, 151770, 151779, 151782, 151803, 151820, 151828, 151842, 151847, 151848, 151861, 151864, 151866, 151875, 151881, 151882, 151886, 151931, 151933, 151962, 151965, 151969, 152023, 152036, 152051, 152064, 152079, 152080, 152091, 152154, 152174, 152176, 152246, 152248, 152262, 152301, 152307, 152308, 152352, 152356, 152373, 152425, 152427, 152433, 152446, 152448, 152454, 152522, 152536, 152549, 152717, 152760, 152787, 152881, 152898, 152939, 152965, 153031, 153194, 153430, 153434, 153595, 152277, 152285, 154356

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 93264).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page