Elekta Medical Linear Accelerator recallZ-0446-2024 · Elekta, Inc. · initiated Oct 25, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Elekta, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of uncontrolled extension.
FDA root cause category: Under Investigation by firm
What the company is doing
Elekta released Field Safety Notice to its consignees on the 10/25/2023 via email. The notice explained the issue, potential risk, and provides information for the safe use of the instrument until Elekta Field Service Representatives can perform the correction.
Recall details
- Recall number
- Z-0446-2024
- Status
- Open, Classified
- Initiated by the company
- Oct 25, 2023
- Posted by the FDA
- Dec 1, 2023
- Recalling firm
- Elekta, Inc.
- Quantity
- 291 units
- Product code
- IYE
- FDA submissions
- K210500
- Distribution
- worldwide
Product description
ELEKTA LIMITED Elekta Infinity, REF XRT 0511, XRT 0521, XRT 0531, XRT 0541; Medical Charged-Particle Radiation Therapy System Accelerator, Linear
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
UDI/DI 05060191071512, Serial Numbers: 153451, 154067, 154072, 153600, 153693, 152596, 152649, 152834, 152956, 152957, 153005, 153054, 153377, 153593, 153602, 153617, 153852, 154048, 154049, 154060, 152971, 152597, 151995, 153685, 154560, 153748, 153979, 153989, 152902, 153670, 152937, 153198, 151839, 151942, 151958, 151959, 151961, 152019, 152103, 152115, 152158, 152207, 152210, 152211, 152223, 152257, 152303, 152364, 152371, 152504, 152520, 152585, 152593, 152604, 152652, 152659, 152715, 152732, 152801, 152802, 152817, 152818, 152828, 152833, 152916, 152927, 152946, 152963, 152970, 152993, 153035, 153046, 153071, 153138, 153141, 153143, 153164, 153200, 153230, 153375, 153450, 153474, 153529, 153534, 153542, 153558, 153727, 153746, 153753, 153761, 153794, 153803, 153824, 153826, 153840, 153871, 153915, 153939, 154051, 154093, 154132, 154135, 154169, 154175, 154204, 154206, 154224, 154242, 154244, 154264, 154312, 154378, 154380, 154422, 154429, 154453, 154536, 154558; UDI/DI 05060191071543, Serial Numbers: 153741, 154363, 152506, 152768, 152969, 153004, 153047, 153048, 153120, 153121, 152478, 152772, 153303, 153402, 153626, 152668, 152669, 152692, 152853, 152979, 152984, 153015, 153017, 153053, 153308, 153802, 154065, 153946, 153960, 154082, 154130, 154144, 154189, 154229, 154289, 154290, 154302, 154311, 154414, 154423, 154475, 153472, 152888, 153298, 154277, 154297, 154310, 154355, 154434, 151044, 152554, 153076, 153237, 153324, 153586, 153961, 154211, 154403, 154231, 153441, 153236, 153511, 153646, 153652, 153655, 153674, 153677, 153845, 153869, 153982, 154134, 154159, 154208, 154209, 154213, 152996, 153010, 153052, 153285, 153288, 153294, 153792, 153799, 154115, 154467, 153350, 153866, 152370, 153878, 154366, 154582, 153933, 153001, 154181, 154205, 153531, 153631, 153692, 153743, 153987, 154045, 154468, 152709, 153615, 153103, 153332, 153907, 153091, 152031, 152163, 152175, 152192, 152200, 152222, 152388, 152392, 152432, 152469, 152481, 152491, 152502, 152556, 152599, 152601, 152618, 152673, 152784, 152792, 152798, 152955, 152959, 153019, 153021, 153032, 153036, 153050, 153061, 153068, 153114, 153136, 153253, 153282, 153319, 153321, 153342, 153354, 153429, 153442, 153448, 153464, 153491, 153493, 153535, 153536, 153548, 153597, 153598, 153681, 153719, 153738, 153749, 153798, 154069, 154117, 154182, 154319, 154387, 154485, 154504, 154512, 154535, 154565, 154265
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Elekta Medical Linear Accelerator 1552092Elekta Solutions AB · Accelerator, linear, medical5 recalls0 reports
- Elekta Medical Linear Accelerator 1550358Elekta Solutions AB · Accelerator, linear, medical4 recalls2 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 93264).
- Z-0443-2024 · Open, ClassifiedELEKTA LIMITED Elekta Synergy, REF MRT 8501, MRT 8511; Medical Charged-Particle Radiation Therapy System Accelerator, Linear
- Z-0445-2024 · Open, ClassifiedELEKTA LIMITED Versa HD, REF XRT 2121, 2131; Medical Charged-Particle Radiation Therapy System Accelerator, Linear
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.