Skip to content

BD Insyte Autoguard recallZ-0400-2024 · Becton Dickinson Infusion Therapy Systems Inc. · initiated Oct 4, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Becton Dickinson Infusion Therapy Systems Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Over-the-needle, intravascular catheters may have needles that do not retract properly into the barrel, which could result in contaminated needlestick injury, blood leaking from the insertion site or catheter, and repeat catheter insertion attempts, due to catheter being inadvertently removed from the insertion site.

FDA root cause category: Nonconforming Material/Component

What the company is doing

On 10/4/23, recall notices were emailed to customers and distributors who were asked to do the following: 1) Destroy all unused product subject to the recall following your institution's process for destruction. 2) For indwelling catheters, no further action is needed. 3) Share this recall notification with all users within your facility network of the product to ensure they are also aware of this recall. 4) Complete and return the customer response form via email to [email protected] Customers with recall questions, product complaints, or technical questions can call 1-844-823-5433, say recall when prompted, Mon-Fri 8:00am to 5:00PM CT, or email [email protected] In addition, distributors are asked to identify all customers within their distribution network that purchased affected product. Then distributors should provide a copy of the customer letter to these customers and advise them of the recall.

Recall details

Recall number
Z-0400-2024
Status
Open, Classified
Initiated by the company
Oct 4, 2023
Posted by the FDA
Nov 24, 2023
Recalling firm
Becton Dickinson Infusion Therapy Systems Inc.
Quantity
528,000
Product code
FOZ
FDA submissions
K201075
Distribution
US Nationwide distribution in the states of CA, NC, MI, WA, CO, FL, NJ, NM, IL, KY, TN, OH, TX, IN, LA, IA, MS, MO, NE, MD, NY, PA, GA, VA, ME, SC, RI, MA, PR.

Product description

BD Insyte Autoguard BC Shielded IV Catheter with Blood Control Technology 20GA X 1.00in, REF: 382533

Code information (models, lots, serial numbers)

UDI-DI: 00382903825332, Lot: 3166524, Expiration: 2026/05/31

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

53 devices cleared under K201075 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates