Skip to content

Big Bore RT with software version V4.8.0.10421- Model:… recallZ-0292-2024 · Philips North America Llc · initiated Oct 9, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Philips North America Llc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

For oncology users: If the user performed off-set reconstruction on CT device a shift could be observed on the contouring area of the primary and secondary images sets when using image fusion (on TPS) and incorrect coordinates may be sent to TPS system when using absolute patient marking. This may lead to incorrect radiation therapy planning and possible growth or spread of cancer due to incorrect early treatment regimen. For radiology users: If the user performed off-set reconstruction on CT devices, after multiple acquisitions, the Relate Position could be inaccurate. This could lead to incorrect diagnosis and treatment.

FDA root cause category: Under Investigation by firm

What the company is doing

Philips issued URGENT Medical Device Correction letter (UMDC) 2023-PD-CTAMI-015) dated 10/9/23. Letter states reason for recall, health risk and action to take: Below are short-term precautions to take until the permanent solution is installed. Enable the Force X/Y to 0/0 option on your device (Refer to the Instructions For Use, Section 3 Preparing for an Exam under Preferences, as well as Section 5 Summary of Scanning Workflows under Scan Workflow). Enabling this option will avoid the issue. After enabling the above referenced option, you may continue to use your system(s) in accordance with the intended use. Please complete and return the attached response form to Philips promptly and no later than 30 days from receipt. Completing this form confirms receipt of the Urgent Medical Device Correction Letter, understanding of the issue, and required actions to be taken. Circulate this Urgent Medical Device Correction Letter to all users of this device so that they are aware of the issue. Please retain this letter with your system(s) until a solution is installed on your system; ensure the letter is in a place likely to be seen/viewed. Philips will contact you to schedule a time for a Philips Field Service Engineer (FSE) to visit your site and install the software update (FCO 72800806) to resolve the issue. If you need any further information or support concerning this issue, please contact the Customer Care Solutions Center (1-800-722-9377).

Recall details

Recall number
Z-0292-2024
Status
Open, Classified
Initiated by the company
Oct 9, 2023
Posted by the FDA
Nov 14, 2023
Recalling firm
Philips North America Llc
Quantity
56 units
Product code
JAK
FDA submissions
K171850
Distribution
Nationwide Foreign: Austria Belgium Canada China Estonia France Germany Hondorus India Italy Mexico Netherlands Philippines Poland Singapore Spain Switzerland Taiwan

Product description

Big Bore RT with software version V4.8.0.10421- Model: 728242 CT scanner and simulator designed for radiation oncology and therapy.

Code information (models, lots, serial numbers)

UDI-DI: (01)00884838095168 Serial Numbers; Serial Number 20094 20078 766037 20064 20051 20063 766015 766013 20053 20052 20039 766003 763011 20033 763006 20023 763004 20092 766075 768020 20093 766073 766074 766072 766071 766070 766069 766068 766065 766067 766066 766064 766061 766062 766063 766060 766059 766058 763030 766057 20086 20090 763033 763032 766080 763031 766081 766079 766078 20091 766076 766077 763010 763018 20040 766047

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 93291).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page