Big Bore RTRecalls & reported problems

Philips Medical Systems Nederland B.V. · Class 2

4 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
4

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00884838095168), received between Jan 25, 2024 and Jan 25, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).

4 FDA recall records name this device's identifier, 4 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 13, 2025Open, Classified

    The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.

    Z-2597-2025 · Root cause: Process control · FDA record ↗

  2. Apr 1, 2025Open, Classified

    Unintended motion issues related to Interventional Control that may lead to collision of the Gantry/table with the operator or patient.

    Z-1660-2025 · Root cause: Software design · FDA record ↗

  3. Oct 9, 2023Open, Classified

    For oncology users: If the user performed off-set reconstruction on CT device a shift could be observed on the contouring area of the primary and secondary images sets when using image fusion (on TPS) and incorrect coordinates may be sent to TPS system when using absolute patient marking. This may lead to incorrect radiation therapy planning and possible growth or spread of cancer due to incorrect early treatment regimen. For radiology users: If the user performed off-set reconstruction on CT devices, after multiple acquisitions, the Relate Position could be inaccurate. This could lead to incorrect diagnosis and treatment.

    Z-0292-2024 · Root cause: Under Investigation by firm · FDA record ↗

  4. May 1, 2021Open, Classified

    Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.

    Z-1983-2021 · Root cause: Device Design · FDA record ↗

Most reported problems

Device problems

  • Insufficient Device Problem Information1
  • Program or Algorithm Execution Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 25, 2024Insufficient Device Problem Information; Program or Algorithm Execution Problem; Insufficient Device Problem Information; Program or Algorithm Execution Problem

    THIS COMPLAINT HAS BEEN EVALUATED BASED ON THE INFORMATION PROVIDED; THERE IS NO ALLEGATION OF DEATH OR SERIOUS INJURY. IT WAS REPORTED THAT DURING THE PHILIPS INTERNAL TESTING PROCESS, IT WAS FOUND THAT THE CT SERIES RECONSTRUCTED WITH A X OR Y RECONSTRUCTION OFFSET PER DICOM PRIVATE TAG (01F1,100C), WHICH CAN BE USED WITHIN TUMOR LOC TO TRANSMIT INVALID ABSOLUTE MARKING COORDINATES TO A LASER MARKING SYSTEM. THESE INVALID COORDINATES MIGHT RESULT IN AN INCORRECT TREATMENT LOCATION ON THE PATIENT¿S SKIN. BASED ON THE AVAILABLE INFORMATION, THIS ISSUE HAS BEEN DETERMINED TO BE A REPORTABLE EVENT. PHILIPS HAS STARTED THE INVESTIGATION OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18585099 ↗