ProCuity recallZ-0433-2024 · Stryker Medical Division of Stryker Corporation · initiated Aug 30, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Stryker Medical Division of Stryker Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Patient beds are missing test values for electrical safety tests presenting risk of tissue burn or electrical shock.
FDA root cause category: Process control
What the company is doing
A customer notification was sent October 20, 2023 via FedEx. Users are to (1) Locate affected units and identify the address where they can be serviced. (2) Return the business reply form by fax +1 269 488-8691 or email to [email protected]. Upon receipt of the completed business reply form, Stryker will contact you to arrange a time to complete the required electrical tests. (3) Forward notification to any new user/location and advise Stryker of the new address or device status if disposed of or obsolete using the reply form. Address questions or concerns to Customer Service at +1 800 327-0770 Monday through Friday from 8:00 a.m. to 6:00 p.m. ET.
Recall details
- Recall number
- Z-0433-2024
- Status
- Open, Classified
- Initiated by the company
- Aug 30, 2023
- Posted by the FDA
- Dec 1, 2023
- Recalling firm
- Stryker Medical Division of Stryker Corporation
- Product code
- FNL
- Distribution
- Distribution US Nationwide including Puerto Rico, Australia, Canada, Chile, China, Mexico, Poland, Spain and Switzerland.
Product description
ProCuity bed series, model number 3009, item number: 3009PX-ZM-600.
Code information (models, lots, serial numbers)
UDI-DI (GTIN): 07613327541311. Serial numbers: 2201000173, 2110001895, 2203001506, 2110000413, 2110000411, 2208002037, 2208002030, 2208002019& 2212000691
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- ProCuity 3009PX-ZM-600Stryker Corporation · BED, AC-POWERED ADJUSTABLE HOSPITAL1 recall9 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 93317).
- Z-0424-2024 · Open, ClassifiedProCuity bed series, model number 3009, item number: 300900000000,
- Z-0425-2024 · Open, ClassifiedProCuity bed series, model number 3009, item number: 3009PX-L-100.
- Z-0426-2024 · Open, ClassifiedProCuity bed series, model number 3009, item number: 3009PX-L-200
- Z-0427-2024 · Open, ClassifiedProCuity bed series, model number 3009, item number: 3009PX-LE-400.
- Z-0428-2024 · Open, ClassifiedProCuity bed series, model number 3009, item number: 3009PX-LE-450.
- Z-0429-2024 · Open, ClassifiedProCuity bed series, model number 3009, item number: 3009PX-LEX-400.
- Z-0430-2024 · Open, ClassifiedProCuity bed series, model number 3009, item number: 3009PX-LEX-450.
- Z-0431-2024 · Open, ClassifiedProCuity bed series, model number 3009, item number: 3009PX-Z-450.
- Z-0432-2024 · Open, ClassifiedProCuity bed series, model number 3009, item number: 3009PX-ZM-450.
- Z-0434-2024 · Open, ClassifiedProCurity bed series, model number 3009, item number: 3009PX-ZMX-450.
- Z-0435-2024 · Open, ClassifiedProCurity bed series, model number 3009, item number: 3009PX-ZMX-600.
- Z-0436-2024 · Open, ClassifiedProCurity bed series, model number 3009, item number: 3009PX-ZX-450.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page