Skip to content

ProCuity 3009PX-ZM-600Recalls & reported problems

Stryker Corporation · ProCuity ZM + Isolibrium · Class 2

All models of ProCuity (17)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
9
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 07613327541311), received between Jul 19, 2023 and Jul 23, 2025 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Intermittent Loss of Power” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 3 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 30, 2023Class IIOpen, Classified

    Patient beds are missing test values for electrical safety tests presenting risk of tissue burn or electrical shock.

    Z-0433-2024 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Intermittent Loss of Power3
  • Incorrect Measurement2
  • Unintended System Motion2
  • Electrical /Electronic Property Problem1
  • No Audible Alarm1

Patient problems

  • No Clinical Signs, Symptoms or Conditions9

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 23, 2025Intermittent Loss of Power; Intermittent Loss of Power

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED ZOOM SUDDENLY STOPS. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22578736 ↗
  • MalfunctionApr 28, 2025Intermittent Loss of Power; Intermittent Loss of Power

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED ZOOM SUDDENLY STOPS. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21919965 ↗
  • MalfunctionOct 28, 2024No Audible Alarm; No Audible Alarm

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED CLINICIAN NOT ALERTED/AWARE OF POSSIBLE PATIENT FALL. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20547385 ↗
  • MalfunctionJul 29, 2024Unintended System Motion; Unintended System Motion

    THIS REPORT SUMMARIZES 0 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED UN-COMMANDED MOTION OF THE ELECTRONIC POSITIONING FUNCTIONS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19850107 ↗
  • MalfunctionJul 26, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED ACCESSIBLE AC CURRENT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19839458 ↗
  • MalfunctionJul 26, 2024Incorrect Measurement; Incorrect Measurement

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED SCALE MEASURED WEIGHT INACCURATE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19838177 ↗
  • MalfunctionJan 24, 2024Incorrect Measurement; Incorrect Measurement

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED SCALE MEASURED WEIGHT INACCURATE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18571842 ↗
  • MalfunctionOct 23, 2023Intermittent Loss of Power; Intermittent Loss of Power

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED ZOOM SUDDENLY STOPS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17990492 ↗
  • MalfunctionJul 19, 2023Unintended System Motion; Unintended System Motion

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED UNINTENDED DIRECTION OF THE ZOOM DRIVE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17350168 ↗