Compress Device recallZ-0374-2024 · Biomet, Inc. · initiated Oct 9, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Anchor plugs potentially have metal burrs in the transverse holes of the device, preventing the drill or pins from passing through, may result in extension of surgery
FDA root cause category: Device Design
What the company is doing
Zimmer Biomet issued Urgent Medical Device Recall letter on 10/09/23 to Distributors and Risk Managers via email/courier . Letter states reason for recall, health risk and action to take: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with the field safety notice for hospitals and ensure documentation. 4. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. This form shall be returned even if you do not have affected products at your facility. 5. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation. 6. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to [email protected]
Recall details
- Recall number
- Z-0374-2024
- Status
- Open, Classified
- Initiated by the company
- Oct 9, 2023
- Posted by the FDA
- Nov 21, 2023
- Recalling firm
- Biomet, Inc.
- Product code
- KWA
- FDA submissions
- K062998
- Distribution
- Worldwide distribution - US Nationwide and the countries of ARGENTINA, AUSTRALIA, CANADA, CHILE, JAPAN, NETHERLANDS, NEW ZEALAND.
Product description
Compress Device Segmental Anchor Plug, 18 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178408
Code information (models, lots, serial numbers)
Lot Numbers/UDI: 017910 (01)00880304259768(17)270807(10)017910; 017910R (01)00880304259768(17)270807(10)017910R; 018060 (01)00880304259768(17)270817(10)018060; 332060 (01)00880304259768(17)270817(10)332060; 608610 (01)00880304259768(17)271204(10)608610; 671740 (01)00880304259768(17)271117(10)671740; 671740R (01)00880304259768(17)271117(10)671740R; 752790 (01)00880304259768(17)270816(10)752790; 752790R (01)00880304259768(17)270816(10)752790R; 965290 (01)00880304259768(17)271030(10)965290;
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Compress® Device 178408Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 recall0 reports
Product lines
Other devices under the same FDA submission
15 devices cleared under K062998 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Compress® Device 178562Biomet Orthopedics, LLC · PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER2 recalls2 reports
- Compress® Device 178552Biomet Orthopedics, LLC · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented1 recall9 reports
- Compress® Device 178554Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 recall5 reports
- Compress® Device 178556Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate1 recall4 reports
- Compress® Device 178558Biomet Orthopedics, LLC · PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER1 recall4 reports
- Compress® Device 178560Biomet Orthopedics, LLC · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 recall1 report
- Compress® Device 178568Biomet Orthopedics, LLC · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented1 recall0 reports
- Compress® Device 178564Biomet Orthopedics, LLC · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented1 recall0 reports
- Compress® Device 178566Biomet Orthopedics, LLC · PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER1 recall0 reports
- Compress® Device 178570Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate1 recall0 reports
- COMPRESS DEVICE 178580Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 recall0 reports
- COMPRESS DEVICE 178577Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 93323).
- Z-0370-2024 · Open, ClassifiedCompress Device Segmental Anchor Plug, 10 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178400
- Z-0371-2024 · Open, ClassifiedCompress Device Segmental Anchor Plug, 12 MM -For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178402
- Z-0372-2024 · Open, ClassifiedCompress Device Segmental Anchor Plug, 14 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178404
- Z-0373-2024 · Open, ClassifiedCompress Device Segmental Anchor Plug, 16 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178406
- Z-0375-2024 · Open, ClassifiedCompress Device Segmental Anchor Plug, 20 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178410
- Z-0376-2024 · Open, ClassifiedCompress Device Segmental Anchor Plug, 22 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178412
- Z-0377-2024 · Open, ClassifiedCompress Device Segmental Anchor Plug, 24 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178414
- Z-0378-2024 · Open, ClassifiedCompress Device Short Anchor Plug, 10 MM-Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement Item Number: 178552
- Z-0379-2024 · Open, ClassifiedCompress Device Short Anchor Plug, 12 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178554
- Z-0380-2024 · Open, ClassifiedCompress Device Short Anchor Plug, 14 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178556
- Z-0381-2024 · Open, ClassifiedCompress Device Short Anchor Plug, 16 MM- Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement Item Number: 178558
- Z-0382-2024 · Open, ClassifiedCompress Device Short Anchor Plug, 18 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178560
- Z-0383-2024 · Open, ClassifiedCompress Device Short Anchor Plug, 20 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178562
- Z-0384-2024 · Open, ClassifiedCompress Device Short Anchor Plug, 22 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178564
- Z-0385-2024 · Open, ClassifiedCompress Device Short Anchor Plug, 23 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178565
- Z-0386-2024 · Open, ClassifiedCompress Device Short Anchor Plug, 24 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178566
- Z-0387-2024 · Open, ClassifiedCompress Device Short Anchor Plug, 26 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178568
- Z-0388-2024 · Open, ClassifiedCompress Device Short Anchor Plug, 28 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178570
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page