Compress® Device 178552Recalls & reported problems
Biomet Orthopedics, LLC · Class 2 · Implant
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00880304461659), received between Jun 30, 2021 and Sep 9, 2025 — classified as 5 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Fracture” (5 reports).
1 FDA recall record name this device's identifier, 1 still open.
20 other recalls cover devices cleared under the same FDA submission (K062998); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Oct 9, 2023Open, Classified
Anchor plugs potentially have metal burrs in the transverse holes of the device, preventing the drill or pins from passing through, may result in extension of surgery
Z-0378-2024 · Root cause: Device Design · FDA record ↗
Most reported problems
Device problems
- Fracture5
- Adverse Event Without Identified Device or Use Problem3
- Difficult to Insert1
- Material Twisted/Bent1
- Migration1
- Physical Resistance/Sticking1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Failure of Implant2
- Insufficient Information2
- Joint Contracture2
- Loss of Range of Motion2
- Ambulation or Postural Difficulties1
- Bone Fracture(s)1
- Pain1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
InjurySep 9, 2025Migration; Migration
IT WAS REPORTED THE PATIENT UNDERWENT AN INITIAL KNEE ARTHROPLASTY. SUBSEQUENTLY, THE PATIENT COMPLAINED OF PAIN OVER A SPAN OF THE LAST YEAR. THE SURGEON BELIEVED THE EXPANDABLE COMPONENT WAS EXPANDING ON ITS OWN, WITHOUT THE REQUIRED SURGERY TO EXPAND IT. THE PATIENT FELT A POP AND COULD NOT BEAR WEAR. THE PATIENT WAS THEN REVISED APPROXIMATELY 3 YEARS POST IMPLANTATION DUE TO A BROKEN ANCHOR PLUG AND THE SPINDLE LOST BONE IN-GROWTH AT THE BONE INTERFACE AND WAS NO LONGER FIXED.
Patient problems: Ambulation or Postural Difficulties; Pain
MAUDE report 23010320 ↗InjuryDec 30, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT IS BEING CONSIDERED FOR A RIGHT KNEE ARTHROPLASTY REVISION TO ADDRESS FLEXION CONTRACTURE AND STIFFNESS APPROXIMATELY TWENTY-TWO (22) MONTHS FOLLOWING KNEE ARTHROPLASTY. ATTEMPTS HAVE BEEN MADE, HOWEVER, NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Insufficient Information; Joint Contracture; Loss of Range of Motion
MAUDE report 21036223 ↗InjuryDec 30, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT IS BEING CONSIDERED FOR A RIGHT KNEE ARTHROPLASTY REVISION TO ADDRESS FLEXION CONTRACTURE AND STIFFNESS APPROXIMATELY TWENTY-TWO (22) MONTHS FOLLOWING KNEE ARTHROPLASTY. ATTEMPTS HAVE BEEN MADE, HOWEVER, NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Insufficient Information; Joint Contracture; Loss of Range of Motion
MAUDE report 21036166 ↗InjuryAug 12, 2024Fracture; Fracture
IT WAS REPORTED THAT PATIENT UNDERWENT A LEFT KNEE REVISION APPROXIMATELY SIXTEEN (16) MONTHS POST-IMPLANTATION DUE TO FRACTURE OF THE ANCHOR PLUG. NO ADDITIONAL PATIENT CONSEQUENCES WERE REPORTED.
Patient problems: Failure of Implant
MAUDE report 19960275 ↗MalfunctionJan 22, 2024Fracture; Difficult to Insert; Physical Resistance/Sticking; Fracture; Difficult to Insert; Physical Resistance/Sticking
IT WAS REPORTED THAT WHILE TRYING TO ALIGN THE DRILL GUIDE AND THE ANCHOR PLUG DURING SURGERY, THE DRILL BIT SEIZED WITH THE GUIDE AND THE DRILL BIT FRACTURED. STAFF WERE UNSURE IF THE GUIDE OR THE ANCHOR PLUG WAS NOT LINING UP EVENLY WITH THE DRILL BITS THROUGH THE BOOM AND ANCHOR IN-SITU. THE SURGERY WAS NOT COMPLETED WITH ANY SECOND DEVICES. THERE WERE NO FOREIGN BODIES RETAINED. ATTEMPTS HAVE BEEN MADE AND ALL AVAILABLE INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18553841 ↗MalfunctionJan 12, 2023Fracture; Fracture
IT WAS REPORTED THE TRACTION BAR OF COMPRESS ANCHOR PLUG FRACTURED WHILE PLACING THE COMPRESSION NUT. NO PIECES FELL INTO THE PATIENT. NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16150098 ↗MalfunctionDec 29, 2022Fracture; Fracture
IT WAS REPORTED THAT DURING A TOTAL RIGHT KNEE PROCEDURE, AN ANCHOR PLUG IMPLANT FRACTURED DURING IMPLANTATION. THERE WAS A SURGICAL DELAY OF FORTY-FIVE (45) MINUTES WHILE THE FRACTURED COMPONENT AND THE ASSOCIATED CONNECTIONS WERE REMOVED AND REPLACEMENT DEVICES WERE IMPLANTED. THERE WAS NO PATIENT IMPACT REPORTED AS A RESULT OF THIS MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16075215 ↗MalfunctionAug 9, 2022Fracture; Fracture
IT WAS REPORTED THAT DURING A RIGHT KNEE PROCEDURE, AN ANCHOR PLUG IMPLANT FRACTURED DURING IMPLANTATION. THERE WAS A SURGICAL DELAY OF FORTY MINUTES WHILE THE FRACTURED COMPONENT AND THE ASSOCIATED CONNECTIONS WERE REMOVED AND REPLACEMENT DEVICES WERE IMPLANTED. THERE WAS NO PATIENT IMPACT REPORTED AS A RESULT OF THIS MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15191150 ↗InjuryJun 30, 2021Material Twisted/Bent; Adverse Event Without Identified Device or Use Problem; Material Twisted/Bent; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THE PATIENT UNDERWENT A RIGHT KNEE PROCEDURE. SUBSEQUENTLY, THE PATIENT WAS SCHEDULED TO BE REVISED DUE TO A FAILED COMPRESS. THE PATIENT THEN UNDERWENT A REVISION SURGERY ONE WEEK AFTER INITIAL IMPLANTATION. THE TRACTION BAR APPEARS TO HAVE BENT. THE SURGEON STATED HE IS UNSURE IF THE FRACTURED BONE WAS AN EFFECT OR A CAUSE OF THE FAILED COMPRESS. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: Bone Fracture(s); Failure of Implant
MAUDE report 12094379 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K062998. One submission often covers many models, so check the product and lot details in each record.
- Z-0387-2024 · Oct 9, 2023 · Open, ClassifiedCompress Device Short Anchor Plug, 26 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178568
- Z-0386-2024 · Oct 9, 2023 · Open, ClassifiedCompress Device Short Anchor Plug, 24 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178566
- Z-0377-2024 · Oct 9, 2023 · Open, ClassifiedCompress Device Segmental Anchor Plug, 24 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178414
- Z-0381-2024 · Oct 9, 2023 · Open, ClassifiedCompress Device Short Anchor Plug, 16 MM- Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement Item Number: 178558
- Z-0383-2024 · Oct 9, 2023 · Open, ClassifiedCompress Device Short Anchor Plug, 20 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178562
- Z-0372-2024 · Oct 9, 2023 · Open, ClassifiedCompress Device Segmental Anchor Plug, 14 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178404
- Z-0384-2024 · Oct 9, 2023 · Open, ClassifiedCompress Device Short Anchor Plug, 22 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178564
- Z-0379-2024 · Oct 9, 2023 · Open, ClassifiedCompress Device Short Anchor Plug, 12 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178554
- Z-0380-2024 · Oct 9, 2023 · Open, ClassifiedCompress Device Short Anchor Plug, 14 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178556
- Z-0376-2024 · Oct 9, 2023 · Open, ClassifiedCompress Device Segmental Anchor Plug, 22 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178412
- Z-0385-2024 · Oct 9, 2023 · Open, ClassifiedCompress Device Short Anchor Plug, 23 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178565
- Z-0373-2024 · Oct 9, 2023 · Open, ClassifiedCompress Device Segmental Anchor Plug, 16 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178406
- Z-0382-2024 · Oct 9, 2023 · Open, ClassifiedCompress Device Short Anchor Plug, 18 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178560
- Z-0374-2024 · Oct 9, 2023 · Open, ClassifiedCompress Device Segmental Anchor Plug, 18 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178408
- Z-0375-2024 · Oct 9, 2023 · Open, ClassifiedCompress Device Segmental Anchor Plug, 20 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178410
…and 5 more.
Other Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented devices
- STAGEONE 431209Biomet Orthopedics, LLC · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented2 recalls0 reports
- SELF CENTERING 1035-45-000Depuy Orthopaedics, Inc. · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented84 reports
- LFIT 6260-9-036Howmedica Osteonics Corp. · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented80 reports
- ACTIS 101011060Depuy (Ireland) · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented66 reports
- RINGLOC BI-POLAR HIP SYSTEM 11-165228Biomet Orthopedics, LLC · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented1 recall1 death12 reports
- SELF CENTERING 1035-44-000Depuy Orthopaedics, Inc. · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented60 reports
- TANDEM 71324046Smith & Nephew, Inc. · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented1 recall9 reports
- TAPERLOC COMPLETE XR 123 MICRO FEMORAL 51-145170Biomet Orthopedics, LLC · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented59 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.