Skip to content

CardioQuip Modular Cooler-Heater, Model:MCH-1000(m), Power:… recallZ-0438-2024 · CardioQuip, LLC · initiated Oct 19, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from CardioQuip, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There is a potential for the heating-chamber to malfunction resulting in melting/burning of the heater chamber components.

FDA root cause category: Device Design

What the company is doing

On 10/19/23, CardioQuip issued a "Urgent: Medical Device Correction" notification to all affected consignees. This notification was an update to the firm's May 22, 2023 communication. CardioQuip ask consignees to take the following actions: 1. Prime the MCH-1000(i) and MCH-1000(m) prior to each use. 2.Never operate the device dry (see communication for further details). 3. Design Changes to be implemented (see communication for further details). A. Relocation of the water Level Sensor (see communication for further details). B. Software Update (see communication for further details). 4. Please fill out and submit the receipt of understanding of the contents of this letter to CardioQuip within 5 business days or receipt. 5. Any adverse events experienced relating to CardioQuip devices should be reported to CardioQuip or through the FDA s MedWatch Program: Phone: +1 (800) FDA-1088 Web: www.fda.gov/medwatch/report.htm

Recall details

Recall number
Z-0438-2024
Status
Open, Classified
Initiated by the company
Oct 19, 2023
Posted by the FDA
Dec 1, 2023
Recalling firm
CardioQuip, LLC
Quantity
150 units
Product code
DWC
FDA submissions
K102147
Distribution
US Distribution: TX, IN, IL, WI, FL, NJ, PA, KY, VA, WV, CT, CA, WA, AL, IA

Product description

CardioQuip Modular Cooler-Heater, Model:MCH-1000(m), Power: 115V-60hz,16A

Code information (models, lots, serial numbers)

Show the full code information (2k characters)

UDI: (01)00860000846110/Serial Numbers: 11160203, 11160204, 11160315, 11160316, 11160317, 11160318, 11160319, 11160320, 11161109, 11161832, 11161836, 11161837, 11161838, 11161839, 11160216, 11160217, 11160218, 11160548, 11160549, 11160550, 11160103, 11160338, 11160339, 11160364, 11160365, 11160366, 11161117, 11161118, 11161827,11161828, 11161829, 11161942, 11161943, 11161944, 11161978, 11162016, 11162018, 11160461, 11160462, 11160463, 11160464, 11160465, 11160579, 11161487, 11161488, 11161489, 11161613, 11161692, 11161695, 11161701, 11161702, 11161696 11161627, 11162262, 11160298, 11160378, 11160380, 11160385, 11160386, 11160387, 11160388, 11160377, 11160379, 11160376, 11160439, 11160440, 11161458, 11161460, 11161815, 11161816, 11161817, 11161818, 11161819, 11161907, 11161916, 11162057, 11162078, 11162079, 11162080, 11160375, 11160083, 11160270, 11160271, 11161018, 11161024, 11161019, 11161023, 11161022, 11161020, 11161021, 11160553, 11160554, 11161143, 11161144, 11161147, 11162002, 11162008, 11162009, 11162010, 11162011, 11160277, 11160290, 11160391, 11160289, 11161667, 11161668, 11161669, 11161670, 11161972, 11161973, 11161294, 11162139, 11162141, 11162142, 11162143, 11162170, 11162171 11160031, 11161076, 11161077, 11161890, 11162128, 11162129, 11161187, 11161189, 11161193, 11161194, 11161545, 11161546, 11161547, 11161549, 11160024, 11160025, 11160027, 11160075, 11161126, 11160314, 11162053, 11162054 11160071 11161083, 11161084, 11161085, 11161086, 11161087, 11161088, 11161089, 11161090, 11161091, 11161092

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

1 device cleared under K102147 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 93347).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page