CELL-DYN recallZ-0604-2024 · Abbott Laboratories · initiated Oct 19, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Abbott Laboratories or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The devices contain a dry natural rubber (latex) subcomponent, which are not labeled as containing dry natural rubber(latex).
FDA root cause category: Under Investigation by firm
What the company is doing
The recalling firm issued the Product Correction letter dated 10/19/2023 via FedEx priority overnight express delivery in the U.S. and either in person or mail, telephone, email, or FAX to the OUS consignees. The product correction is for instruments distributed in the U.S. and Brazil only. The letter explains the issue, lists all products involved (some of which have different 806 report numbers and are being submitted by Division III) and that a precautionary label indicating the presence of dry natural rubber (latex) will be applied to the devices by the local Abbott representative. The Impact to User Safety is provided and actions to be taken by the customer, which included: Follow good laboratory practice by wearing impervious gloves and other personal protective equipment when working on the listed Abbott products; complete and return the Customer Reply Form; if the consignee has forwarded the product(s) listed to other laboratories, inform them of this Product Correction and provide a copy of the letter; and retain this letter for their laboratory records.
Recall details
- Recall number
- Z-0604-2024
- Status
- Open, Classified
- Initiated by the company
- Oct 19, 2023
- Posted by the FDA
- Dec 21, 2023
- Recalling firm
- Abbott Laboratories
- Quantity
- 670 devices (329 US and 341 OUS)
- Product code
- GKZ
- FDA submissions
- K061667, K051215
- Distribution
- Distribution was made to AL, AR, AZ, CA, CO, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NV, NY, OH, OK, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY, including Puerto Rico. There was also government and military distribution. Foreign distribution was made to Brazil.
Product description
Abbott CELL-DYN Ruby, Model CD-Ruby, List number 08H67-01, and Abbott CELL-DYN Sapphire, Model CD Sapphire, list number 08H00-01. Multi-parameter, automated hematology analyzer.
Code information (models, lots, serial numbers)
All serial numbers are involved. CELL-DYN Ruby - UDI/DI 00380740017170; CELL-DYN Sapphire - UDI/DI 00380740016616.
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- CELL-DYN 08H67-01Abbott Laboratories · Counter, differential cell6 recalls269 reports
- CELL-DYN 08H00-01Abbott Laboratories · Counter, differential cell1 recall90 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page