CELL-DYN 08H67-01Recalls & reported problems
Abbott Laboratories · CELL-DYN Ruby · Class 2
All models of CELL-DYN (12)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
6 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Abbott CELL-DYN Ruby blood analyzer has 268 reports, with no deaths and 10 injuries. Many describe pressure problems, and some describe test results that are too high or too low, smoke from the analyzer, or staff cut or splashed while servicing it. Six open recalls name this device, covering software that can accept expired test chemicals without warning and a latex part not labeled as latex.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 269
- Deaths reported
- 0
- Injuries reported
- 10
- Recalls naming it
- 6
Summary
The FDA's MAUDE database contains 269 adverse event reports that list this exact device (DI 00380740017170), received between Sep 15, 2016 and Sep 4, 2026 — classified as 10 injuries, 259 malfunctions.
The most frequently reported specific device problem is “Pressure Problem” (105 reports).
6 FDA recall records name this device's identifier, 6 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Mar 18, 2025Class IIOpen, Classified
When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use of expired reagents is against the guidelines of the Operator's Manual.
Z-1655-2025 · Root cause: Software change control · Recall details · FDA record ↗
- Mar 18, 2025Class IIOpen, Classified
When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use of expired reagents is against the guidelines of the Operator's Manual.
Z-1652-2025 · Root cause: Software change control · Recall details · FDA record ↗
- Mar 18, 2025Class IIOpen, Classified
When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use of expired reagents is against the guidelines of the Operator's Manual.
Z-1653-2025 · Root cause: Software change control · Recall details · FDA record ↗
- Mar 18, 2025Class IIOpen, Classified
When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use of expired reagents is against the guidelines of the Operator's Manual.
Z-1656-2025 · Root cause: Software change control · Recall details · FDA record ↗
- Mar 18, 2025Class IIOpen, Classified
When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use of expired reagents is against the guidelines of the Operator's Manual.
Z-1654-2025 · Root cause: Software change control · Recall details · FDA record ↗
- Oct 19, 2023Class IIOpen, Classified
The devices contain a dry natural rubber (latex) subcomponent, which are not labeled as containing dry natural rubber(latex).
Z-0604-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Pressure Problem105
- Low Test Results57
- High Test Results37
- Leak/Splash13
- Smoking12
- Incorrect, Inadequate or Imprecise Result or Readings9
- Incorrect Or Inadequate Test Results8
- Thermal Decomposition of Device6
Patient problems
- No Consequences Or Impact To Patient113
- No Clinical Signs, Symptoms or Conditions61
- No Known Impact Or Consequence To Patient54
- No Patient Involvement18
- No Code Available16
- Needle Stick/Puncture4
- Muscle/Tendon Damage2
- Eye Injury1
Recent adverse event reports
The 25 most recent of 269 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 4, 2026Low Test Results
The customer observed falsely decreased hemoglobin (Hgb) results by the CELL DYN Ruby analyzer for a 82yrs old female patient. The results provided were:On (b)(6)2026 SID (b)(6) Hemoglobin: Initial=6.44 g/dL /repeated=6.31 g/dL.Redrawn and repeated SID (b)(6) =10.1 g/dL.On (b)(6)2026=9.4 g/dL.ON (b)(6)2026=10.2 g/dL.On (b)(6)2026=11.0 g/dL.ON (b)(6)2026=12.3 g/dL.On (b)(6)2026=10.7 g/dL.There was no reported impact to patient management.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50601572 ↗MalfunctionMar 26, 2026Smoking
THE CUSTOMER HEARD AN EXPLOSION/LOUD BANG, SMELLED A BURNING ODOR, AND OBSERVED SMOKE COMING OUT OF THE CELL-DYN RUBY ANALYZER. THE INSTRUMENT SHUT DOWN AND THE CUSTOMER IMMEDIATELY DISCONNECTED THE INSTRUMENT. THERE WAS NO VISIBLE EVIDENCE SHOWING SIGNS OF FIRE ON THE OUTSIDE OF THE CELL-DYN RUBY ANALYZER. ADDITIONALLY, THE CUSTOMER REPORTED THERE WAS NO HARM WHATSOEVER TO THE USERS OR THE FACILITIES IN THE LABORATORY. FURTHER INFORMATION WAS PROVIDED THAT DURING THE VISUAL INSPECTION; THERE WAS EVIDENCE INDICATING THAT THE CAUSE WAS THE LAMBDA POWER SUPPLY WHICH WAS THEN REPLACED. IT WAS CONFIRMED THAT THE INSTRUMENT NO LONGER OVERHEATED OR SHUT DOWN. BACKGROUNDS WERE PROCESSED WITHIN SPECIFICATIONS AND CONTROLS WERE PROCESSED WITHIN RANGE. THE INSTRUMENT WAS DELIVERED IN GOOD WORKING ORDER AND WAS APPROVED FOR ROUTINE USE. NO IMPACT TO PATIENT MANAGEMENT OR IMPACT TO USER SAFETY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24698343 ↗MalfunctionMar 16, 2026Low Test Results
THE CUSTOMER OBSERVED IMPRECISE HEMOGLOBIN RESULTS ON A CELL-DYN RUBY ANALYZER FOR ONE SAMPLE. THE FOLLOWING RESULTS WERE PROVIDED: REP 1 HGB = 8.38. REP 2 HGB = 12.7. REP 3 HGB = 5.94. REP 4 HGB = 8.25. REP 5 HGB = 7.98 THE SAMPLE WAS REPEATED ON ANOTHER INSTRUMENT GENERATING A HGB RESULT OF 13.6 THERE WAS NO IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24605232 ↗InjuryDec 22, 2025Use of Device Problem; Use of Device Problem
WHILE WORKING ON A CELL-DYN RUBY ANALYZER, AN INDIVIDUAL MADE CONTACT WITH A SAMPLE PROBE. THE CONTACT CUT THROUGH THE GLOVE THE INDIVIDUAL WAS WEARING AND CAUSED A SMALL CUT ON THE RIGHT INDEX FINGER. THE INDIVIDUAL HAD THEIR BLOOD DRAWN AND WAS PROVIDED THE FOLLOWING REGIMEN OF ANTI-RETROVIRAL THERAPY (TENOFOVIR 300MG, DOLUTEGRAVIR 50MG + LAMIVUDINE 300MG) TO TAKE DAILY INITIALLY FOR 3 DAYS. THERE WAS NO FURTHER USER IMPACT REPORTED.
Patient problems: Needle Stick/Puncture
MAUDE report 23864387 ↗MalfunctionOct 23, 2025High Test Results; High Test Results
THE CUSTOMER OBSERVED A FALSELY ELEVATED WBC RESULT GENERATED ON THE CELL-DYN RUBY FOR A PATIENT SAMPLE. THE FOLLOWING DATA WAS PROVIDED: SAMPLE ID (B)(6), INITIAL RESULT = 60.7 10E3/UL, REPEAT RESULT = 15.1 10E3/UL . NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23372831 ↗MalfunctionSep 16, 2025High Test Results; High Test Results
THE CUSTOMER OBSERVED FALSELY ELEVATED PLATELET RESULTS GENERATED ON THE CELL DYN RUBY ANALYZER AFTER IT ENTERS A PROLONGED IDLE STATE. THE FOLLOWING DATA WAS PROVIDED (REFERENCE VALUES PLT 150-450 10E3/UL; RBC 4.2-5.4 10E6/UL; WBC 5-10 10E3/UL): PATIENT 1 SID (B)(6) 03/09/2025 7:04. WBC 6.24 10E3/UL. RBC 1.88 10E6/UL. PLT 989 10E3/UL. SID NO ID (IT IS THE SAME SAMPLE) 03/09/205 7:30. WBC 5.88 10E3/UL. RBC 2.33 10E6/UL. PLT 570 10E3/UL. SID (B)(6) 03/09/2025 7:44. WBC 9.65 10E3/UL. RBC 3.40 10E6/UL. PLT 323 10E3/UL. PATIENT 2 SID (B)(6) 03/09/2025 3:37. WBC 5.25 10E3/UL. RBC 2.12 10E6/UL. PLT 984 10E3/UL. SID NO ID (IT IS THE SAME SAMPLE) 03/09/205 3:40. WBC 5.17 10E3/UL. RBC 3.08 10E6/UL. PLT 277 10E3/UL. SID (B)(6) 03/09/2025 3:42. WBC 5.56 10E3/UL RBC 3.20 10E6/UL. PLT 285 10E3/UL. PATIENT 3 SID (B)(6) 02/09/2025 6:50. PLT 1055 10E3/UL RBC 1.56 10E6/UL WBC 5.88 10E3/UL SID (B)(6) 02/09/2025 7:01. PLT 341 10E3/UL RBC 2.57 10E6/UL WBC 6.27 10E3/UL SID (B)(6) 02/09/2025 7:08. PLT 336 10E3/UL RBC 2.57 10E6/UL WBC 6.15 10E3/UL SID (B)(6) 02/09/2025 7:11. PLT 1016 10E3/UL RBC 1.67 10E6/UL WBC 6.42 10E3/UL SID (B)(6) 02/09/2025 7:14. PLT 663 10E3/UL RBC 2.06 10E6/UL WBC 6.03 10E3/UL SI
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23067259 ↗InjuryAug 15, 2025Leak/Splash; Leak/Splash
THE ABBOTT TECHNICIAN WAS PERFORMING PREVENTATIVE MAINTENANCE ON THE CELL DYN RUBY ANALYZER AND WAS SPLASHED IN THE EYE WHEN CHANGING THE HEMOGLOBIN REAGENT SYRINGE. THE TECHNICIAN WAS NOT WEARING GOGGLES. SHE FLUSHED HER EYES WITH WATER AND WENT TO THE CLINIC TO GET MEDICAL CARE. THE TECHNICIAN WAS GIVEN EYE DROPS FOR TREATMENT (MOXIFLOXACINA 0.5% HYLO GEL EYES DROPS). THERE WAS NO REPORTED IMPACT TO PATIENT RESULTS OR PATIENT MANAGEMENT. THE TECHNICIAN WAS GIVEN EYE DROPS FOR TREATMENT (MOXIFLOXACINA 0.5% HYLO GEL EYES DROPS).
Patient problems: Eye Injury
MAUDE report 22814280 ↗InjuryJun 30, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE FIELD SERVICE REPRESENTATIVE (FSR) WAS ONSITE TROUBLESHOOTING INSTRUMENT ERROR MESSAGES ON THE CELL DYN RUBY ANALYZER AND IDENTIFIED THE MIXER HEAD HAD A LEAK. DURING REPLACEMENT OF A SPARE PART, THE FSR WAS PUNCTURED WITH THE SAMPLE PROBE. HE IMMEDIATELY WASHED THE AREA WITH SOAP, APPLIED PRESSURE TO THE LACERATION, AND CLEANED IT WITH AN ANTISEPTIC WIPE. THE FSR WAS TREATED IN THE EMERGENCY DEPARTMENT AND WAS PRESCRIBED A ONE MONTH COURSE OF ANTI-RETROVIRAL THERAPY. NO ADDITIONAL IMPACT/TREATMENT TO THE FIELD SERVICE REPRESENTATIVE WAS REPORTED.
Patient problems: Needle Stick/Puncture
MAUDE report 22358625 ↗MalfunctionJun 16, 2025Low Test Results; Low Test Results
THE CUSTOMER OBSERVED A FALSELY DEPRESSED HEMOGLOBIN CELL-DYN RUBY RESULTS FOR ONE PATIENT. THE RESULT WAS HIGHER WHEN REPEATED ON THE SAME INSTRUMENT. THE FOLLOWING DATA WAS PROVIDED: RUBY SERIAL NUMBER (B)(6), SID (B)(6), PROCESSED ON (B)(6) 2025: RESULTS FROM 7:54 AM: HEMOGLOBIN = 7.12 G/DL. OTHER RESULTS: RBC = 206 10E4/UL, MCV = 106, WBC = 95.5 10E2/UL, PLT = 27.4 10E4/UL. RESULTS FROM 8:45 AM NEW SAMPLE SAME PATIENT: HEMOGLOBIN = 8.90 G/DL. OTHER RESULTS: RBC = 260, MCV = 107, WBC = 84.6, PLT = 23.3. RESULTS FROM 9:37 AM NEW SAMPLE SAME PATIENT: HEMOGLOBIN = 9.60 G/DL. OTHER RESULTS: RBC = 283, MCV = 105, WBC = 71.7, PLT = 22.1 THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22212672 ↗MalfunctionApr 22, 2025Low Test Results; Low Test Results
THE CUSTOMER OBSERVED A FALSELY DEPRESSED WBC ON THE CELL DYN RUBY ANALYZER FOR ONE PATIENT THAT WAS NOT REPORTED OUT OF THE LABORATORY. THE CUSTOMER REPORTED THAT THE CELL DYN ANALYZER WOULD GENERATE LOW RESULTS FOR ALL ANALYTES, BUT WHEN THE SAMPLE WAS REPEATED THE RESULTS WERE HIGHER. THE FOLLOWING RESULTS WERE PROVIDED (REFERENCE RANGE WBC 6.1 ¿ 7.6 10E3/UL): INITIAL WBC RESULT= 0.64 10E3/UL; REPEAT RESULT= 8.16 10E3/UL. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21881318 ↗MalfunctionMar 31, 2025Incorrect, Inadequate or Imprecise Result or Readings; High Test Results; Incorrect, Inadequate or Imprecise Result or Readings; High Test Results
THE CUSTOMER OBSERVED FALSELY ELEVATED WBC RESULTS AND DISCREPANT HEMOGLOBIN RESULTS GENERATED BY THE CELL-DYN RUBY ANALYZER FOR A PATIENT, WHICH WERE INCONSISTENT WITH THE RESULTS FROM THE SYSMEX. THE BLOOD SMEAR WAS REVIEWED AND SHOWED THE PRESENCE OF NUCLEATED RBCS (NRBCS). THE FOLLOWING DATA WERE PROVIDED: ON (B)(6) 2025 SAMPLE ID: (B)(6) SEQUENCE#: 4570 (CBC+ RRBC TEST MODE). RUBY WBC RESULT = 92.7 X 10E3/ L (INVALIDATED WITH AN ASTERISK). RUBY HGB RESULT = 8.15 G/DL. FLAGS PRESENT: (NOC)WBC, FWBC/NWBC, DFLT (NLMEB), VAR LYM, RBC MORPH. THE CUSTOMER RELEASED THE RESULTS FROM SEQUENCE #4570 OUT OF THE LAB. UPON REVIEWING THE BLOOD SMEAR, NUMEROUS NRBCS WERE SEEN. THE SAMPLE WAS REPEATED ON THE CELL-DYN RUBY IN CBC + NOC TEST MODE AS WELL AS ON THE SYSMEX PLATFORM. THE FOLLOWING DATA WERE PROVIDED: ON (B)(6) 2025 SAMPLE ID: (B)(6) SEQUENCE#: 4571 (CBC+ NOC TEST MODE). RUBY WBC RESULT = 89.5 X 10E3/ L. RUBY HGB RESULT = 7.25 G/DL. FLAGS PRESENT: NOC, FWBC, DFLT (NLMEB), RBC MORPH. ON (B)(6) 2025 SYSMEX RESULTS: WBC = 16.31 X 10E3/ L. HGB = 5.8 G/DL. FLAGS PRESENT: WBC ABN SCATTERGRAM, LYMPHOCYTOSIS, MONOCYTOSIS, NRBC PRESENT, IG PRESENT, BLASTS/ABN LYMPHO, ANISOCYTOSIS, MACROCYTO
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21729864 ↗MalfunctionSep 23, 2024Fire; Fire
THE CUSTOMER STATED THAT THE POWER CABLE ON THE CELL DYN RUBY ANALYZER SUDDENLY CAUGHT FIRE. THE POWER CORD AND THE CELERON COMPUTER ASSEMBLY WERE REPLACED. THERE WAS NO VISIBLE DAMAGE TO THE INSTRUMENT, LAB AND TO ANY PERSONNEL. NO REPORT OF INJURIES/EXPOSURE OR IMPACT TO USER SAFETY AND NO IMPACT TO ANALYTICAL PATIENT RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20280034 ↗MalfunctionSep 19, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER OBSERVED IMPRECISE HEMOGLOBIN RESULTS GENERATED ON THE CELL DYN RUBY FOR A NEONATAL PATIENT SAMPLE. THE FOLLOWING DATA WAS PROVIDED (CUSTOMER¿S REFERENCE RANGE 12.9-14.2 G/DL): INITIAL RESULT = 16.8 G/DL, REPEAT RESULT = 6.75 G/DL, NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED. 17OCT2024 ADDITIONAL INFORMATION WAS PROVIDED FOR PATIENT PREVIOUSLY REPORTED ON (B)(6) 2024 SAMPLE ID (B)(6). ADDITIONAL RESULTS PROVIDED: WBC INITIAL RESULT = 10.4 10E3/UL, REPEAT RESULT = 13.6 10E3/UL, NEUTROPHIL INITIAL RESULT = 3.49, REPEAT RESULT = 4.13, PLATELET INITIAL RESULT = 78.6 10E3/UL, REPEAT RESULT = 129 10E3/UL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20255777 ↗MalfunctionSep 9, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER OBSERVED IMPRECISE HEMOGLOBIN RESULTS GENERATED ON THE CELL DYN RUBY ANALYZER FOR ONE PEDIATRIC SAMPLE. THE FOLLOWING DATA WAS PROVIDED: CBC RRBC MODE HEMOGLOBIN RESULT WAS 7.94; THE RESULT FROM THE PREVIOUS DAY WAS 11. REPEATED AGAIN AND THE RESULT WAS 11.5; USING THE GASOMETER THE RESULT WAS 6.7. THEY REPROCESS CBC + RRBC IN RUBY GIVING 6.75, CBC + NOC GIVES 6.8. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20171589 ↗MalfunctionAug 13, 2024Leak/Splash; Leak/Splash
THE CUSTOMER OBSERVED THE WASTE CHAMBERS AND VENT CHAMBER WERE FULL ON THE CELL-DYN RUBY ANALYZER. THE CUSTOMER WENT TO PERFORM A CLEANING OF WASTE CHAMBER ON THE ANALYZER AND A TUBING CAME OFF CAUSING A SPRAY. THE SPRAY FROM THE TUBING WENT ONTO THE CEILING AND THEN DRIPPED FROM THE CEILING ONTO THE TOP OF THE CUSTOMER¿S HEAD. THE CUSTOMER WAS WEARING GOGGLES AND A LAB COAT. THE CUSTOMER WASHED HER HAIR AND HAD NO OTHER TREATMENT. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED. THERE WAS NO ADDITIONAL HARM TO THE CUSTOMER REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19972246 ↗MalfunctionJun 28, 2024Low Test Results; Low Test Results
THE CUSTOMER OBSERVED A FALSELY DEPRESSED PLATELET ON THE CELL-DYN RUBY FOR ONE PATIENT THAT WAS NOT REPORTED OUT OF THE LABORATORY. THE FOLLOWING RESULTS WERE PROVIDED: INITIAL PLATELET= 17.5 (10E3/UL); REPEAT PLATELET= 105 (10E3/UL) THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19632978 ↗MalfunctionMay 14, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER OBSERVED IMPRECISE HEMOGLOBIN RESULTS GENERATED ON THE CELL DYN RUBY ANALYZER. THE FOLLOWING DATA WAS PROVIDED (NORMAL RANGE 12 TO 14.5): SAMPLES FROM ONE PATIENT: APRIL 14 12.7. APRIL 14 8.7. APRIL 15 8.5. APRIL 16 8.6. APRIL 16 7.8. APRIL 17 9.1. APRIL 17 8.6. APRIL 17 9.1. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19310269 ↗MalfunctionApr 30, 2024High Test Results; High Test Results
THE CUSTOMER OBSERVED ELEVATED HEMOGLOBIN (HGB) AND HEMATOCRIT (HCT) RESULTS FOR ONE PATIENT ON A CELL-DYN RUBY ANALYZER. THE FOLLOWING DATA WAS PROVIDED: SAMPLE ID (B)(4). INITIAL HGB RESULT WAS 23.6, REPEAT RESULTS WERE 15.1 AND 15.3 G/DL; INITIAL HCT RESULT WAS 71.4, REPEAT RESULTS WERE 46.9 AND 47.5%. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19212805 ↗MalfunctionMar 1, 2024Low Test Results; Low Test Results
THE CUSTOMER OBSERVED FALSELY DECREASED HEMOGLOBIN RESULTS GENERATED ON THE CELL-DYN RUBY ANALYZER FOR ONE HEMODIALYSIS PATIENT FROM MULTIPLE DRAWS. THE RESULTS PROVIDED WERE: ON (B)(6) 2024 INITIAL RESULT ON CLOSE MODE=6.8 G/DL, REPEATED SAME SPECIMEN AT ANOTHER LABORATORY ON (B)(6) 2024 (UNKNOWN PLATFORM) = 8.0 G/DL ON (B)(6) 2024 TWO SPECIMENS DRAWN FROM THE PATIENT VEIN AND CATHETER; VEIN R2V ON OPEN MODE=6.6 G/DL; CATHETER R2C ON OPEN MODE = 6.8 G/DL, REPEATED ON 09FEB2024 R2V=6.78 G/DL, SAMPLE REPEATED ON BLOOD BANK RUBY=7.17 G/DL ((B)(6)), SAME SPECIMEN PROCESS AT ANOTHER REFERENCE LABORATORY (UNKNOWN PLATFORM) =7.30 G/DL, AFTER TROUBLESHOOTING REPEATED = 6.67 G/DL ON (B)(6) 2024 NEW SPECIMEN REDRAWN AND REPEATED AT ANOTHER LAB (UNKNOWN PLATFORM) =8.0 G/DL THE CUSTOMER REPORTED THE RESULTS QUESTIONED BY THE DOCTOR, DUE TO THE LOW VALUE, HE ASKED THAT THE PATIENT BE TRANSFUSED; HOWEVER, PRIOR TO THE PROCEDURE, THE RESULT WAS CORROBORATED AND WAS HIGHER, THEREFORE THE TRANSFUSION WAS CANCELED. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18815536 ↗MalfunctionFeb 15, 2024Low Test Results; Low Test Results
THE CUSTOMER OBSERVED FALSELY DECREASED CBC RESULTS GENERATED ON THE CELL-DYN RUBY ANALYZER FOR PATIENT SAMPLES. THE FOLLOWING DATA WAS PROVIDED: PATIENT 2 PLATELET RESULT = 6.0, SLIDE REVIEW RESULT = 150 (NORMAL). NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18715227 ↗InjuryFeb 13, 2024Unintended Movement; Unintended Movement
THE ABBOTT FIELD SERVICE REPRESENTATIVE (FSR) WAS INJURED WHEN INSTALLING A CELL-DYN RUBY ANALYZER. THE FSR WAS ASSISTING THE MOVERS WITH PUTTING LIFTING HANDLES ON BOTH SIDES OF THE ANALYZER WHILE IT WAS SITTING ON A PALLET. THE FSR WAS SLIGHTLY LIFTING ONE SIDE OF THE ANALYZER SO THE MOVER COULD SLIDE THE ANALYZER OVER TO ONE SIDE OF THE PALLET, WHEN THE MOVER LIFTED THE ANALYZER, IT PUT ALL THE ANALYZER¿S WEIGHT ON THE FSR¿S ARM. THE FSR HEARD A POP AND FELT EXTREME PAIN SHOOT THROUGH THE LEFT ARM FROM BICEP MUSCLE DOWN THROUGH THE LEFT ARM. THE FSR WENT TO AN URGENT CARE FACILITY WHERE AN X-RAY FOUND NOTHING WAS BROKEN IN THE ARM, HOWEVER, THE FSR WENT TO AN ORTHOPEDIC DOCTOR THE NEXT MORNING WHERE AN ULTRASOUND CONCLUDED THAT THE TENDON FOR THE LEFT BICEP MUSCLE HAD TORN AWAY FROM THE MUSCLE AND NEEDED SURGERY TO REPAIR. THE SURGERY WAS SUCCESSFUL.
Patient problems: Muscle/Tendon Damage
MAUDE report 18699191 ↗InjuryFeb 13, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DURING THE DE-INSTALLATION PROCESS OF THE CELL DYN RUBY ANALYZER ON (B)(6) 2023, THE TECHNICAL SERVICE SPECIALIST (TSS) (49-YEAR-OLD MALE) WAS INJURED WHEN MOVING THE EQUIPMENT. THE TSS WAS CARRYING OUT THE PROCESS OF DEMOBILIZING DIAGNOSTIC EQUIPMENT CELL DYN RUBY (SERIAL (B)(6) AND AFTER UNINSTALLING THE EQUIPMENT, THE TSS WENT TO REMOVE THE EQUIPMENT FROM THE BENCH AND POSITION IT ON A CART FOR TRANSPORTATION. DUE TO THE WEIGHT (APPROXIMATELY 100KG), IT WAS NOT POSSIBLE TO MOVE ALONE, SO THE TSS REQUESTED HELP FROM A SERVICE PROVIDER FROM ANOTHER COMPANY THAT WORKS FOR THE CLIENT TO MOVE THE EQUIPMENT FROM THE BENCH (APPROXIMATELY 1.20M HIGH) TO A TRANSPORT CART (APPROXIMATELY 0.20M HIGH). WHEN THE TSS LIFTED THE EQUIPMENT, HE LOST STRENGTH IN HIS RIGHT ARM LETTING THE EQUIPMENT FALL TO THE FLOOR. AFTER A FEW MINUTES HE REGAINED STRENGTH AND FINISHED THE TASK. THE EQUIPMENT WAS TAKEN TO THE LOCATION FOR STORAGE AND SUBSEQUENT COLLECTION FROM THE CUSTOMER. HE FELT PAIN IN HIS RIGHT ARM, WHICH IMPROVED SPONTANEOUSLY. WHEN HE REMOVED HIS COAT, HE SAW THAT HIS BICEPS MUSCLE WAS RETRACTED AND DECIDED TO SEEK MEDICAL ATTENTION AT (B)(6) HOSPITAL WHERE HE WAS DIAGNOSED WITH A TENDON RU
Patient problems: Muscle/Tendon Damage
MAUDE report 18694715 ↗MalfunctionJan 26, 2024Low Test Results; Low Test Results
THE CUSTOMER OBSERVED A FALSELY DEPRESSED HEMOGLOBIN CELL-DYN RUBY RESULT WHEN COMPARED TO THE ALINITY HQ RESULT FOR ONE PATIENT. THE FOLLOWING DATA WAS PROVIDED: RUBY HEMOGLOBIN = 7.85 G/DL ALINITY HQ HEMOGLOBIN = 8.71 G/DL THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18588033 ↗MalfunctionJan 16, 2024Low Test Results; Low Test Results
THE CUSTOMER OBSERVED FALSELY DECREASED WBC AND PLATELET RESULTS GENERATED ON THE CELL DYN RUBY FOR MULTIPLE SAMPLES. THE INITIAL WBC RESULTS FOR THESE PARAMETERS WERE 0.00 OR CLOSE TO 0.00. THE DISCREPANT RESULTS WERE NOT REPORTED OUT OF THE LABORATORY. THE FOLLOWING DATA WAS PROVIDED: (B)(6) 2023 SID (B)(6) PLATELET: INITIAL RESULT = 9.02 10E3/UL, REPEAT = 184 10E3/UL SID (B)(6) WBC: INITIAL RESULT = 0.023 10E3/UL, REPEATS = 10.5 10E3/UL. (B)(6) 2023 SID (B)(6) PLATELET: INITIAL RESULT = 0.00 10E3/UL, REPEAT = 297 10E3/UL. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18522643 ↗MalfunctionJan 16, 2024Low Test Results; Low Test Results
THE CUSTOMER OBSERVED FALSELY DEPRESSED WBC AND PLATELET RESULTS GENERATED ON THE CELL DYN RUBY FOR MULTIPLE SAMPLES. THE INITIAL RESULTS FOR THESE PARAMETERS WERE 0.00 OR CLOSE TO 0.00. THE DISCREPANT RESULTS WERE NOT REPORTED OUT OF THE LABORATORY. THE FOLLOWING DATA WAS PROVIDED: (B)(6) 2023 SID (B)(6): PLATELET: INITIAL RESULT = 0.00 K/UL, REPEAT = 209 K/UL (B)(6) 2023 SID (B)(6): WBC: INITIAL RESULT = 0.001 K/UL, REPEAT = 2.04 K/UL (B)(6) 2023 SID (B)(6): WBC: INITIAL RESULT = 0.00 K/UL, REPEAT = 10.5 K/UL (B)(6) 2023 SID (B)(6): WBC: INITIAL RESULT = 0.001 K/UL, REPEATS = 6.30 K/UL AND 6.37 K/UL (B)(6) 2023 SID (B)(6): WBC: INITIAL RESULT = 0.001 K/UL, REPEAT = 6.13 K/UL (B)(6) 2023 SID (B)(6) : WBC: INITIAL RESULT = 0.00 K/UL, REPEAT = 11.3 K/UL PLATELET: INITIAL RESULT = 0.00 K/UL, REPEAT = 298 K/UL (B)(6) 2023 SID (B)(6) : PLATELET: INITIAL RESULT = 0.00 K/UL, REPEAT = 432 K/UL, 411 K/UL, AND 434 K/UL NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18516734 ↗
FDA clearance and classification
The FDA regulates product code GKZ (counter, differential cell) under 21 CFR 864.5220 as a Class II device. 21 CFR 864.5220 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the CELL-DYN 08H67-01
Has the CELL-DYN 08H67-01 been recalled?
Yes. 6 FDA recall records name this device's identifier: 6 still open. The most recent, Z-1655-2025 (Class II), began on Mar 18, 2025: When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use of expired reagents is… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the CELL-DYN 08H67-01?
The FDA's MAUDE database lists 269 adverse event reports for this device, received between Sep 15, 2016 and Sep 4, 2026. In those reports, the most common specific device problems are “Pressure Problem” (105), “Low Test Results” (57) and “High Test Results” (37); the most common patient problems are “Needle Stick/Puncture” (4), “Muscle/Tendon Damage” (2) and “Eye Injury” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the CELL-DYN 08H67-01?
None of the reports are classified as deaths, 10 as injuries and 259 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the CELL-DYN 08H67-01, and is it still sold?
The FDA's device database lists Abbott Laboratories as the company responsible for it (brand CELL-DYN, model 08H67-01). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Counter, differential cell devices
- Cytomics FC500 UPS (CXP)Beckman Coulter, Inc. · Counter, Differential Cell1,196 reports
- Alinity 09P6801Abbott Laboratories · Counter, differential cell2 recalls96 reports
- FACSCanto 338960Becton, Dickinson And Company · Counter, differential cell1 recall117 reports
- Cytomics FC500 UPS (MPL)Beckman Coulter, Inc. · Counter, Differential Cell141 reports
- CELL-DYN 08H00-01Abbott Laboratories · Counter, differential cell1 recall90 reports
- CELL-DYN 09H39-01Abbott Laboratories · Counter, differential cell123 reports
- BD Multitest 662995Becton, Dickinson And Company · Counter, differential cell2 recalls1 report
- UniCel DxH 800Beckman Coulter, Inc. · Counter, Differential Cell1 recall33 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page