Skip to content

Medtronic SynchroMed, Model A10 recallZ-0522-2024 · Medtronic Inc. · initiated Jul 13, 2020

Completed

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medtronic Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The previous software application version (1.1.300) is missing a decimal separator (a comma) for parameter range guidance values displayed on some of the programming screens: Catheter, Reservoir, Infusion, Bolus, myPTM, and Alarm.

FDA root cause category: Software Design Change

What the company is doing

Medtronic issued a SAFETY NOTIFICATION to its consignees on 07/13/2020 by letter. The notice explained the problem with the device and requested the software be upgraded to version 1.1.342.

Recall details

Recall number
Z-0522-2024
Status
Completed
Initiated by the company
Jul 13, 2020
Posted by the FDA
Dec 9, 2023
Recalling firm
Medtronic Inc.
Quantity
2543 units
Product code
LKK
FDA submissions
P860004
Distribution
Foreign Distribution

Product description

Medtronic SynchroMed, Model A10

Code information (models, lots, serial numbers)

GTIN 00643169771031, All serial numbers with software version 1.1.300.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

50 devices cleared under P860004 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates