Skip to content

CardioQuip Modular Cooler-Heater, Model: MCH-1000(i)… recallZ-0565-2024 · CardioQuip, LLC · initiated Oct 24, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from CardioQuip, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There is the potential that cooler-heater devices may contain bacterial contamination which could result in patient infection.

FDA root cause category: Process control

What the company is doing

On October 24, 2023, CardioQuip issued a "Urgent: Medical Device Correction/Removal" notification via USPS to all affected consignees. CardioQuip ask consignees to take the following actions: 1. CardioQuip recommends that you remove your potentially contaminated MCH device from service immediately. 2. CardioQuip Customer Service will contact you by November 3, 2023 to provide your facility with a loaner device and to schedule an IWPR procedure for your MCH device. 3. Please complete and return the Receipt of Communication/Response Form attached to this letter to CardioQuip as soon as possible. 4. Any adverse events experienced relating to CardioQuip devices should be reported to CardioQuip or through the FDA s MedWatch Program:

Recall details

Recall number
Z-0565-2024
Status
Open, Classified
Initiated by the company
Oct 24, 2023
Posted by the FDA
Dec 15, 2023
Recalling firm
CardioQuip, LLC
Quantity
4 units
Product code
DWC
FDA submissions
K102147
Distribution
US: California, Texas, Pennsylvania, Washington, Ohio, Nebraska OUS: None

Product description

CardioQuip Modular Cooler-Heater, Model: MCH-1000(i), Power: 115V-60Hz, 16A

Code information (models, lots, serial numbers)

UDI: 00860000846103 /Lot #:10161316, 10161479, 10160241, 10161040

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

1 device cleared under K102147 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 93446).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page