MAQUET CARDIOHELP recallZ-0555-2024 · Maquet Medical Systems USA · initiated Nov 6, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Maquet Medical Systems USA or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Certain factory settings for the device were incorrectly described in the IFU-related to the warning limits for PVen and PAux, and the default settings for the venous bubble sensor (VBS) and the automatic lock in the Minimized Extracorporeal Circulation (MECC) ThApp
FDA root cause category: Labeling design
What the company is doing
Maquet Cardiopulmonary GmbH (MCP)/Getinge issued Urgent Medical Device- Correction Customer Letter to US consignees issued on 11/6/23 via express mail. Letter states reason for recall, health risk and action to take: Please examine your inventory immediately to determine if you have any affected CARDIOHELP-i in your inventory. " Please review, download, and print the updated IFU for these products at: https://us.getinge.com/l/950532/2023-10-13/ppb1d2 " Please remove and destroy any previous versions of the IFU at your facility. " As noted in the Key User Functions section of the IFU: " Before beginning the application, ensure that the selected warning and alarm limits, as well as interventions, are suitable and safe for the patient and the current situation. " Before beginning the application, check every intervention selected by simulating an alarm condition. " Should you have any questions or need additional copies of the IFU, please contact Getinge Customer Support at (888) 943-8872 (select option 2), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone). " Please forward this information to all current and potential CARDIOHELP-i users within your hospital / facility. " If you are a distributor who has shipped any affected products to customers, please forward this document to their attention for appropriate action. Whether or not your facility has affected product(s) listed in this notice, please complete and sign the attached MEDICAL DEVICE - CORRECTION RESPONSE FORM (Page 5) to acknowledge that you have received this notification. Return the completed form to Getinge by e-mailing a scanned copy to [email protected] or by faxing the form to 1- 800-861-0842. Type of Action by Getinge: The corrected IFU for the CARDIOHELP-i Heart Lung Support System is available at the web link included in this letter. Getinge will also send a hard copy of the IFU to those customers who request one.
Recall details
- Recall number
- Z-0555-2024
- Status
- Open, Classified
- Initiated by the company
- Nov 6, 2023
- Posted by the FDA
- Dec 14, 2023
- Recalling firm
- Maquet Medical Systems USA
- Quantity
- 1595 units
- Product code
- DTQ
- FDA submissions
- K133598
- Distribution
- Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Brazil, Brunei, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, French Polynesia, Germany, Ghana, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Macau, Malaysia, Maldives, Martinique, Mauritius, Mayotte, Mexico, Nepal, Netherlands, New Zealand, Norway, Oman, Pakistan, Panama, Philippines, Poland, Portugal, Qatar, Republic of Korea, R¿union, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, Vietnam.
Product description
CARDIOHELP-i Heart Lung Support System- A blood oxygenation and carbon dioxide removal system used to pump blood through the extracorporeal bypass circuit for circulatory and/or pulmonary support for periods appropriate to cardiopulmonary bypass Part Number: 701072780, (CARDIOHELP-I US); 701048012 (CARDIOHELP-I NON US)
Code information (models, lots, serial numbers)
UDI Code: 04058863074863 (CARDIOHELP-I US); 04037691658384 (CARDIOHELP-I NONUS). All serial numbers up to and including 90415177.
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- MAQUET CARDIOHELP CARDIOHELP-iMaquet Cardiopulmonary AG · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS5 recalls2 deaths64 reports
- MAQUET CARDIOHELP CARDIOHELP SystemMaquet Cardiopulmonary AG · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS4 recalls18 deaths107 reports
Product lines
Other devices under the same FDA submission
3 devices cleared under K133598 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- MAQUET CARDIOHELP BS 3/8x3/32 L1.7Maquet Cardiopulmonary AG · Console, Heart-Lung Machine, Cardiopulmonary Bypass2 recalls0 reports
- MAQUET CARDIOHELP CARDIOHELP Emergency Drive (w/o holder)Maquet Cardiopulmonary AG · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS1 recall0 reports
- MAQUET CARDIOHELP CARDIOHELP Base UnitMaquet Cardiopulmonary AG · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS2 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page