RAPIDPoint 500 Measurement Cartridge recallZ-0606-2024 · Siemens Healthcare Diagnostics Inc · initiated Oct 26, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

RAPIDPoint 500 Systems Measurement Cartridges (with lactate) has the potential to affect the Sodium (Na+) sensor to produce a negative bias that could result in serious injury due to a delayed diagnosis of hypernatremia or unnecessary intervention for hyponatremia and result in iatrogenic hypernatremia especially if the true sodium result is near the thresholds of severe alterations, as well as cause a Question Result -----? error flag for multiple electrolytes on patient samples and quality control that may cause a delay of diagnosis or treatment.

FDA root cause category: Under Investigation by firm

What the company is doing

Siemens Healthineers issued Urgent Medical Device Correction POC 24-004.A.US on November 1, 2023 to US customers. Letter states reason for recall, health risk and action to take: Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. If you experience this issue, you may request a no-charge replacement or receive a credit from your local Siemens Healthineers representative. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. We apologize for the inconvenience this situation may cause. Siemens Healthineers is conducting a comprehensive evaluation with a cross-functional team to promptly resolve the issue with utmost priority. If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. Health Products Regulatory Authority (HPRA) has been notified by the Siemens Authorized Representative in Ireland. Local representatives for EU and non-EU regions have also been notified of this Urgent Field Safety Notice (OUS Version) and will follow local requirements for contacting their respective National Competent Authorities to advise them of the issue.

Recall details

Recall number
Z-0606-2024
Status
Open, Classified
Initiated by the company
Oct 26, 2023
Posted by the FDA
Dec 21, 2023
Recalling firm
Siemens Healthcare Diagnostics Inc
Quantity
0 cartridges US
Product code
JIX
FDA submissions
K122398
Distribution
Nationwide with Worldwide Distribution to: Albania Algeria Angola Argentina Armenia Australia Austria Azerbaijan Bahrain Bangladesh Belarus Belgium Bolivia Bosnia and Herzegovina Botswana Brazil Brunei Darussalam Bulgaria Burkina Faso Cambodia Canada Chile China Colombia Congo, Republic Costa Rica C¿te dIvoire Croatia Cura¿ao Czech Republic Denmark Ecuador Egypt Estonia Fiji Finland France French Polynesia Georgia Germany Ghana Greece Guadeloupe Honduras Hong Kong Hungary Iceland India Indonesia Iran, Islamic Republic of Iraq Ireland Israel Italy Japan Jordan Kazakhstan Kenya Kosovo Kuwait Latvia Lebanon Lesotho Lithuania Luxembourg Macedonia, former Yugoslav Rep. of Malaysia Maldives Mali Martinique Mexico Moldova, Rep. of Montenegro Namibia Nepal Netherlands New Caledonia New Zealand Norway Oman Pakistan Palestinian Territory, Occupied Paraguay Peru Philippines Poland Portugal Qatar Republic Korea Romania Russian Federation Saudi Arabia Serbia Singapore Slovakia Slovenia South Africa Spain Sweden Switzerland Taiwan, Province of China Thailand Tunisia Turkey Uganda Ukraine United Arab Emirates United Kingdom Great Britain United Republic of Tanzania Uruguay Uzbekistan Vietnam

Product description

RAPIDPoint 500 Systems Measurement Cartridge (with Lactate) 100 -Intended for in vitro diagnostic use and is designed to provide the determination in whole blood. SMN: 10844813

Code information (models, lots, serial numbers)

UDI-DI: 00630414947556 All serial numbers

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

3 devices cleared under K122398 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 93473).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page