RAPIDPoint 500 Measurement Cartridge recallZ-0609-2024 · Siemens Healthcare Diagnostics Inc · initiated Oct 26, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
RAPIDPoint 500 Systems Measurement Cartridges (with lactate) has the potential to affect the Sodium (Na+) sensor to produce a negative bias that could result in serious injury due to a delayed diagnosis of hypernatremia or unnecessary intervention for hyponatremia and result in iatrogenic hypernatremia especially if the true sodium result is near the thresholds of severe alterations, as well as cause a Question Result -----? error flag for multiple electrolytes on patient samples and quality control that may cause a delay of diagnosis or treatment.
FDA root cause category: Under Investigation by firm
What the company is doing
Siemens Healthineers issued Urgent Medical Device Correction POC 24-004.A.US on November 1, 2023 to US customers. Letter states reason for recall, health risk and action to take: Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. If you experience this issue, you may request a no-charge replacement or receive a credit from your local Siemens Healthineers representative. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. We apologize for the inconvenience this situation may cause. Siemens Healthineers is conducting a comprehensive evaluation with a cross-functional team to promptly resolve the issue with utmost priority. If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. Health Products Regulatory Authority (HPRA) has been notified by the Siemens Authorized Representative in Ireland. Local representatives for EU and non-EU regions have also been notified of this Urgent Field Safety Notice (OUS Version) and will follow local requirements for contacting their respective National Competent Authorities to advise them of the issue.
Recall details
- Recall number
- Z-0609-2024
- Status
- Open, Classified
- Initiated by the company
- Oct 26, 2023
- Posted by the FDA
- Dec 21, 2023
- Recalling firm
- Siemens Healthcare Diagnostics Inc
- Quantity
- 3505 Cartridges US
- Product code
- JIX
- FDA submissions
- K122398
- Distribution
- Nationwide with Worldwide Distribution to: Albania Algeria Angola Argentina Armenia Australia Austria Azerbaijan Bahrain Bangladesh Belarus Belgium Bolivia Bosnia and Herzegovina Botswana Brazil Brunei Darussalam Bulgaria Burkina Faso Cambodia Canada Chile China Colombia Congo, Republic Costa Rica C¿te dIvoire Croatia Cura¿ao Czech Republic Denmark Ecuador Egypt Estonia Fiji Finland France French Polynesia Georgia Germany Ghana Greece Guadeloupe Honduras Hong Kong Hungary Iceland India Indonesia Iran, Islamic Republic of Iraq Ireland Israel Italy Japan Jordan Kazakhstan Kenya Kosovo Kuwait Latvia Lebanon Lesotho Lithuania Luxembourg Macedonia, former Yugoslav Rep. of Malaysia Maldives Mali Martinique Mexico Moldova, Rep. of Montenegro Namibia Nepal Netherlands New Caledonia New Zealand Norway Oman Pakistan Palestinian Territory, Occupied Paraguay Peru Philippines Poland Portugal Qatar Republic Korea Romania Russian Federation Saudi Arabia Serbia Singapore Slovakia Slovenia South Africa Spain Sweden Switzerland Taiwan, Province of China Thailand Tunisia Turkey Uganda Ukraine United Arab Emirates United Kingdom Great Britain United Republic of Tanzania Uruguay Uzbekistan Vietnam
Product description
RAPIDPoint 500 Systems Measurement Cartridge (with Lactate) 750-Intended for in vitro diagnostic use and is designed to provide the determination in whole blood. SMN: 10491449
Code information (models, lots, serial numbers)
UDI-DI: 00630414589770 All serial numbers
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- RAPIDPoint® 500 Measurement Cartridge 10491449Siemens Healthcare Diagnostics Inc. · CALIBRATOR, MULTI-ANALYTE MIXTURE1 recall3 reports
Product lines
Other devices under the same FDA submission
3 devices cleared under K122398 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- RAPIDPoint® 500 Measurement Cartridge 10491447Siemens Healthcare Diagnostics Inc. · CALIBRATOR, MULTI-ANALYTE MIXTURE1 recall0 reports
- RAPIDPoint® 500 Measurement Cartridge 10491448Siemens Healthcare Diagnostics Inc. · CALIBRATOR, MULTI-ANALYTE MIXTURE1 recall0 reports
- RAPIDPoint® 500 Measurement Cartridge 10844813Siemens Healthcare Diagnostics Inc. · CALIBRATOR, MULTI-ANALYTE MIXTURE1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 93473).
- Z-0606-2024 · Open, ClassifiedRAPIDPoint 500 Systems Measurement Cartridge (with Lactate) 100 -Intended for in vitro diagnostic use and is designed to provide the determination in whole blood. SMN: 10844813
- Z-0607-2024 · Open, ClassifiedRAPIDPoint 500 Systems Measurement Cartridge (with Lactate) 250-Intended for in vitro diagnostic use and is designed to provide the determination in whole blood. SMN: 10491447
- Z-0608-2024 · Open, ClassifiedRAPIDPoint 500 Systems Measurement Cartridge (with Lactate) 400-Intended for in vitro diagnostic use and is designed to provide the determination in whole blood. SMN: 10491448
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page