Maquet Rolite recallZ-0760-2024 · Getinge Usa Sales Inc · initiated Nov 8, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Getinge Usa Sales Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A potential for a light system to fall in the operating room.

FDA root cause category: Equipment maintenance

What the company is doing

An URGENT MEDICAL DEVICE CORRECTION letter dated 11/8/23 was sent to customers. Actions to be taken by the customer: Our records indicate that you have been identified as having purchased at least one model of one of these product ranges. Please identify your affected units using the referenced devices on page 1 and follow the instructions below. Please distribute this notice and forward the link below to access manuals to anyone within your organization who may need to use the manuals, and especially to your service or service provider who performs your maintenance. Light system fixing and other replacements: Regarding: - the fixing of the light systems: suspensions fixing screws, adapter fixing screws, bushing fixing screws - the brake screws - the safety segments - the batteries, a dedicated instruction to illustrate and facilitate understanding of what to replace and what to order for replacement has been issued. The instruction is available on a dedicated space on the web portal https://www.getinge.com/int/campaigns/maintenance-and-service-on-or-light-system/, and provided upon request by emailing [email protected] Should you have questions or require additional information, please do not hesitate to contact Maquet SAS/Getinge or your local representative. Please complete and sign the attached MEDICAL DEVICE CORRECTION - RESPONSE FORM (page 7) to acknowledge that you have received this notification. Return the completed form to Maquet/Getinge by e-mailing a scanned copy to [email protected] or by faxing the form to (866) 350-0897. If you are a distributor who has shipped any affected products to customers, please forward this document to their attention for appropriate action. Actions to be taken by Maquet SAS / Getinge: As the majority of customers may not have regular contact with Getinge services or authorized Getinge representatives and therefore may not be aware of current maintenance recommendations, Maquet SAS/Getinge is providin

Recall details

Recall number
Z-0760-2024
Status
Open, Classified
Initiated by the company
Nov 8, 2023
Posted by the FDA
Jan 19, 2024
Recalling firm
Getinge Usa Sales Inc
Quantity
236793 units in total
Product code
FTD
Distribution
US Nationwide. Global Distribution.

Product description

Maquet Rolite OR Light Systems, Model Numbers and Catalog Numbers ROL PWDII50SF - ARDPWT309001A, ROL PWDII50SF - ARDPWT309002A, ROL PWDII50SF - ARDPWT309009A, ROL PWDII50SF - ARDPWT309010A, ROL PWDII70SF - ARDPWT309007A, ROL PWDII70SF - ARDPWT309008A, ROL PWDII70SF - ARDPWT309015A, ROL PWDII70SF - ARDPWT309016A, ROL VCS40SF - ARDROL309031A, ROL VCS60SF - ARDROL309032A, ROL VST40SF - ARDROL309033A, ROL VST40SF - ARDROL309034A, ROL VST40SF B - ARDROL309037A, ROL VST60SF - ARDROL309035A, ROL VST60SF - ARDROL309036A, ROL VST60SF B - ARDROL309038A, ROL VCSII40SF - ARDROL309039A, ROL VCSII40SF - ARDROL309040A, ROL VCSII60SF - ARDROL309041A, ROL VCSII60SF - ARDROL309042A, ROL PWDII50SF - ARDROL309043A, ROL PWDII50SF - ARDROL309044A, ROL PWDII70SF - ARDROL309045A, ROL PWDII70SF - ARDROL309046A, ROL PWDII50SF - ARDROL309047A, ROL PWDII50SF - ARDROL309048A, ROL PWDII70SF - ARDROL309049A, ROL PWDII70SF - ARDROL309050A, ROL VSTII40SF - ARDROL309051A, ROL VSTII60SF - ARDROL309052A, ROL VSTII40SF - ARDROL309053A, ROL VSTII60SF - ARDROL309054A

Code information (models, lots, serial numbers)

All systems affected UDI-DI: 3700712411978 3700712411985 3700712412050 3700712412067 3700712412036 3700712412043 3700712412111 3700712412128 3700712417062 3700712417079 3700712417086 3700712417093 3700712419240 3700712417109 3700712417116 3700712419257 3700712422707 3700712422714 3700712422721 3700712422738 3700712422745 3700712422752 3700712422769 3700712422776 3700712422783 3700712422790 3700712422806 3700712422813 3700712422820 3700712422837 3700712422844 3700712422851

Devices named in this recall

15 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 93567).